Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster

Purpose. To compare anterior chamber depth (ACD), representing the distance between the anterior corneal surface and anterior lens surface measurements between the Galilei Dual Scheimpflug Analyzer and the IOLMaster. Methods. A retrospective review of 65 individual patient eyes with normal anterior...

Descripción completa

Detalles Bibliográficos
Autores principales: Patel, Roma P., Pandit, Rahul T.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3270446/
https://www.ncbi.nlm.nih.gov/pubmed/22315660
http://dx.doi.org/10.1155/2012/430249
_version_ 1782222580933984256
author Patel, Roma P.
Pandit, Rahul T.
author_facet Patel, Roma P.
Pandit, Rahul T.
author_sort Patel, Roma P.
collection PubMed
description Purpose. To compare anterior chamber depth (ACD), representing the distance between the anterior corneal surface and anterior lens surface measurements between the Galilei Dual Scheimpflug Analyzer and the IOLMaster. Methods. A retrospective review of 65 individual patient eyes with normal anterior segments, and no prior ocular surgery was performed. Patients underwent ACD measurements with both devices during the same session by a trained examiner. Interdevice agreement was evaluated using paired two-tailed t-tests, Pearson correlation coefficient, and Bland-Altman analysis. Results. The mean ± standard deviation (SD) ACD for the Galilei and IOLMaster was 3.37 ± 0.36 mm (range from 2.62 to 4.13) and 3.25 ± 0.38 mm (range from 2.34 to 3.92), respectively (Pearson correlation coefficient = 0.96). ACD mean difference was 0.12 mm (P < 0.0001); 95% limits of agreement was from −0.09 to 0.34. The Galilei measured slightly longer ACD values than the IOLMaster. There was no relationship between axial length and interdevice difference. Conclusion. ACD measurements correlate well between the Galilei and IOLMaster, with Galilei values on average 0.12 mm longer than the IOLMaster.
format Online
Article
Text
id pubmed-3270446
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-32704462012-02-07 Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster Patel, Roma P. Pandit, Rahul T. J Ophthalmol Clinical Study Purpose. To compare anterior chamber depth (ACD), representing the distance between the anterior corneal surface and anterior lens surface measurements between the Galilei Dual Scheimpflug Analyzer and the IOLMaster. Methods. A retrospective review of 65 individual patient eyes with normal anterior segments, and no prior ocular surgery was performed. Patients underwent ACD measurements with both devices during the same session by a trained examiner. Interdevice agreement was evaluated using paired two-tailed t-tests, Pearson correlation coefficient, and Bland-Altman analysis. Results. The mean ± standard deviation (SD) ACD for the Galilei and IOLMaster was 3.37 ± 0.36 mm (range from 2.62 to 4.13) and 3.25 ± 0.38 mm (range from 2.34 to 3.92), respectively (Pearson correlation coefficient = 0.96). ACD mean difference was 0.12 mm (P < 0.0001); 95% limits of agreement was from −0.09 to 0.34. The Galilei measured slightly longer ACD values than the IOLMaster. There was no relationship between axial length and interdevice difference. Conclusion. ACD measurements correlate well between the Galilei and IOLMaster, with Galilei values on average 0.12 mm longer than the IOLMaster. Hindawi Publishing Corporation 2012 2012-01-19 /pmc/articles/PMC3270446/ /pubmed/22315660 http://dx.doi.org/10.1155/2012/430249 Text en Copyright © 2012 R. P. Patel and R. T. Pandit. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Clinical Study
Patel, Roma P.
Pandit, Rahul T.
Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster
title Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster
title_full Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster
title_fullStr Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster
title_full_unstemmed Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster
title_short Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster
title_sort comparison of anterior chamber depth measurements from the galilei dual scheimpflug analyzer with iolmaster
topic Clinical Study
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3270446/
https://www.ncbi.nlm.nih.gov/pubmed/22315660
http://dx.doi.org/10.1155/2012/430249
work_keys_str_mv AT patelromap comparisonofanteriorchamberdepthmeasurementsfromthegalileidualscheimpfluganalyzerwithiolmaster
AT panditrahult comparisonofanteriorchamberdepthmeasurementsfromthegalileidualscheimpfluganalyzerwithiolmaster