Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers
Infrared tympanic thermometers (ITTs) are easy to use and have a quick response time. They are widely used for temperature measurement of the human body. The accuracy and uncertainty of measurement is the importance performance indicator for these meters. The performance of two infrared tympanic the...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Molecular Diversity Preservation International (MDPI)
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3257961/ https://www.ncbi.nlm.nih.gov/pubmed/22319287 http://dx.doi.org/10.3390/s100403073 |
_version_ | 1782221218050473984 |
---|---|
author | Chung, Wenbin Chen, Chiachung |
author_facet | Chung, Wenbin Chen, Chiachung |
author_sort | Chung, Wenbin |
collection | PubMed |
description | Infrared tympanic thermometers (ITTs) are easy to use and have a quick response time. They are widely used for temperature measurement of the human body. The accuracy and uncertainty of measurement is the importance performance indicator for these meters. The performance of two infrared tympanic thermometers, Braun THT-3020 and OMRON MC-510, were evaluated in this study. The cell of a temperature calibrator was modified to serve as the standard temperature of the blackbody. The errors of measurement for the two meters were reduced by the calibration equation. The predictive values could meet the requirements of the ASTM standard. The sources of uncertainty include the standard deviations of replication at fixed temperature or the predicted values of calibration equation, reference standard values and resolution. The uncertainty analysis shows that the uncertainty of calibration equation is the main source for combined uncertainty. Ambient temperature did not have the significant effects on the measured performance. The calibration equations could improve the accuracy of ITTs. However, these equations did not improve the uncertainty of ITTs. |
format | Online Article Text |
id | pubmed-3257961 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2010 |
publisher | Molecular Diversity Preservation International (MDPI) |
record_format | MEDLINE/PubMed |
spelling | pubmed-32579612012-02-08 Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers Chung, Wenbin Chen, Chiachung Sensors (Basel) Article Infrared tympanic thermometers (ITTs) are easy to use and have a quick response time. They are widely used for temperature measurement of the human body. The accuracy and uncertainty of measurement is the importance performance indicator for these meters. The performance of two infrared tympanic thermometers, Braun THT-3020 and OMRON MC-510, were evaluated in this study. The cell of a temperature calibrator was modified to serve as the standard temperature of the blackbody. The errors of measurement for the two meters were reduced by the calibration equation. The predictive values could meet the requirements of the ASTM standard. The sources of uncertainty include the standard deviations of replication at fixed temperature or the predicted values of calibration equation, reference standard values and resolution. The uncertainty analysis shows that the uncertainty of calibration equation is the main source for combined uncertainty. Ambient temperature did not have the significant effects on the measured performance. The calibration equations could improve the accuracy of ITTs. However, these equations did not improve the uncertainty of ITTs. Molecular Diversity Preservation International (MDPI) 2010-03-31 /pmc/articles/PMC3257961/ /pubmed/22319287 http://dx.doi.org/10.3390/s100403073 Text en © 2010 by the authors; licensee Molecular Diversity Preservation International, Basel, Switzerland. This article is an open-access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/). |
spellingShingle | Article Chung, Wenbin Chen, Chiachung Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers |
title | Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers |
title_full | Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers |
title_fullStr | Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers |
title_full_unstemmed | Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers |
title_short | Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers |
title_sort | evaluation of performance and uncertainty of infrared tympanic thermometers |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3257961/ https://www.ncbi.nlm.nih.gov/pubmed/22319287 http://dx.doi.org/10.3390/s100403073 |
work_keys_str_mv | AT chungwenbin evaluationofperformanceanduncertaintyofinfraredtympanicthermometers AT chenchiachung evaluationofperformanceanduncertaintyofinfraredtympanicthermometers |