Integration of data from remote monitoring systems and programmers into the hospital electronic health record system based on international standards

Remote follow-up of implanted ICDs may offer a solution to the problem of overcrowded outpatient clinics. All major device companies have developed a remote follow-up solution. Data obtained from the remote follow-up systems are stored in a central database system, operated and owned by the device c...

Descripción completa

Detalles Bibliográficos
Autores principales: van der Velde, E. T., Foeken, H., Witteman, T. A., van Erven, L., Schalij, M. J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Bohn Stafleu van Loghum 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3265695/
https://www.ncbi.nlm.nih.gov/pubmed/22231151
http://dx.doi.org/10.1007/s12471-011-0234-x
Descripción
Sumario:Remote follow-up of implanted ICDs may offer a solution to the problem of overcrowded outpatient clinics. All major device companies have developed a remote follow-up solution. Data obtained from the remote follow-up systems are stored in a central database system, operated and owned by the device company and accessible for the physician or technician. However, the problem now arises that part of the patient’s clinical information is stored in the local electronic health record (EHR) system in the hospital, while another part is only available in the remote monitoring database. This may potentially result in patient safety issues. Ideally all information should become available in the EHR system. IHE (Integrating the Healthcare Enterprise) is an initiative to improve the way computer systems in healthcare share information. To address the requirement of integrating remote monitoring data in the local EHR, the IHE Implantable Device Cardiac Observation (IDCO) profile has been developed. In our hospital, we have implemented the IHE IDCO profile to import data from the remote databases from two device vendors into the departmental Cardiology Information System. Data are exchanged via an HL7/XML communication protocol, as defined in the IHE IDCO profile.