Buffer substitution in malaria rapid diagnostic tests causes false-positive results

BACKGROUND: Malaria rapid diagnostic tests (RDTs) are kits that generally include 20 to 25 test strips or cassettes, but only a single buffer vial. In field settings, laboratory staff occasionally uses saline, distilled water (liquids for parenteral drugs dilution) or tap water as substitutes for th...

Descripción completa

Detalles Bibliográficos
Autores principales: Gillet, Philippe, Mori, Marcella, Van den Ende, Jef, Jacobs, Jan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3224932/
https://www.ncbi.nlm.nih.gov/pubmed/20650003
http://dx.doi.org/10.1186/1475-2875-9-215
_version_ 1782217462723379200
author Gillet, Philippe
Mori, Marcella
Van den Ende, Jef
Jacobs, Jan
author_facet Gillet, Philippe
Mori, Marcella
Van den Ende, Jef
Jacobs, Jan
author_sort Gillet, Philippe
collection PubMed
description BACKGROUND: Malaria rapid diagnostic tests (RDTs) are kits that generally include 20 to 25 test strips or cassettes, but only a single buffer vial. In field settings, laboratory staff occasionally uses saline, distilled water (liquids for parenteral drugs dilution) or tap water as substitutes for the RDT kit's buffer to compensate for the loss of a diluent bottle. The present study assessed the effect of buffer substitution on the RDT results. METHODS: Twenty-seven RDT brands were run with EDTA-blood samples of five malaria-free subjects, who were negative for rheumatoid factor and antinuclear antibodies. Saline, distilled water and tap water were used as substitute liquids. RDTs were also run with distilled water, without adding blood. Results were compared to those obtained with the RDT kit's buffer and Plasmodium positive samples. RESULTS: Only eight cassettes (in four RDT brands) showed no control line and were considered invalid. Visible test lines occurred for at least one malaria-free sample and one of the substitutes in 20/27 (74%) RDT brands (saline: n = 16; distilled water: n = 17; and tap water: n = 20), and in 15 RDTs which were run with distilled water only. They occurred for all Plasmodium antigens and RDT formats (two-, three- and four-band RDTs). Clearance of the background of the strip was excellent except for saline. The aspects (colour, intensity and crispness) of the control and the false-positive test lines were similar to those obtained with the RDT kits' buffer and Plasmodium positive samples. CONCLUSION: Replacement of the RDT kit's dedicated buffer by saline, distilled water and tap water can cause false-positive test results.
format Online
Article
Text
id pubmed-3224932
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-32249322011-11-29 Buffer substitution in malaria rapid diagnostic tests causes false-positive results Gillet, Philippe Mori, Marcella Van den Ende, Jef Jacobs, Jan Malar J Methodology BACKGROUND: Malaria rapid diagnostic tests (RDTs) are kits that generally include 20 to 25 test strips or cassettes, but only a single buffer vial. In field settings, laboratory staff occasionally uses saline, distilled water (liquids for parenteral drugs dilution) or tap water as substitutes for the RDT kit's buffer to compensate for the loss of a diluent bottle. The present study assessed the effect of buffer substitution on the RDT results. METHODS: Twenty-seven RDT brands were run with EDTA-blood samples of five malaria-free subjects, who were negative for rheumatoid factor and antinuclear antibodies. Saline, distilled water and tap water were used as substitute liquids. RDTs were also run with distilled water, without adding blood. Results were compared to those obtained with the RDT kit's buffer and Plasmodium positive samples. RESULTS: Only eight cassettes (in four RDT brands) showed no control line and were considered invalid. Visible test lines occurred for at least one malaria-free sample and one of the substitutes in 20/27 (74%) RDT brands (saline: n = 16; distilled water: n = 17; and tap water: n = 20), and in 15 RDTs which were run with distilled water only. They occurred for all Plasmodium antigens and RDT formats (two-, three- and four-band RDTs). Clearance of the background of the strip was excellent except for saline. The aspects (colour, intensity and crispness) of the control and the false-positive test lines were similar to those obtained with the RDT kits' buffer and Plasmodium positive samples. CONCLUSION: Replacement of the RDT kit's dedicated buffer by saline, distilled water and tap water can cause false-positive test results. BioMed Central 2010-07-22 /pmc/articles/PMC3224932/ /pubmed/20650003 http://dx.doi.org/10.1186/1475-2875-9-215 Text en Copyright ©2010 Gillet et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Methodology
Gillet, Philippe
Mori, Marcella
Van den Ende, Jef
Jacobs, Jan
Buffer substitution in malaria rapid diagnostic tests causes false-positive results
title Buffer substitution in malaria rapid diagnostic tests causes false-positive results
title_full Buffer substitution in malaria rapid diagnostic tests causes false-positive results
title_fullStr Buffer substitution in malaria rapid diagnostic tests causes false-positive results
title_full_unstemmed Buffer substitution in malaria rapid diagnostic tests causes false-positive results
title_short Buffer substitution in malaria rapid diagnostic tests causes false-positive results
title_sort buffer substitution in malaria rapid diagnostic tests causes false-positive results
topic Methodology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3224932/
https://www.ncbi.nlm.nih.gov/pubmed/20650003
http://dx.doi.org/10.1186/1475-2875-9-215
work_keys_str_mv AT gilletphilippe buffersubstitutioninmalariarapiddiagnostictestscausesfalsepositiveresults
AT morimarcella buffersubstitutioninmalariarapiddiagnostictestscausesfalsepositiveresults
AT vandenendejef buffersubstitutioninmalariarapiddiagnostictestscausesfalsepositiveresults
AT jacobsjan buffersubstitutioninmalariarapiddiagnostictestscausesfalsepositiveresults