Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy

The objective of this study was to evaluate the acute response of Sencil(™), a fiber-optic sensor, in point accuracy for glucose monitoring in vivo on healthy dogs under anesthesia. A total of four dogs with clinically normal glycemia were implanted with one sensor each in the chest region to measur...

Descripción completa

Detalles Bibliográficos
Autores principales: Liao, Kuo-Chih, Chang, Shih-Chieh, Chiu, Cheng-Yang, Chou, Yu-Hsiang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Molecular Diversity Preservation International (MDPI) 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3231153/
https://www.ncbi.nlm.nih.gov/pubmed/22163627
http://dx.doi.org/10.3390/s100807789
_version_ 1782218156004081664
author Liao, Kuo-Chih
Chang, Shih-Chieh
Chiu, Cheng-Yang
Chou, Yu-Hsiang
author_facet Liao, Kuo-Chih
Chang, Shih-Chieh
Chiu, Cheng-Yang
Chou, Yu-Hsiang
author_sort Liao, Kuo-Chih
collection PubMed
description The objective of this study was to evaluate the acute response of Sencil(™), a fiber-optic sensor, in point accuracy for glucose monitoring in vivo on healthy dogs under anesthesia. A total of four dogs with clinically normal glycemia were implanted with one sensor each in the chest region to measure the interstitial glucose concentration during the ovariohysterectomy procedure. The data was acquired every 10 seconds after initiation, and was compared to the concentration of venous plasma glucose sampled during the surgery procedures for accuracy of agreement analysis. In the four trials with a range of 71–297 mg/dL plasma glucose, the collected 21 pairs of ISF readings from the Sencil™ and the plasma reference showed superior dispersion of residue values than the conventional system, and a linear correlation (the Pearson correlation coefficient is 0.9288 and the y-intercept is 14.22 mg/dL). The MAD (17.6 mg/dL) and RMAD (16.16%) of Sencil™ measurements were in the comparable range of the conventional system. The Clarke error grid analysis indicated that 100% of the paired points were in the clinically acceptable zone A (61.9%) and B (38.1%).
format Online
Article
Text
id pubmed-3231153
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher Molecular Diversity Preservation International (MDPI)
record_format MEDLINE/PubMed
spelling pubmed-32311532011-12-07 Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy Liao, Kuo-Chih Chang, Shih-Chieh Chiu, Cheng-Yang Chou, Yu-Hsiang Sensors (Basel) Article The objective of this study was to evaluate the acute response of Sencil(™), a fiber-optic sensor, in point accuracy for glucose monitoring in vivo on healthy dogs under anesthesia. A total of four dogs with clinically normal glycemia were implanted with one sensor each in the chest region to measure the interstitial glucose concentration during the ovariohysterectomy procedure. The data was acquired every 10 seconds after initiation, and was compared to the concentration of venous plasma glucose sampled during the surgery procedures for accuracy of agreement analysis. In the four trials with a range of 71–297 mg/dL plasma glucose, the collected 21 pairs of ISF readings from the Sencil™ and the plasma reference showed superior dispersion of residue values than the conventional system, and a linear correlation (the Pearson correlation coefficient is 0.9288 and the y-intercept is 14.22 mg/dL). The MAD (17.6 mg/dL) and RMAD (16.16%) of Sencil™ measurements were in the comparable range of the conventional system. The Clarke error grid analysis indicated that 100% of the paired points were in the clinically acceptable zone A (61.9%) and B (38.1%). Molecular Diversity Preservation International (MDPI) 2010-08-20 /pmc/articles/PMC3231153/ /pubmed/22163627 http://dx.doi.org/10.3390/s100807789 Text en © 2010 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/).
spellingShingle Article
Liao, Kuo-Chih
Chang, Shih-Chieh
Chiu, Cheng-Yang
Chou, Yu-Hsiang
Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy
title Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy
title_full Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy
title_fullStr Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy
title_full_unstemmed Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy
title_short Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy
title_sort acute response in vivo of a fiber-optic sensor for continuous glucose monitoring from canine studies on point accuracy
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3231153/
https://www.ncbi.nlm.nih.gov/pubmed/22163627
http://dx.doi.org/10.3390/s100807789
work_keys_str_mv AT liaokuochih acuteresponseinvivoofafiberopticsensorforcontinuousglucosemonitoringfromcaninestudiesonpointaccuracy
AT changshihchieh acuteresponseinvivoofafiberopticsensorforcontinuousglucosemonitoringfromcaninestudiesonpointaccuracy
AT chiuchengyang acuteresponseinvivoofafiberopticsensorforcontinuousglucosemonitoringfromcaninestudiesonpointaccuracy
AT chouyuhsiang acuteresponseinvivoofafiberopticsensorforcontinuousglucosemonitoringfromcaninestudiesonpointaccuracy