The effects of cutouts on output, mean energy and percentage depth dose of 12 and 14 MeV electrons
Electron field-shaping cerrobend cutouts on the linear accelerator applicator have some effects on the output and percentage depth dose. These effects which arise from the lateral scatter nonequilibrium are particularly evident in higher energies and in cutouts with smaller radius. Dose measurements...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3249732/ https://www.ncbi.nlm.nih.gov/pubmed/22228930 http://dx.doi.org/10.4103/0971-6203.89970 |
_version_ | 1782220374281289728 |
---|---|
author | Khaledy, Navid Arbabi, Azim Sardari, Dariush |
author_facet | Khaledy, Navid Arbabi, Azim Sardari, Dariush |
author_sort | Khaledy, Navid |
collection | PubMed |
description | Electron field-shaping cerrobend cutouts on the linear accelerator applicator have some effects on the output and percentage depth dose. These effects which arise from the lateral scatter nonequilibrium are particularly evident in higher energies and in cutouts with smaller radius. Dose measurements for circular, square, and triangular cutouts as well as open field was performed in a 10 × 10 cm applicator, using plane parallel type ion chamber with a 100 cm source surface distance. The Percentage Depth Doses curves were drawn and the outputs were measured for each of these cutouts. The output factors, normalized to open 10 × 10 cm field, varied between 0.891 and 0.996 depending on the energy, cutout shape, and cavity area. With the use of cutouts, R(100) shifted toward the surface. The shifts ranged from 9 to 0 mm and from 13 to 0 mm for 12 and 14 MeV, respectively, depending on the shape and cavity area. For R(90), R(80), and R(50) the ranges for observed shifts narrowed down and practically no shifts were observed for R(20). We present these changes in the form of predictive formulas, which would be useful in clinical applications. |
format | Online Article Text |
id | pubmed-3249732 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2011 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-32497322012-01-06 The effects of cutouts on output, mean energy and percentage depth dose of 12 and 14 MeV electrons Khaledy, Navid Arbabi, Azim Sardari, Dariush J Med Phys Original Article Electron field-shaping cerrobend cutouts on the linear accelerator applicator have some effects on the output and percentage depth dose. These effects which arise from the lateral scatter nonequilibrium are particularly evident in higher energies and in cutouts with smaller radius. Dose measurements for circular, square, and triangular cutouts as well as open field was performed in a 10 × 10 cm applicator, using plane parallel type ion chamber with a 100 cm source surface distance. The Percentage Depth Doses curves were drawn and the outputs were measured for each of these cutouts. The output factors, normalized to open 10 × 10 cm field, varied between 0.891 and 0.996 depending on the energy, cutout shape, and cavity area. With the use of cutouts, R(100) shifted toward the surface. The shifts ranged from 9 to 0 mm and from 13 to 0 mm for 12 and 14 MeV, respectively, depending on the shape and cavity area. For R(90), R(80), and R(50) the ranges for observed shifts narrowed down and practically no shifts were observed for R(20). We present these changes in the form of predictive formulas, which would be useful in clinical applications. Medknow Publications & Media Pvt Ltd 2011 /pmc/articles/PMC3249732/ /pubmed/22228930 http://dx.doi.org/10.4103/0971-6203.89970 Text en Copyright: © Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Khaledy, Navid Arbabi, Azim Sardari, Dariush The effects of cutouts on output, mean energy and percentage depth dose of 12 and 14 MeV electrons |
title | The effects of cutouts on output, mean energy and percentage depth dose of 12 and 14 MeV electrons |
title_full | The effects of cutouts on output, mean energy and percentage depth dose of 12 and 14 MeV electrons |
title_fullStr | The effects of cutouts on output, mean energy and percentage depth dose of 12 and 14 MeV electrons |
title_full_unstemmed | The effects of cutouts on output, mean energy and percentage depth dose of 12 and 14 MeV electrons |
title_short | The effects of cutouts on output, mean energy and percentage depth dose of 12 and 14 MeV electrons |
title_sort | effects of cutouts on output, mean energy and percentage depth dose of 12 and 14 mev electrons |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3249732/ https://www.ncbi.nlm.nih.gov/pubmed/22228930 http://dx.doi.org/10.4103/0971-6203.89970 |
work_keys_str_mv | AT khaledynavid theeffectsofcutoutsonoutputmeanenergyandpercentagedepthdoseof12and14mevelectrons AT arbabiazim theeffectsofcutoutsonoutputmeanenergyandpercentagedepthdoseof12and14mevelectrons AT sardaridariush theeffectsofcutoutsonoutputmeanenergyandpercentagedepthdoseof12and14mevelectrons AT khaledynavid effectsofcutoutsonoutputmeanenergyandpercentagedepthdoseof12and14mevelectrons AT arbabiazim effectsofcutoutsonoutputmeanenergyandpercentagedepthdoseof12and14mevelectrons AT sardaridariush effectsofcutoutsonoutputmeanenergyandpercentagedepthdoseof12and14mevelectrons |