Laboratory evaluation on the sensitivity and specificity of a novel and rapid detection method for malaria diagnosis based on magneto-optical technology (MOT)

BACKGROUND: This study describes the laboratory evaluation of a novel diagnostic platform for malaria. The Magneto Optical Test (MOT) is based on the bio-physical detection of haemozoin in clinical samples. Having an assay time of around one minute, it offers the potential of high throughput screeni...

Descripción completa

Detalles Bibliográficos
Autores principales: Mens, Petra F, Matelon, Raphael J, Nour, Bakri YM, Newman, Dave M, Schallig, Henk DFH
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3224991/
https://www.ncbi.nlm.nih.gov/pubmed/20642834
http://dx.doi.org/10.1186/1475-2875-9-207
_version_ 1782217471644663808
author Mens, Petra F
Matelon, Raphael J
Nour, Bakri YM
Newman, Dave M
Schallig, Henk DFH
author_facet Mens, Petra F
Matelon, Raphael J
Nour, Bakri YM
Newman, Dave M
Schallig, Henk DFH
author_sort Mens, Petra F
collection PubMed
description BACKGROUND: This study describes the laboratory evaluation of a novel diagnostic platform for malaria. The Magneto Optical Test (MOT) is based on the bio-physical detection of haemozoin in clinical samples. Having an assay time of around one minute, it offers the potential of high throughput screening. METHODS: Blood samples of confirmed malaria patients from different regions of Africa, patients with other diseases and healthy non-endemic controls were used in the present study. The samples were analysed with two reference tests, i.e. an histidine rich protein-2 based rapid diagnostic test (RDT) and a conventional Pan-Plasmodium PCR, and the MOT as index test. Data were entered in 2 × 2 tables and analysed for sensitivity and specificity. The agreement between microscopy, RDT and PCR and the MOT assay was determined by calculating Kappa values with a 95% confidence interval. RESULTS: The observed sensitivity/specificity of the MOT test in comparison with clinical description, RDT or PCR ranged from 77.2 - 78.8% (sensitivity) and from 72.5 - 74.6% (specificity). In general, the agreement between MOT and the other assays is around 0.5 indicating a moderate agreement between the reference and the index test. However, when RDT and PCR are compared to each other, an almost perfect agreement can be observed (k = 0.97) with a sensitivity and specificity of >95%. CONCLUSIONS: Although MOT sensitivity and specificity are currently not yet at a competing level compared to other diagnostic test, such as PCR and RDTs, it has a potential to rapidly screen patients for malaria in endemic as well as non-endemic countries.
format Online
Article
Text
id pubmed-3224991
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-32249912011-11-30 Laboratory evaluation on the sensitivity and specificity of a novel and rapid detection method for malaria diagnosis based on magneto-optical technology (MOT) Mens, Petra F Matelon, Raphael J Nour, Bakri YM Newman, Dave M Schallig, Henk DFH Malar J Research BACKGROUND: This study describes the laboratory evaluation of a novel diagnostic platform for malaria. The Magneto Optical Test (MOT) is based on the bio-physical detection of haemozoin in clinical samples. Having an assay time of around one minute, it offers the potential of high throughput screening. METHODS: Blood samples of confirmed malaria patients from different regions of Africa, patients with other diseases and healthy non-endemic controls were used in the present study. The samples were analysed with two reference tests, i.e. an histidine rich protein-2 based rapid diagnostic test (RDT) and a conventional Pan-Plasmodium PCR, and the MOT as index test. Data were entered in 2 × 2 tables and analysed for sensitivity and specificity. The agreement between microscopy, RDT and PCR and the MOT assay was determined by calculating Kappa values with a 95% confidence interval. RESULTS: The observed sensitivity/specificity of the MOT test in comparison with clinical description, RDT or PCR ranged from 77.2 - 78.8% (sensitivity) and from 72.5 - 74.6% (specificity). In general, the agreement between MOT and the other assays is around 0.5 indicating a moderate agreement between the reference and the index test. However, when RDT and PCR are compared to each other, an almost perfect agreement can be observed (k = 0.97) with a sensitivity and specificity of >95%. CONCLUSIONS: Although MOT sensitivity and specificity are currently not yet at a competing level compared to other diagnostic test, such as PCR and RDTs, it has a potential to rapidly screen patients for malaria in endemic as well as non-endemic countries. BioMed Central 2010-07-19 /pmc/articles/PMC3224991/ /pubmed/20642834 http://dx.doi.org/10.1186/1475-2875-9-207 Text en Copyright ©2010 Mens et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Mens, Petra F
Matelon, Raphael J
Nour, Bakri YM
Newman, Dave M
Schallig, Henk DFH
Laboratory evaluation on the sensitivity and specificity of a novel and rapid detection method for malaria diagnosis based on magneto-optical technology (MOT)
title Laboratory evaluation on the sensitivity and specificity of a novel and rapid detection method for malaria diagnosis based on magneto-optical technology (MOT)
title_full Laboratory evaluation on the sensitivity and specificity of a novel and rapid detection method for malaria diagnosis based on magneto-optical technology (MOT)
title_fullStr Laboratory evaluation on the sensitivity and specificity of a novel and rapid detection method for malaria diagnosis based on magneto-optical technology (MOT)
title_full_unstemmed Laboratory evaluation on the sensitivity and specificity of a novel and rapid detection method for malaria diagnosis based on magneto-optical technology (MOT)
title_short Laboratory evaluation on the sensitivity and specificity of a novel and rapid detection method for malaria diagnosis based on magneto-optical technology (MOT)
title_sort laboratory evaluation on the sensitivity and specificity of a novel and rapid detection method for malaria diagnosis based on magneto-optical technology (mot)
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3224991/
https://www.ncbi.nlm.nih.gov/pubmed/20642834
http://dx.doi.org/10.1186/1475-2875-9-207
work_keys_str_mv AT menspetraf laboratoryevaluationonthesensitivityandspecificityofanovelandrapiddetectionmethodformalariadiagnosisbasedonmagnetoopticaltechnologymot
AT matelonraphaelj laboratoryevaluationonthesensitivityandspecificityofanovelandrapiddetectionmethodformalariadiagnosisbasedonmagnetoopticaltechnologymot
AT nourbakriym laboratoryevaluationonthesensitivityandspecificityofanovelandrapiddetectionmethodformalariadiagnosisbasedonmagnetoopticaltechnologymot
AT newmandavem laboratoryevaluationonthesensitivityandspecificityofanovelandrapiddetectionmethodformalariadiagnosisbasedonmagnetoopticaltechnologymot
AT schallighenkdfh laboratoryevaluationonthesensitivityandspecificityofanovelandrapiddetectionmethodformalariadiagnosisbasedonmagnetoopticaltechnologymot