Cargando…
The Apparent and the Effective PR Interval, Insights for Cardiac Pacing
A 75-year-old-male patient with dual chamber pacemaker presented with a bizarre EKG showing a unique spike within the QRS complex. Apparent PR interval was 160 ms and effective atrio-right ventricular delay was 210 ms due to right bundle branch block. Sensed AV delay was set at 180 ms causing pseudo...
Autor principal: | |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Libertas Academica
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3235988/ https://www.ncbi.nlm.nih.gov/pubmed/22174592 http://dx.doi.org/10.4137/CCRep.S8227 |
_version_ | 1782218669908033536 |
---|---|
author | Kossaify, Antoine |
author_facet | Kossaify, Antoine |
author_sort | Kossaify, Antoine |
collection | PubMed |
description | A 75-year-old-male patient with dual chamber pacemaker presented with a bizarre EKG showing a unique spike within the QRS complex. Apparent PR interval was 160 ms and effective atrio-right ventricular delay was 210 ms due to right bundle branch block. Sensed AV delay was set at 180 ms causing pseudofusions. Insights regarding cardiac pacing are presented. |
format | Online Article Text |
id | pubmed-3235988 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2011 |
publisher | Libertas Academica |
record_format | MEDLINE/PubMed |
spelling | pubmed-32359882011-12-15 The Apparent and the Effective PR Interval, Insights for Cardiac Pacing Kossaify, Antoine Clin Med Insights Case Rep Case Report A 75-year-old-male patient with dual chamber pacemaker presented with a bizarre EKG showing a unique spike within the QRS complex. Apparent PR interval was 160 ms and effective atrio-right ventricular delay was 210 ms due to right bundle branch block. Sensed AV delay was set at 180 ms causing pseudofusions. Insights regarding cardiac pacing are presented. Libertas Academica 2011-11-14 /pmc/articles/PMC3235988/ /pubmed/22174592 http://dx.doi.org/10.4137/CCRep.S8227 Text en © the author(s), publisher and licensee Libertas Academica Ltd. This is an open access article. Unrestricted non-commercial use is permitted provided the original work is properly cited. |
spellingShingle | Case Report Kossaify, Antoine The Apparent and the Effective PR Interval, Insights for Cardiac Pacing |
title | The Apparent and the Effective PR Interval, Insights for Cardiac Pacing |
title_full | The Apparent and the Effective PR Interval, Insights for Cardiac Pacing |
title_fullStr | The Apparent and the Effective PR Interval, Insights for Cardiac Pacing |
title_full_unstemmed | The Apparent and the Effective PR Interval, Insights for Cardiac Pacing |
title_short | The Apparent and the Effective PR Interval, Insights for Cardiac Pacing |
title_sort | apparent and the effective pr interval, insights for cardiac pacing |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3235988/ https://www.ncbi.nlm.nih.gov/pubmed/22174592 http://dx.doi.org/10.4137/CCRep.S8227 |
work_keys_str_mv | AT kossaifyantoine theapparentandtheeffectiveprintervalinsightsforcardiacpacing AT kossaifyantoine apparentandtheeffectiveprintervalinsightsforcardiacpacing |