Cargando…
Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents
PURPOSE: To validate the Dinamap ProCare 200 blood pressure (BP) monitor against a mercury sphygmomanometer in children 7 to 18 years old in accordance with the 2010 International Protocol of European Society of Hypertension (ESH-IP2) and the British Hypertension Society (BHS) protocol. METHODS: For...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Pediatric Society
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3254892/ https://www.ncbi.nlm.nih.gov/pubmed/22253643 http://dx.doi.org/10.3345/kjp.2011.54.11.463 |
_version_ | 1782220918899081216 |
---|---|
author | Lee, Chong Guk Park, Hyang Mi Shin, Hye Jung Moon, Jin Soo Hong, Yeong Mi Kim, Nam Soo Ha, Il Soo Chang, Myeong Jin Oh, Kyeong Won |
author_facet | Lee, Chong Guk Park, Hyang Mi Shin, Hye Jung Moon, Jin Soo Hong, Yeong Mi Kim, Nam Soo Ha, Il Soo Chang, Myeong Jin Oh, Kyeong Won |
author_sort | Lee, Chong Guk |
collection | PubMed |
description | PURPOSE: To validate the Dinamap ProCare 200 blood pressure (BP) monitor against a mercury sphygmomanometer in children 7 to 18 years old in accordance with the 2010 International Protocol of European Society of Hypertension (ESH-IP2) and the British Hypertension Society (BHS) protocol. METHODS: Forty-five children were recruited for the study. A validation procedure was performed following the protocol based on the ESH-IP2 and BHS protocols for children and adolescents. Each subject underwent 7 sequential BP measurements alternatively with a mercury sphygmomanometer and the test device by trained nurses. The results were analyzed according to the validation criteria of ESH-IP2. RESULTS: The mean (±SD) difference in the absolute BP values between test device and mercury sphygmomanometer readings was 1.85±1.65 mmHg for systolic BP (SBP) and 4.41±3.53 mmHg for diastolic BP (DBP). These results fulfilled the Association for the Advancement of Medical Instrumentation criterion of a mean±SD below 5±8 mmHg for both SBP and DBP. The percentages of test device-observer mercury sphygmomanometer BP differences within 5, 10, and 15 mmHg were 96%, 100%, and 100% for SBP, and 69%, 92%, and 100% for DBP, respectively, in the part 1 analysis; both SBP and DBP passed the part 1 criteria. In the part 2 analysis, SBP passed the criteria but DBP failed. CONCLUSION: Although the Dinamap ProCare 200 BP monitor failed an adapted ESH-IP2, SBP passed. When comparing BP readings measured by oscillometers and mercury sphygmomanometers, one has to consider the differences between them, particularly in DBP, because DBP can be underestimated. |
format | Online Article Text |
id | pubmed-3254892 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2011 |
publisher | The Korean Pediatric Society |
record_format | MEDLINE/PubMed |
spelling | pubmed-32548922012-01-17 Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents Lee, Chong Guk Park, Hyang Mi Shin, Hye Jung Moon, Jin Soo Hong, Yeong Mi Kim, Nam Soo Ha, Il Soo Chang, Myeong Jin Oh, Kyeong Won Korean J Pediatr Original Article PURPOSE: To validate the Dinamap ProCare 200 blood pressure (BP) monitor against a mercury sphygmomanometer in children 7 to 18 years old in accordance with the 2010 International Protocol of European Society of Hypertension (ESH-IP2) and the British Hypertension Society (BHS) protocol. METHODS: Forty-five children were recruited for the study. A validation procedure was performed following the protocol based on the ESH-IP2 and BHS protocols for children and adolescents. Each subject underwent 7 sequential BP measurements alternatively with a mercury sphygmomanometer and the test device by trained nurses. The results were analyzed according to the validation criteria of ESH-IP2. RESULTS: The mean (±SD) difference in the absolute BP values between test device and mercury sphygmomanometer readings was 1.85±1.65 mmHg for systolic BP (SBP) and 4.41±3.53 mmHg for diastolic BP (DBP). These results fulfilled the Association for the Advancement of Medical Instrumentation criterion of a mean±SD below 5±8 mmHg for both SBP and DBP. The percentages of test device-observer mercury sphygmomanometer BP differences within 5, 10, and 15 mmHg were 96%, 100%, and 100% for SBP, and 69%, 92%, and 100% for DBP, respectively, in the part 1 analysis; both SBP and DBP passed the part 1 criteria. In the part 2 analysis, SBP passed the criteria but DBP failed. CONCLUSION: Although the Dinamap ProCare 200 BP monitor failed an adapted ESH-IP2, SBP passed. When comparing BP readings measured by oscillometers and mercury sphygmomanometers, one has to consider the differences between them, particularly in DBP, because DBP can be underestimated. The Korean Pediatric Society 2011-11 2011-11-30 /pmc/articles/PMC3254892/ /pubmed/22253643 http://dx.doi.org/10.3345/kjp.2011.54.11.463 Text en Copyright © 2011 by The Korean Pediatric Society http://creativecommons.org/licenses/by-nc/3.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Lee, Chong Guk Park, Hyang Mi Shin, Hye Jung Moon, Jin Soo Hong, Yeong Mi Kim, Nam Soo Ha, Il Soo Chang, Myeong Jin Oh, Kyeong Won Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents |
title | Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents |
title_full | Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents |
title_fullStr | Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents |
title_full_unstemmed | Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents |
title_short | Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents |
title_sort | validation study of the dinamap procare 200 upper arm blood pressure monitor in children and adolescents |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3254892/ https://www.ncbi.nlm.nih.gov/pubmed/22253643 http://dx.doi.org/10.3345/kjp.2011.54.11.463 |
work_keys_str_mv | AT leechongguk validationstudyofthedinamapprocare200upperarmbloodpressuremonitorinchildrenandadolescents AT parkhyangmi validationstudyofthedinamapprocare200upperarmbloodpressuremonitorinchildrenandadolescents AT shinhyejung validationstudyofthedinamapprocare200upperarmbloodpressuremonitorinchildrenandadolescents AT moonjinsoo validationstudyofthedinamapprocare200upperarmbloodpressuremonitorinchildrenandadolescents AT hongyeongmi validationstudyofthedinamapprocare200upperarmbloodpressuremonitorinchildrenandadolescents AT kimnamsoo validationstudyofthedinamapprocare200upperarmbloodpressuremonitorinchildrenandadolescents AT hailsoo validationstudyofthedinamapprocare200upperarmbloodpressuremonitorinchildrenandadolescents AT changmyeongjin validationstudyofthedinamapprocare200upperarmbloodpressuremonitorinchildrenandadolescents AT ohkyeongwon validationstudyofthedinamapprocare200upperarmbloodpressuremonitorinchildrenandadolescents |