The objective assessment of cough frequency: accuracy of the LR102 device

BACKGROUND: The measurement of cough frequency is problematic and most often based on subjective assessment. The aim of the study was to assess the accuracy of the automatic identification of cough episodes by LR102, a cough frequency meter based on electromyography and audio sensors. METHODS: Ten a...

Descripción completa

Detalles Bibliográficos
Autores principales: Leconte, Sophie, Liistro, Giuseppe, Lebecque, Patrick, Degryse, Jean-Marie
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3268712/
https://www.ncbi.nlm.nih.gov/pubmed/22132691
http://dx.doi.org/10.1186/1745-9974-7-11
_version_ 1782222405292261376
author Leconte, Sophie
Liistro, Giuseppe
Lebecque, Patrick
Degryse, Jean-Marie
author_facet Leconte, Sophie
Liistro, Giuseppe
Lebecque, Patrick
Degryse, Jean-Marie
author_sort Leconte, Sophie
collection PubMed
description BACKGROUND: The measurement of cough frequency is problematic and most often based on subjective assessment. The aim of the study was to assess the accuracy of the automatic identification of cough episodes by LR102, a cough frequency meter based on electromyography and audio sensors. METHODS: Ten adult patients complaining of cough were recruited in primary care and hospital settings. Participants were asked to wear LR102 for 4 consecutive hours during which they were also filmed. RESULTS: Measures of cough frequency by LR102 and manual counting were closely correlated (r = 0.87 for number of cough episodes per hour; r = 0.89 for number of single coughs per hour) but LR102 overestimated cough frequency. Bland-Altman plots indicate that differences between the two measurements were not influenced by cough frequency. CONCLUSIONS: LR102 offers a useful estimate of cough frequency in adults in their own environment, while significantly reducing the time required for analysis.
format Online
Article
Text
id pubmed-3268712
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-32687122012-01-31 The objective assessment of cough frequency: accuracy of the LR102 device Leconte, Sophie Liistro, Giuseppe Lebecque, Patrick Degryse, Jean-Marie Cough Research BACKGROUND: The measurement of cough frequency is problematic and most often based on subjective assessment. The aim of the study was to assess the accuracy of the automatic identification of cough episodes by LR102, a cough frequency meter based on electromyography and audio sensors. METHODS: Ten adult patients complaining of cough were recruited in primary care and hospital settings. Participants were asked to wear LR102 for 4 consecutive hours during which they were also filmed. RESULTS: Measures of cough frequency by LR102 and manual counting were closely correlated (r = 0.87 for number of cough episodes per hour; r = 0.89 for number of single coughs per hour) but LR102 overestimated cough frequency. Bland-Altman plots indicate that differences between the two measurements were not influenced by cough frequency. CONCLUSIONS: LR102 offers a useful estimate of cough frequency in adults in their own environment, while significantly reducing the time required for analysis. BioMed Central 2011-12-01 /pmc/articles/PMC3268712/ /pubmed/22132691 http://dx.doi.org/10.1186/1745-9974-7-11 Text en Copyright ©2011 Leconte et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Leconte, Sophie
Liistro, Giuseppe
Lebecque, Patrick
Degryse, Jean-Marie
The objective assessment of cough frequency: accuracy of the LR102 device
title The objective assessment of cough frequency: accuracy of the LR102 device
title_full The objective assessment of cough frequency: accuracy of the LR102 device
title_fullStr The objective assessment of cough frequency: accuracy of the LR102 device
title_full_unstemmed The objective assessment of cough frequency: accuracy of the LR102 device
title_short The objective assessment of cough frequency: accuracy of the LR102 device
title_sort objective assessment of cough frequency: accuracy of the lr102 device
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3268712/
https://www.ncbi.nlm.nih.gov/pubmed/22132691
http://dx.doi.org/10.1186/1745-9974-7-11
work_keys_str_mv AT lecontesophie theobjectiveassessmentofcoughfrequencyaccuracyofthelr102device
AT liistrogiuseppe theobjectiveassessmentofcoughfrequencyaccuracyofthelr102device
AT lebecquepatrick theobjectiveassessmentofcoughfrequencyaccuracyofthelr102device
AT degrysejeanmarie theobjectiveassessmentofcoughfrequencyaccuracyofthelr102device
AT lecontesophie objectiveassessmentofcoughfrequencyaccuracyofthelr102device
AT liistrogiuseppe objectiveassessmentofcoughfrequencyaccuracyofthelr102device
AT lebecquepatrick objectiveassessmentofcoughfrequencyaccuracyofthelr102device
AT degrysejeanmarie objectiveassessmentofcoughfrequencyaccuracyofthelr102device