Visual performance of the Tecnis one-piece lens ZCB00

SYNOPSIS: The aspheric, one-piece design of the Tecnis(®) ZCB00 intraocular lens provided higher mesopic contrast sensitivity and relatively constant refraction at different pupil sizes compared to the Sensar(®) AR40e. PURPOSE: Comparing the impact on functional vision of the aspheric Tecnis(®) ZCB0...

Descripción completa

Detalles Bibliográficos
Autores principales: Wahba, Sherein S, Riad, Riad F, Morkos, Fathy F, Hassouna, Abdallah K, Roshdy, Maged M
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3258090/
https://www.ncbi.nlm.nih.gov/pubmed/22267915
http://dx.doi.org/10.2147/OPTH.S27324
Descripción
Sumario:SYNOPSIS: The aspheric, one-piece design of the Tecnis(®) ZCB00 intraocular lens provided higher mesopic contrast sensitivity and relatively constant refraction at different pupil sizes compared to the Sensar(®) AR40e. PURPOSE: Comparing the impact on functional vision of the aspheric Tecnis(®) ZCB00 one-piece lens compared to the spherical Sensar(®) AR40e three-piece lens. SETTING: National Eye Hospital, Cairo, Egypt. METHODS: In this prospective study, we implanted 32 Tecnis ZCB00 (Abbott Medical Optics, Santa Ana, CA) and 30 Sensar AR40e (Pharamacia and Upjohn LLC, North Peapack, NJ) lenses. Twenty-three patients in the ZCB00 group and 13 patients in the AR40e group provided reliable, reproducible data, and were therefore included in the statistical analysis. After 8 weeks, we measured photopic contrast sensitivity (PCS) at 85 cd/m(2) and mesopic contrast sensitivity (MCS) at 5 cd/m(2). Uncorrected and best-corrected visual acuity were measured. Myopic shift was analyzed by comparing the manifest refraction spherical equivalent at 3 days and 8 weeks after surgery in the normal and dilated pupil. RESULTS: The Tecnis ZCB00 group showed significantly better MCS than the Sensar AR40e group at a spatial frequency of six cycles per degree (cpd) (P = 0.037), but not at 3, 12, or 18 cpd (P = 0.299, P = 0.226, and P = 0.396, respectively). There were no significant differences between groups in corrected distance visual acuity (P = 0.175) or PCS at 3, 6, 12, or 18 cpd spatial frequencies (P = 0.440, P = 0.176, P = 0.365, and P = 0.251, respectively). The ZCB00 group showed less myopic shift in normal and dilated pupils between 3 days and 8 weeks after surgery compared to the AR40e group, but this difference was not significant. CONCLUSION: The Tecnis one-piece ZCB00 provided higher MCS at moderate spatial frequency than the three-piece Sensar AR40e, and was associated with relatively constant refraction at different pupil sizes.