Buffer substitution in malaria rapid diagnostic tests causes false-positive results
BACKGROUND: Malaria rapid diagnostic tests (RDTs) are kits that generally include 20 to 25 test strips or cassettes, but only a single buffer vial. In field settings, laboratory staff occasionally uses saline, distilled water (liquids for parenteral drugs dilution) or tap water as substitutes for th...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3224932/ https://www.ncbi.nlm.nih.gov/pubmed/20650003 http://dx.doi.org/10.1186/1475-2875-9-215 |
_version_ | 1782217462723379200 |
---|---|
author | Gillet, Philippe Mori, Marcella Van den Ende, Jef Jacobs, Jan |
author_facet | Gillet, Philippe Mori, Marcella Van den Ende, Jef Jacobs, Jan |
author_sort | Gillet, Philippe |
collection | PubMed |
description | BACKGROUND: Malaria rapid diagnostic tests (RDTs) are kits that generally include 20 to 25 test strips or cassettes, but only a single buffer vial. In field settings, laboratory staff occasionally uses saline, distilled water (liquids for parenteral drugs dilution) or tap water as substitutes for the RDT kit's buffer to compensate for the loss of a diluent bottle. The present study assessed the effect of buffer substitution on the RDT results. METHODS: Twenty-seven RDT brands were run with EDTA-blood samples of five malaria-free subjects, who were negative for rheumatoid factor and antinuclear antibodies. Saline, distilled water and tap water were used as substitute liquids. RDTs were also run with distilled water, without adding blood. Results were compared to those obtained with the RDT kit's buffer and Plasmodium positive samples. RESULTS: Only eight cassettes (in four RDT brands) showed no control line and were considered invalid. Visible test lines occurred for at least one malaria-free sample and one of the substitutes in 20/27 (74%) RDT brands (saline: n = 16; distilled water: n = 17; and tap water: n = 20), and in 15 RDTs which were run with distilled water only. They occurred for all Plasmodium antigens and RDT formats (two-, three- and four-band RDTs). Clearance of the background of the strip was excellent except for saline. The aspects (colour, intensity and crispness) of the control and the false-positive test lines were similar to those obtained with the RDT kits' buffer and Plasmodium positive samples. CONCLUSION: Replacement of the RDT kit's dedicated buffer by saline, distilled water and tap water can cause false-positive test results. |
format | Online Article Text |
id | pubmed-3224932 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2010 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-32249322011-11-29 Buffer substitution in malaria rapid diagnostic tests causes false-positive results Gillet, Philippe Mori, Marcella Van den Ende, Jef Jacobs, Jan Malar J Methodology BACKGROUND: Malaria rapid diagnostic tests (RDTs) are kits that generally include 20 to 25 test strips or cassettes, but only a single buffer vial. In field settings, laboratory staff occasionally uses saline, distilled water (liquids for parenteral drugs dilution) or tap water as substitutes for the RDT kit's buffer to compensate for the loss of a diluent bottle. The present study assessed the effect of buffer substitution on the RDT results. METHODS: Twenty-seven RDT brands were run with EDTA-blood samples of five malaria-free subjects, who were negative for rheumatoid factor and antinuclear antibodies. Saline, distilled water and tap water were used as substitute liquids. RDTs were also run with distilled water, without adding blood. Results were compared to those obtained with the RDT kit's buffer and Plasmodium positive samples. RESULTS: Only eight cassettes (in four RDT brands) showed no control line and were considered invalid. Visible test lines occurred for at least one malaria-free sample and one of the substitutes in 20/27 (74%) RDT brands (saline: n = 16; distilled water: n = 17; and tap water: n = 20), and in 15 RDTs which were run with distilled water only. They occurred for all Plasmodium antigens and RDT formats (two-, three- and four-band RDTs). Clearance of the background of the strip was excellent except for saline. The aspects (colour, intensity and crispness) of the control and the false-positive test lines were similar to those obtained with the RDT kits' buffer and Plasmodium positive samples. CONCLUSION: Replacement of the RDT kit's dedicated buffer by saline, distilled water and tap water can cause false-positive test results. BioMed Central 2010-07-22 /pmc/articles/PMC3224932/ /pubmed/20650003 http://dx.doi.org/10.1186/1475-2875-9-215 Text en Copyright ©2010 Gillet et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Methodology Gillet, Philippe Mori, Marcella Van den Ende, Jef Jacobs, Jan Buffer substitution in malaria rapid diagnostic tests causes false-positive results |
title | Buffer substitution in malaria rapid diagnostic tests causes false-positive results |
title_full | Buffer substitution in malaria rapid diagnostic tests causes false-positive results |
title_fullStr | Buffer substitution in malaria rapid diagnostic tests causes false-positive results |
title_full_unstemmed | Buffer substitution in malaria rapid diagnostic tests causes false-positive results |
title_short | Buffer substitution in malaria rapid diagnostic tests causes false-positive results |
title_sort | buffer substitution in malaria rapid diagnostic tests causes false-positive results |
topic | Methodology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3224932/ https://www.ncbi.nlm.nih.gov/pubmed/20650003 http://dx.doi.org/10.1186/1475-2875-9-215 |
work_keys_str_mv | AT gilletphilippe buffersubstitutioninmalariarapiddiagnostictestscausesfalsepositiveresults AT morimarcella buffersubstitutioninmalariarapiddiagnostictestscausesfalsepositiveresults AT vandenendejef buffersubstitutioninmalariarapiddiagnostictestscausesfalsepositiveresults AT jacobsjan buffersubstitutioninmalariarapiddiagnostictestscausesfalsepositiveresults |