Effect of low-density heterogeneities in telecobalt therapy and validation of dose calculation algorithm of a treatment planning system

Telecobalt machines are still prominently used for the treatment of a variety of cancer cases in developing countries. The human body is a heterogeneous composition of variety of tissues and cavities which vary widely in their physical and radiological properties. The presence of heterogeneities in...

Descripción completa

Detalles Bibliográficos
Autores principales: Kumar, Anuj, Sharma, Sunil Dutt, Arya, A. K., Gupta, Surabhi, Shrotriya, Deepak
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3249730/
https://www.ncbi.nlm.nih.gov/pubmed/22228928
http://dx.doi.org/10.4103/0971-6203.89967
Descripción
Sumario:Telecobalt machines are still prominently used for the treatment of a variety of cancer cases in developing countries. The human body is a heterogeneous composition of variety of tissues and cavities which vary widely in their physical and radiological properties. The presence of heterogeneities in the path of telecobalt beam presents an altered dose distribution in the region of clinical interests. A computerized treatment planning system (TPS) is generally used for calculating the dose distribution in the patient. Experimental measurements were carried out in a telecobalt beam with the objectives to study the effects of low-density heterogeneities and to verify the ability of the ASHA radiotherapy TPS in predicting the altered dose distribution along the central axis and off-axis of the beam. Locally available kailwood was tested for its lung equivalence and measurements were carried out in a polymethyl methacrylate phantom by introducing lung equivalent and air gap heterogeneities. A comparison of experimentally measured and TPS calculated dose values indicates that the TPS overestimates the dose by 11.6% in lung equivalent (kailwood) heterogeneity along the central axis. Similarly, it was found that the TPS overestimates the dose by 3.9% and 5.9%, respectively, with air heterogeneity of 1.0 and 2.0 cm. While testing the adequacy of TPS in off-axis region, it was found that the TPS calculation does not indicate the widening of the beam profile in the low-density heterogeneity region. This study suggests that the effective path length based algorithm of the ASHA radiotherapy TPS is unable to achieve the recommended 3% accuracy of clinical dose calculation in heterogeneous media.