The Exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (RSA)

BACKGROUND: Due to its collarless, double-tapered polished design, the Exeter femoral stem is known to migrate distally in the first 5 years after implantation. However, its long-term migration pattern has not been investigated. PATIENTS AND METHODS: 39 consecutive patients (41 total hip arthroplast...

Descripción completa

Detalles Bibliográficos
Autores principales: Nieuwenhuijse, Marc J, Valstar, Edward R, Kaptein, Bart L, Nelissen, Rob G H H
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Informa Healthcare 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3339525/
https://www.ncbi.nlm.nih.gov/pubmed/22401676
http://dx.doi.org/10.3109/17453674.2012.672093
Descripción
Sumario:BACKGROUND: Due to its collarless, double-tapered polished design, the Exeter femoral stem is known to migrate distally in the first 5 years after implantation. However, its long-term migration pattern has not been investigated. PATIENTS AND METHODS: 39 consecutive patients (41 total hip arthroplasties) received a cemented Exeter stem and had prospective clinical and RSA follow-up. Patients were evaluated postoperatively at 6, 12, 26, and 52 weeks, and annually thereafter. Short-term results have been reported. In this study, the mean length of follow-up was 9.4 years (SD 3.2 years). No patients were lost to follow-up. 15 patients died during follow-up. RESULTS: No stems were revised. In 4 stems, fractures of the cement mantle were noted within the first 3 postoperative years. In 3 stems, this resulted in a complete circumferential cement mantle discontinuity. For the 37 well-performing stems, continuous but small migration was measured between 2 and 12 years of follow-up. Continued subsidence of 0.08 mm/year (95% CI: 0.05–0.12, p < 0.001) was seen in combination with continued rotation in retroversion of 0.07°/year (95% CI: 0.02–0.12, p = 0.01). At 10 years of follow-up, mean subsidence was 2.1 (SD 1.2) mm and mean retroversion was 1.8° (SD 2.0). Two-thirds of this occurred during the first 2 postoperative years. In the 3 stems with a complete circumferential cement fracture, a sudden and disproportionately high increase in subsidence was measured in the time period of occurrence. INTERPRETATION: The Exeter femoral stem continues to migrate during the first decade after implantation. Absolute stability is not required for good long-term survival if this is compatible with the design of the implant.