Sensitivities of a Standard Test Method for the Determination of the pHe of Bioethanol and Suggestions for Improvement
An assessment of the sensitivities of the critical parameters in the ASTM D6423 documentary standard method for the measurement of pHe in (bio)ethanol has been undertaken. Repeatability of measurements made using the same glass electrode and reproducibility between different glass electrodes have be...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Molecular Diversity Preservation International (MDPI)
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3231013/ https://www.ncbi.nlm.nih.gov/pubmed/22163451 http://dx.doi.org/10.3390/s101109982 |
_version_ | 1782218123440553984 |
---|---|
author | Brown, Richard J. C. Keates, Adam C. Brewer, Paul J. |
author_facet | Brown, Richard J. C. Keates, Adam C. Brewer, Paul J. |
author_sort | Brown, Richard J. C. |
collection | PubMed |
description | An assessment of the sensitivities of the critical parameters in the ASTM D6423 documentary standard method for the measurement of pHe in (bio)ethanol has been undertaken. Repeatability of measurements made using the same glass electrode and reproducibility between different glass electrodes have been identified as the main contributors to the uncertainty of the values produced. Strategies to reduce the uncertainty of the measurement have been identified and tested. Both increasing the time after which the pHe measurement is made following immersion in the sample, and rinsing the glass electrode with ethanol prior to immersion in the sample, have been shown to be effective in reducing the uncertainty of the numerical value produced. However, it is acknowledged that the values produced using these modified approaches may not be directly compared with those obtained using the documentary ASTM method since pHe is defined operationally by the process used to measure it. |
format | Online Article Text |
id | pubmed-3231013 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2010 |
publisher | Molecular Diversity Preservation International (MDPI) |
record_format | MEDLINE/PubMed |
spelling | pubmed-32310132011-12-07 Sensitivities of a Standard Test Method for the Determination of the pHe of Bioethanol and Suggestions for Improvement Brown, Richard J. C. Keates, Adam C. Brewer, Paul J. Sensors (Basel) Article An assessment of the sensitivities of the critical parameters in the ASTM D6423 documentary standard method for the measurement of pHe in (bio)ethanol has been undertaken. Repeatability of measurements made using the same glass electrode and reproducibility between different glass electrodes have been identified as the main contributors to the uncertainty of the values produced. Strategies to reduce the uncertainty of the measurement have been identified and tested. Both increasing the time after which the pHe measurement is made following immersion in the sample, and rinsing the glass electrode with ethanol prior to immersion in the sample, have been shown to be effective in reducing the uncertainty of the numerical value produced. However, it is acknowledged that the values produced using these modified approaches may not be directly compared with those obtained using the documentary ASTM method since pHe is defined operationally by the process used to measure it. Molecular Diversity Preservation International (MDPI) 2010-11-09 /pmc/articles/PMC3231013/ /pubmed/22163451 http://dx.doi.org/10.3390/s101109982 Text en © 2010 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/. (http://creativecommons.org/licenses/by/3.0/) ) |
spellingShingle | Article Brown, Richard J. C. Keates, Adam C. Brewer, Paul J. Sensitivities of a Standard Test Method for the Determination of the pHe of Bioethanol and Suggestions for Improvement |
title | Sensitivities of a Standard Test Method for the Determination of the pHe of Bioethanol and Suggestions for Improvement |
title_full | Sensitivities of a Standard Test Method for the Determination of the pHe of Bioethanol and Suggestions for Improvement |
title_fullStr | Sensitivities of a Standard Test Method for the Determination of the pHe of Bioethanol and Suggestions for Improvement |
title_full_unstemmed | Sensitivities of a Standard Test Method for the Determination of the pHe of Bioethanol and Suggestions for Improvement |
title_short | Sensitivities of a Standard Test Method for the Determination of the pHe of Bioethanol and Suggestions for Improvement |
title_sort | sensitivities of a standard test method for the determination of the phe of bioethanol and suggestions for improvement |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3231013/ https://www.ncbi.nlm.nih.gov/pubmed/22163451 http://dx.doi.org/10.3390/s101109982 |
work_keys_str_mv | AT brownrichardjc sensitivitiesofastandardtestmethodforthedeterminationofthepheofbioethanolandsuggestionsforimprovement AT keatesadamc sensitivitiesofastandardtestmethodforthedeterminationofthepheofbioethanolandsuggestionsforimprovement AT brewerpaulj sensitivitiesofastandardtestmethodforthedeterminationofthepheofbioethanolandsuggestionsforimprovement |