Evaluation of imaging performance of major image guidance systems
PURPOSE: The imaging characteristics of two popular kV cone-beam CT (CBCT) and two MVCT systems utilised in image-guided radiation therapy (IGRT) were evaluated. MATERIALS AND METHODS: The study was performed on Varian Clinac iX, Elekta Synergy S, Siemens Oncor, and Tomotherapy. A CT phantom (Catpha...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Malaysia
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3265149/ https://www.ncbi.nlm.nih.gov/pubmed/22287985 http://dx.doi.org/10.2349/biij.7.2.e11 |
_version_ | 1782222045740793856 |
---|---|
author | Chan, MF Yang, J Song, Y Burman, C Chan, P Li, S |
author_facet | Chan, MF Yang, J Song, Y Burman, C Chan, P Li, S |
author_sort | Chan, MF |
collection | PubMed |
description | PURPOSE: The imaging characteristics of two popular kV cone-beam CT (CBCT) and two MVCT systems utilised in image-guided radiation therapy (IGRT) were evaluated. MATERIALS AND METHODS: The study was performed on Varian Clinac iX, Elekta Synergy S, Siemens Oncor, and Tomotherapy. A CT phantom (Catphan-504, Phantom Laboratory, Salem, NY) was scanned for measurements of image quality including image noise, uniformity, density accuracy, spatial resolution, contrast linearity, and contrast resolution. The measurement results were analysed using in-house image analysis software. Reproducibility, position correction, and geometric accuracy were also evaluated with markers in a smaller alignment phantom. The performance evaluation compared volumetric image properties from these four systems with those from a conventional diagnostic CT (CCT). RESULTS: It was shown that the linearity of the two kV CBCT was fairly consistent with CCT. The Elekta CBCT with half-circle 27-cm FOV had higher CT numbers than the other three systems. The image noises of the Elekta kV CBCT, Siemens MV CBCT, and Tomotherapy fan-beam CT (FBCT) are about 2–4 times higher than that of the Varian CBCT. The spatial resolutions of two kV CBCTs and two MV CBCTs were 8-11 lp/cm and 3-5 lp/cm, respectively. CONCLUSION: Elekta CBCT provided a faster image reconstruction and low dose per scan for half-circle scanning. Varian CBCT had relatively lower image noise. Tomotherapy FBCT had the best uniformity. |
format | Online Article Text |
id | pubmed-3265149 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2011 |
publisher | Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Malaysia |
record_format | MEDLINE/PubMed |
spelling | pubmed-32651492012-01-27 Evaluation of imaging performance of major image guidance systems Chan, MF Yang, J Song, Y Burman, C Chan, P Li, S Biomed Imaging Interv J Original Article PURPOSE: The imaging characteristics of two popular kV cone-beam CT (CBCT) and two MVCT systems utilised in image-guided radiation therapy (IGRT) were evaluated. MATERIALS AND METHODS: The study was performed on Varian Clinac iX, Elekta Synergy S, Siemens Oncor, and Tomotherapy. A CT phantom (Catphan-504, Phantom Laboratory, Salem, NY) was scanned for measurements of image quality including image noise, uniformity, density accuracy, spatial resolution, contrast linearity, and contrast resolution. The measurement results were analysed using in-house image analysis software. Reproducibility, position correction, and geometric accuracy were also evaluated with markers in a smaller alignment phantom. The performance evaluation compared volumetric image properties from these four systems with those from a conventional diagnostic CT (CCT). RESULTS: It was shown that the linearity of the two kV CBCT was fairly consistent with CCT. The Elekta CBCT with half-circle 27-cm FOV had higher CT numbers than the other three systems. The image noises of the Elekta kV CBCT, Siemens MV CBCT, and Tomotherapy fan-beam CT (FBCT) are about 2–4 times higher than that of the Varian CBCT. The spatial resolutions of two kV CBCTs and two MV CBCTs were 8-11 lp/cm and 3-5 lp/cm, respectively. CONCLUSION: Elekta CBCT provided a faster image reconstruction and low dose per scan for half-circle scanning. Varian CBCT had relatively lower image noise. Tomotherapy FBCT had the best uniformity. Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Malaysia 2011-04-01 /pmc/articles/PMC3265149/ /pubmed/22287985 http://dx.doi.org/10.2349/biij.7.2.e11 Text en © 2011 Biomedical Imaging and Intervention Journal http://creativecommons.org/licenses/by/2.5/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Chan, MF Yang, J Song, Y Burman, C Chan, P Li, S Evaluation of imaging performance of major image guidance systems |
title | Evaluation of imaging performance of major image guidance systems |
title_full | Evaluation of imaging performance of major image guidance systems |
title_fullStr | Evaluation of imaging performance of major image guidance systems |
title_full_unstemmed | Evaluation of imaging performance of major image guidance systems |
title_short | Evaluation of imaging performance of major image guidance systems |
title_sort | evaluation of imaging performance of major image guidance systems |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3265149/ https://www.ncbi.nlm.nih.gov/pubmed/22287985 http://dx.doi.org/10.2349/biij.7.2.e11 |
work_keys_str_mv | AT chanmf evaluationofimagingperformanceofmajorimageguidancesystems AT yangj evaluationofimagingperformanceofmajorimageguidancesystems AT songy evaluationofimagingperformanceofmajorimageguidancesystems AT burmanc evaluationofimagingperformanceofmajorimageguidancesystems AT chanp evaluationofimagingperformanceofmajorimageguidancesystems AT lis evaluationofimagingperformanceofmajorimageguidancesystems |