Toroidal coil chromatography: The effect of scale-up and “g” field on stage efficiency

Selected test results have been taken from various publications and resolution and stage efficiency measured using an established model. All experiments used the same sample and, where possible, the same sample loading. The results show that stage mixing efficiencies have increased from 1.1% in 1998...

Descripción completa

Detalles Bibliográficos
Autores principales: Sutherland, Ian, Hewitson, Peter, de Folter, Joost
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3223526/
https://www.ncbi.nlm.nih.gov/pubmed/21227439
http://dx.doi.org/10.1016/j.chroma.2010.12.090
Descripción
Sumario:Selected test results have been taken from various publications and resolution and stage efficiency measured using an established model. All experiments used the same sample and, where possible, the same sample loading. The results show that stage mixing efficiencies have increased from 1.1% in 1998 to greater than 25% in the latest scaled-up version of a Toroidal coil chromatography (TCC) instrument working at 240 g.