High-frequency irreversible electroporation (H-FIRE) for non-thermal ablation without muscle contraction

BACKGROUND: Therapeutic irreversible electroporation (IRE) is an emerging technology for the non-thermal ablation of tumors. The technique involves delivering a series of unipolar electric pulses to permanently destabilize the plasma membrane of cancer cells through an increase in transmembrane pote...

Descripción completa

Detalles Bibliográficos
Autores principales: Arena, Christopher B, Sano, Michael B, Rossmeisl, John H, Caldwell, John L, Garcia, Paulo A, Rylander, Marissa Nichole, Davalos, Rafael V
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3258292/
https://www.ncbi.nlm.nih.gov/pubmed/22104372
http://dx.doi.org/10.1186/1475-925X-10-102
_version_ 1782221266062671872
author Arena, Christopher B
Sano, Michael B
Rossmeisl, John H
Caldwell, John L
Garcia, Paulo A
Rylander, Marissa Nichole
Davalos, Rafael V
author_facet Arena, Christopher B
Sano, Michael B
Rossmeisl, John H
Caldwell, John L
Garcia, Paulo A
Rylander, Marissa Nichole
Davalos, Rafael V
author_sort Arena, Christopher B
collection PubMed
description BACKGROUND: Therapeutic irreversible electroporation (IRE) is an emerging technology for the non-thermal ablation of tumors. The technique involves delivering a series of unipolar electric pulses to permanently destabilize the plasma membrane of cancer cells through an increase in transmembrane potential, which leads to the development of a tissue lesion. Clinically, IRE requires the administration of paralytic agents to prevent muscle contractions during treatment that are associated with the delivery of electric pulses. This study shows that by applying high-frequency, bipolar bursts, muscle contractions can be eliminated during IRE without compromising the non-thermal mechanism of cell death. METHODS: A combination of analytical, numerical, and experimental techniques were performed to investigate high-frequency irreversible electroporation (H-FIRE). A theoretical model for determining transmembrane potential in response to arbitrary electric fields was used to identify optimal burst frequencies and amplitudes for in vivo treatments. A finite element model for predicting thermal damage based on the electric field distribution was used to design non-thermal protocols for in vivo experiments. H-FIRE was applied to the brain of rats, and muscle contractions were quantified via accelerometers placed at the cervicothoracic junction. MRI and histological evaluation was performed post-operatively to assess ablation. RESULTS: No visual or tactile evidence of muscle contraction was seen during H-FIRE at 250 kHz or 500 kHz, while all IRE protocols resulted in detectable muscle contractions at the cervicothoracic junction. H-FIRE produced ablative lesions in brain tissue that were characteristic in cellular morphology of non-thermal IRE treatments. Specifically, there was complete uniformity of tissue death within targeted areas, and a sharp transition zone was present between lesioned and normal brain. CONCLUSIONS: H-FIRE is a feasible technique for non-thermal tissue ablation that eliminates muscle contractions seen in IRE treatments performed with unipolar electric pulses. Therefore, it has the potential to be performed clinically without the administration of paralytic agents.
format Online
Article
Text
id pubmed-3258292
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-32582922012-01-18 High-frequency irreversible electroporation (H-FIRE) for non-thermal ablation without muscle contraction Arena, Christopher B Sano, Michael B Rossmeisl, John H Caldwell, John L Garcia, Paulo A Rylander, Marissa Nichole Davalos, Rafael V Biomed Eng Online Research BACKGROUND: Therapeutic irreversible electroporation (IRE) is an emerging technology for the non-thermal ablation of tumors. The technique involves delivering a series of unipolar electric pulses to permanently destabilize the plasma membrane of cancer cells through an increase in transmembrane potential, which leads to the development of a tissue lesion. Clinically, IRE requires the administration of paralytic agents to prevent muscle contractions during treatment that are associated with the delivery of electric pulses. This study shows that by applying high-frequency, bipolar bursts, muscle contractions can be eliminated during IRE without compromising the non-thermal mechanism of cell death. METHODS: A combination of analytical, numerical, and experimental techniques were performed to investigate high-frequency irreversible electroporation (H-FIRE). A theoretical model for determining transmembrane potential in response to arbitrary electric fields was used to identify optimal burst frequencies and amplitudes for in vivo treatments. A finite element model for predicting thermal damage based on the electric field distribution was used to design non-thermal protocols for in vivo experiments. H-FIRE was applied to the brain of rats, and muscle contractions were quantified via accelerometers placed at the cervicothoracic junction. MRI and histological evaluation was performed post-operatively to assess ablation. RESULTS: No visual or tactile evidence of muscle contraction was seen during H-FIRE at 250 kHz or 500 kHz, while all IRE protocols resulted in detectable muscle contractions at the cervicothoracic junction. H-FIRE produced ablative lesions in brain tissue that were characteristic in cellular morphology of non-thermal IRE treatments. Specifically, there was complete uniformity of tissue death within targeted areas, and a sharp transition zone was present between lesioned and normal brain. CONCLUSIONS: H-FIRE is a feasible technique for non-thermal tissue ablation that eliminates muscle contractions seen in IRE treatments performed with unipolar electric pulses. Therefore, it has the potential to be performed clinically without the administration of paralytic agents. BioMed Central 2011-11-21 /pmc/articles/PMC3258292/ /pubmed/22104372 http://dx.doi.org/10.1186/1475-925X-10-102 Text en Copyright ©2011 Arena et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Arena, Christopher B
Sano, Michael B
Rossmeisl, John H
Caldwell, John L
Garcia, Paulo A
Rylander, Marissa Nichole
Davalos, Rafael V
High-frequency irreversible electroporation (H-FIRE) for non-thermal ablation without muscle contraction
title High-frequency irreversible electroporation (H-FIRE) for non-thermal ablation without muscle contraction
title_full High-frequency irreversible electroporation (H-FIRE) for non-thermal ablation without muscle contraction
title_fullStr High-frequency irreversible electroporation (H-FIRE) for non-thermal ablation without muscle contraction
title_full_unstemmed High-frequency irreversible electroporation (H-FIRE) for non-thermal ablation without muscle contraction
title_short High-frequency irreversible electroporation (H-FIRE) for non-thermal ablation without muscle contraction
title_sort high-frequency irreversible electroporation (h-fire) for non-thermal ablation without muscle contraction
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3258292/
https://www.ncbi.nlm.nih.gov/pubmed/22104372
http://dx.doi.org/10.1186/1475-925X-10-102
work_keys_str_mv AT arenachristopherb highfrequencyirreversibleelectroporationhfirefornonthermalablationwithoutmusclecontraction
AT sanomichaelb highfrequencyirreversibleelectroporationhfirefornonthermalablationwithoutmusclecontraction
AT rossmeisljohnh highfrequencyirreversibleelectroporationhfirefornonthermalablationwithoutmusclecontraction
AT caldwelljohnl highfrequencyirreversibleelectroporationhfirefornonthermalablationwithoutmusclecontraction
AT garciapauloa highfrequencyirreversibleelectroporationhfirefornonthermalablationwithoutmusclecontraction
AT rylandermarissanichole highfrequencyirreversibleelectroporationhfirefornonthermalablationwithoutmusclecontraction
AT davalosrafaelv highfrequencyirreversibleelectroporationhfirefornonthermalablationwithoutmusclecontraction