Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers

Infrared tympanic thermometers (ITTs) are easy to use and have a quick response time. They are widely used for temperature measurement of the human body. The accuracy and uncertainty of measurement is the importance performance indicator for these meters. The performance of two infrared tympanic the...

Descripción completa

Detalles Bibliográficos
Autores principales: Chung, Wenbin, Chen, Chiachung
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Molecular Diversity Preservation International (MDPI) 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3257961/
https://www.ncbi.nlm.nih.gov/pubmed/22319287
http://dx.doi.org/10.3390/s100403073
_version_ 1782221218050473984
author Chung, Wenbin
Chen, Chiachung
author_facet Chung, Wenbin
Chen, Chiachung
author_sort Chung, Wenbin
collection PubMed
description Infrared tympanic thermometers (ITTs) are easy to use and have a quick response time. They are widely used for temperature measurement of the human body. The accuracy and uncertainty of measurement is the importance performance indicator for these meters. The performance of two infrared tympanic thermometers, Braun THT-3020 and OMRON MC-510, were evaluated in this study. The cell of a temperature calibrator was modified to serve as the standard temperature of the blackbody. The errors of measurement for the two meters were reduced by the calibration equation. The predictive values could meet the requirements of the ASTM standard. The sources of uncertainty include the standard deviations of replication at fixed temperature or the predicted values of calibration equation, reference standard values and resolution. The uncertainty analysis shows that the uncertainty of calibration equation is the main source for combined uncertainty. Ambient temperature did not have the significant effects on the measured performance. The calibration equations could improve the accuracy of ITTs. However, these equations did not improve the uncertainty of ITTs.
format Online
Article
Text
id pubmed-3257961
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher Molecular Diversity Preservation International (MDPI)
record_format MEDLINE/PubMed
spelling pubmed-32579612012-02-08 Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers Chung, Wenbin Chen, Chiachung Sensors (Basel) Article Infrared tympanic thermometers (ITTs) are easy to use and have a quick response time. They are widely used for temperature measurement of the human body. The accuracy and uncertainty of measurement is the importance performance indicator for these meters. The performance of two infrared tympanic thermometers, Braun THT-3020 and OMRON MC-510, were evaluated in this study. The cell of a temperature calibrator was modified to serve as the standard temperature of the blackbody. The errors of measurement for the two meters were reduced by the calibration equation. The predictive values could meet the requirements of the ASTM standard. The sources of uncertainty include the standard deviations of replication at fixed temperature or the predicted values of calibration equation, reference standard values and resolution. The uncertainty analysis shows that the uncertainty of calibration equation is the main source for combined uncertainty. Ambient temperature did not have the significant effects on the measured performance. The calibration equations could improve the accuracy of ITTs. However, these equations did not improve the uncertainty of ITTs. Molecular Diversity Preservation International (MDPI) 2010-03-31 /pmc/articles/PMC3257961/ /pubmed/22319287 http://dx.doi.org/10.3390/s100403073 Text en © 2010 by the authors; licensee Molecular Diversity Preservation International, Basel, Switzerland. This article is an open-access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/).
spellingShingle Article
Chung, Wenbin
Chen, Chiachung
Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers
title Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers
title_full Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers
title_fullStr Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers
title_full_unstemmed Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers
title_short Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers
title_sort evaluation of performance and uncertainty of infrared tympanic thermometers
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3257961/
https://www.ncbi.nlm.nih.gov/pubmed/22319287
http://dx.doi.org/10.3390/s100403073
work_keys_str_mv AT chungwenbin evaluationofperformanceanduncertaintyofinfraredtympanicthermometers
AT chenchiachung evaluationofperformanceanduncertaintyofinfraredtympanicthermometers