The objective assessment of cough frequency: accuracy of the LR102 device
BACKGROUND: The measurement of cough frequency is problematic and most often based on subjective assessment. The aim of the study was to assess the accuracy of the automatic identification of cough episodes by LR102, a cough frequency meter based on electromyography and audio sensors. METHODS: Ten a...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3268712/ https://www.ncbi.nlm.nih.gov/pubmed/22132691 http://dx.doi.org/10.1186/1745-9974-7-11 |
_version_ | 1782222405292261376 |
---|---|
author | Leconte, Sophie Liistro, Giuseppe Lebecque, Patrick Degryse, Jean-Marie |
author_facet | Leconte, Sophie Liistro, Giuseppe Lebecque, Patrick Degryse, Jean-Marie |
author_sort | Leconte, Sophie |
collection | PubMed |
description | BACKGROUND: The measurement of cough frequency is problematic and most often based on subjective assessment. The aim of the study was to assess the accuracy of the automatic identification of cough episodes by LR102, a cough frequency meter based on electromyography and audio sensors. METHODS: Ten adult patients complaining of cough were recruited in primary care and hospital settings. Participants were asked to wear LR102 for 4 consecutive hours during which they were also filmed. RESULTS: Measures of cough frequency by LR102 and manual counting were closely correlated (r = 0.87 for number of cough episodes per hour; r = 0.89 for number of single coughs per hour) but LR102 overestimated cough frequency. Bland-Altman plots indicate that differences between the two measurements were not influenced by cough frequency. CONCLUSIONS: LR102 offers a useful estimate of cough frequency in adults in their own environment, while significantly reducing the time required for analysis. |
format | Online Article Text |
id | pubmed-3268712 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2011 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-32687122012-01-31 The objective assessment of cough frequency: accuracy of the LR102 device Leconte, Sophie Liistro, Giuseppe Lebecque, Patrick Degryse, Jean-Marie Cough Research BACKGROUND: The measurement of cough frequency is problematic and most often based on subjective assessment. The aim of the study was to assess the accuracy of the automatic identification of cough episodes by LR102, a cough frequency meter based on electromyography and audio sensors. METHODS: Ten adult patients complaining of cough were recruited in primary care and hospital settings. Participants were asked to wear LR102 for 4 consecutive hours during which they were also filmed. RESULTS: Measures of cough frequency by LR102 and manual counting were closely correlated (r = 0.87 for number of cough episodes per hour; r = 0.89 for number of single coughs per hour) but LR102 overestimated cough frequency. Bland-Altman plots indicate that differences between the two measurements were not influenced by cough frequency. CONCLUSIONS: LR102 offers a useful estimate of cough frequency in adults in their own environment, while significantly reducing the time required for analysis. BioMed Central 2011-12-01 /pmc/articles/PMC3268712/ /pubmed/22132691 http://dx.doi.org/10.1186/1745-9974-7-11 Text en Copyright ©2011 Leconte et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Leconte, Sophie Liistro, Giuseppe Lebecque, Patrick Degryse, Jean-Marie The objective assessment of cough frequency: accuracy of the LR102 device |
title | The objective assessment of cough frequency: accuracy of the LR102 device |
title_full | The objective assessment of cough frequency: accuracy of the LR102 device |
title_fullStr | The objective assessment of cough frequency: accuracy of the LR102 device |
title_full_unstemmed | The objective assessment of cough frequency: accuracy of the LR102 device |
title_short | The objective assessment of cough frequency: accuracy of the LR102 device |
title_sort | objective assessment of cough frequency: accuracy of the lr102 device |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3268712/ https://www.ncbi.nlm.nih.gov/pubmed/22132691 http://dx.doi.org/10.1186/1745-9974-7-11 |
work_keys_str_mv | AT lecontesophie theobjectiveassessmentofcoughfrequencyaccuracyofthelr102device AT liistrogiuseppe theobjectiveassessmentofcoughfrequencyaccuracyofthelr102device AT lebecquepatrick theobjectiveassessmentofcoughfrequencyaccuracyofthelr102device AT degrysejeanmarie theobjectiveassessmentofcoughfrequencyaccuracyofthelr102device AT lecontesophie objectiveassessmentofcoughfrequencyaccuracyofthelr102device AT liistrogiuseppe objectiveassessmentofcoughfrequencyaccuracyofthelr102device AT lebecquepatrick objectiveassessmentofcoughfrequencyaccuracyofthelr102device AT degrysejeanmarie objectiveassessmentofcoughfrequencyaccuracyofthelr102device |