Cargando…

Non-Linear Signal Detection Improvement by Radiation Damping in Single-Pulse NMR Spectra

When NMR lines overlap and at least one of them is affected by radiation damping, the resonance line shapes of all lines are no longer Lorentzian. We report the appearance of narrow signal distortions, which resemble hole-burnt spectra. This new experimental phenomenon facilitates the detection of t...

Descripción completa

Detalles Bibliográficos
Autores principales: Schlagnitweit, Judith, Morgan, Steven W, Nausner, Martin, Müller, Norbert, Desvaux, Hervé
Formato: Online Artículo Texto
Lenguaje:English
Publicado: WILEY-VCH Verlag 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3307601/
https://www.ncbi.nlm.nih.gov/pubmed/22266720
http://dx.doi.org/10.1002/cphc.201100724
Descripción
Sumario:When NMR lines overlap and at least one of them is affected by radiation damping, the resonance line shapes of all lines are no longer Lorentzian. We report the appearance of narrow signal distortions, which resemble hole-burnt spectra. This new experimental phenomenon facilitates the detection of tiny signals hidden below the main resonance. Theoretical analysis based on modified Maxwell–Bloch equations shows that the presence of strong transverse magnetization creates a feedback through the coil, which influences the magnetization of all spins with overlapping resonance lines. In the time domain this leads to cross-precession terms between magnetization densities, which ultimately cause non-linear behavior. Numerical simulations corroborate this interpretation.