Low Resolution Limits and Inaccurate Algorithms Decrease Significantly the Value of Late Loss in Current Drug-Eluting Stent Trials

Quantitative coronary and vascular angiography (QCA resp., QVA) remains the current gold standard for evaluation of restenosis. Late loss as one of the most commonly accepted parameters to highlight efficacy of the various devices has shown high correlation to clinical parameters but, surprisingly,...

Descripción completa

Detalles Bibliográficos
Autores principales: Dahm, Johannes B., van Buuren, Frank
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3317211/
https://www.ncbi.nlm.nih.gov/pubmed/22489271
http://dx.doi.org/10.1155/2012/417250
_version_ 1782228529561206784
author Dahm, Johannes B.
van Buuren, Frank
author_facet Dahm, Johannes B.
van Buuren, Frank
author_sort Dahm, Johannes B.
collection PubMed
description Quantitative coronary and vascular angiography (QCA resp., QVA) remains the current gold standard for evaluation of restenosis. Late loss as one of the most commonly accepted parameters to highlight efficacy of the various devices has shown high correlation to clinical parameters but, surprisingly, has no impact on the evaluation of the remaining amount of restenostic tissue. The current clinical practice leads to unrealistic late loss calculations. Smaller late loss differences are usually not greater than the inherited resolution limits of QCA, which is especially the case in small differences between the various stents in the drug-eluting stent era. Late loss include additional systematic and random errors, due to the fact that measurements were taken at two different time points including the inherited resolution and calibration limits of QCA on two occasions. Due to the limited value of late loss in discriminating the small differences between the one and other DES, late lumen area loss and clearly defined calculation algorithms (e.g., MLD-relocation) should be used in future DES studies also to fulfill the more stringent regulatory requirements.
format Online
Article
Text
id pubmed-3317211
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-33172112012-04-09 Low Resolution Limits and Inaccurate Algorithms Decrease Significantly the Value of Late Loss in Current Drug-Eluting Stent Trials Dahm, Johannes B. van Buuren, Frank Int J Vasc Med Research Article Quantitative coronary and vascular angiography (QCA resp., QVA) remains the current gold standard for evaluation of restenosis. Late loss as one of the most commonly accepted parameters to highlight efficacy of the various devices has shown high correlation to clinical parameters but, surprisingly, has no impact on the evaluation of the remaining amount of restenostic tissue. The current clinical practice leads to unrealistic late loss calculations. Smaller late loss differences are usually not greater than the inherited resolution limits of QCA, which is especially the case in small differences between the various stents in the drug-eluting stent era. Late loss include additional systematic and random errors, due to the fact that measurements were taken at two different time points including the inherited resolution and calibration limits of QCA on two occasions. Due to the limited value of late loss in discriminating the small differences between the one and other DES, late lumen area loss and clearly defined calculation algorithms (e.g., MLD-relocation) should be used in future DES studies also to fulfill the more stringent regulatory requirements. Hindawi Publishing Corporation 2012 2012-03-20 /pmc/articles/PMC3317211/ /pubmed/22489271 http://dx.doi.org/10.1155/2012/417250 Text en Copyright © 2012 J. B. Dahm and F. van Buuren. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Dahm, Johannes B.
van Buuren, Frank
Low Resolution Limits and Inaccurate Algorithms Decrease Significantly the Value of Late Loss in Current Drug-Eluting Stent Trials
title Low Resolution Limits and Inaccurate Algorithms Decrease Significantly the Value of Late Loss in Current Drug-Eluting Stent Trials
title_full Low Resolution Limits and Inaccurate Algorithms Decrease Significantly the Value of Late Loss in Current Drug-Eluting Stent Trials
title_fullStr Low Resolution Limits and Inaccurate Algorithms Decrease Significantly the Value of Late Loss in Current Drug-Eluting Stent Trials
title_full_unstemmed Low Resolution Limits and Inaccurate Algorithms Decrease Significantly the Value of Late Loss in Current Drug-Eluting Stent Trials
title_short Low Resolution Limits and Inaccurate Algorithms Decrease Significantly the Value of Late Loss in Current Drug-Eluting Stent Trials
title_sort low resolution limits and inaccurate algorithms decrease significantly the value of late loss in current drug-eluting stent trials
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3317211/
https://www.ncbi.nlm.nih.gov/pubmed/22489271
http://dx.doi.org/10.1155/2012/417250
work_keys_str_mv AT dahmjohannesb lowresolutionlimitsandinaccuratealgorithmsdecreasesignificantlythevalueoflatelossincurrentdrugelutingstenttrials
AT vanbuurenfrank lowresolutionlimitsandinaccuratealgorithmsdecreasesignificantlythevalueoflatelossincurrentdrugelutingstenttrials