Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy
The objective of this study was to evaluate the acute response of Sencil(™), a fiber-optic sensor, in point accuracy for glucose monitoring in vivo on healthy dogs under anesthesia. A total of four dogs with clinically normal glycemia were implanted with one sensor each in the chest region to measur...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Molecular Diversity Preservation International (MDPI)
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3231153/ https://www.ncbi.nlm.nih.gov/pubmed/22163627 http://dx.doi.org/10.3390/s100807789 |
_version_ | 1782218156004081664 |
---|---|
author | Liao, Kuo-Chih Chang, Shih-Chieh Chiu, Cheng-Yang Chou, Yu-Hsiang |
author_facet | Liao, Kuo-Chih Chang, Shih-Chieh Chiu, Cheng-Yang Chou, Yu-Hsiang |
author_sort | Liao, Kuo-Chih |
collection | PubMed |
description | The objective of this study was to evaluate the acute response of Sencil(™), a fiber-optic sensor, in point accuracy for glucose monitoring in vivo on healthy dogs under anesthesia. A total of four dogs with clinically normal glycemia were implanted with one sensor each in the chest region to measure the interstitial glucose concentration during the ovariohysterectomy procedure. The data was acquired every 10 seconds after initiation, and was compared to the concentration of venous plasma glucose sampled during the surgery procedures for accuracy of agreement analysis. In the four trials with a range of 71–297 mg/dL plasma glucose, the collected 21 pairs of ISF readings from the Sencil™ and the plasma reference showed superior dispersion of residue values than the conventional system, and a linear correlation (the Pearson correlation coefficient is 0.9288 and the y-intercept is 14.22 mg/dL). The MAD (17.6 mg/dL) and RMAD (16.16%) of Sencil™ measurements were in the comparable range of the conventional system. The Clarke error grid analysis indicated that 100% of the paired points were in the clinically acceptable zone A (61.9%) and B (38.1%). |
format | Online Article Text |
id | pubmed-3231153 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2010 |
publisher | Molecular Diversity Preservation International (MDPI) |
record_format | MEDLINE/PubMed |
spelling | pubmed-32311532011-12-07 Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy Liao, Kuo-Chih Chang, Shih-Chieh Chiu, Cheng-Yang Chou, Yu-Hsiang Sensors (Basel) Article The objective of this study was to evaluate the acute response of Sencil(™), a fiber-optic sensor, in point accuracy for glucose monitoring in vivo on healthy dogs under anesthesia. A total of four dogs with clinically normal glycemia were implanted with one sensor each in the chest region to measure the interstitial glucose concentration during the ovariohysterectomy procedure. The data was acquired every 10 seconds after initiation, and was compared to the concentration of venous plasma glucose sampled during the surgery procedures for accuracy of agreement analysis. In the four trials with a range of 71–297 mg/dL plasma glucose, the collected 21 pairs of ISF readings from the Sencil™ and the plasma reference showed superior dispersion of residue values than the conventional system, and a linear correlation (the Pearson correlation coefficient is 0.9288 and the y-intercept is 14.22 mg/dL). The MAD (17.6 mg/dL) and RMAD (16.16%) of Sencil™ measurements were in the comparable range of the conventional system. The Clarke error grid analysis indicated that 100% of the paired points were in the clinically acceptable zone A (61.9%) and B (38.1%). Molecular Diversity Preservation International (MDPI) 2010-08-20 /pmc/articles/PMC3231153/ /pubmed/22163627 http://dx.doi.org/10.3390/s100807789 Text en © 2010 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/). |
spellingShingle | Article Liao, Kuo-Chih Chang, Shih-Chieh Chiu, Cheng-Yang Chou, Yu-Hsiang Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy |
title | Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy |
title_full | Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy |
title_fullStr | Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy |
title_full_unstemmed | Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy |
title_short | Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy |
title_sort | acute response in vivo of a fiber-optic sensor for continuous glucose monitoring from canine studies on point accuracy |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3231153/ https://www.ncbi.nlm.nih.gov/pubmed/22163627 http://dx.doi.org/10.3390/s100807789 |
work_keys_str_mv | AT liaokuochih acuteresponseinvivoofafiberopticsensorforcontinuousglucosemonitoringfromcaninestudiesonpointaccuracy AT changshihchieh acuteresponseinvivoofafiberopticsensorforcontinuousglucosemonitoringfromcaninestudiesonpointaccuracy AT chiuchengyang acuteresponseinvivoofafiberopticsensorforcontinuousglucosemonitoringfromcaninestudiesonpointaccuracy AT chouyuhsiang acuteresponseinvivoofafiberopticsensorforcontinuousglucosemonitoringfromcaninestudiesonpointaccuracy |