Cargando…
The Exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (RSA)
BACKGROUND: Due to its collarless, double-tapered polished design, the Exeter femoral stem is known to migrate distally in the first 5 years after implantation. However, its long-term migration pattern has not been investigated. PATIENTS AND METHODS: 39 consecutive patients (41 total hip arthroplast...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Informa Healthcare
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3339525/ https://www.ncbi.nlm.nih.gov/pubmed/22401676 http://dx.doi.org/10.3109/17453674.2012.672093 |
_version_ | 1782231370611818496 |
---|---|
author | Nieuwenhuijse, Marc J Valstar, Edward R Kaptein, Bart L Nelissen, Rob G H H |
author_facet | Nieuwenhuijse, Marc J Valstar, Edward R Kaptein, Bart L Nelissen, Rob G H H |
author_sort | Nieuwenhuijse, Marc J |
collection | PubMed |
description | BACKGROUND: Due to its collarless, double-tapered polished design, the Exeter femoral stem is known to migrate distally in the first 5 years after implantation. However, its long-term migration pattern has not been investigated. PATIENTS AND METHODS: 39 consecutive patients (41 total hip arthroplasties) received a cemented Exeter stem and had prospective clinical and RSA follow-up. Patients were evaluated postoperatively at 6, 12, 26, and 52 weeks, and annually thereafter. Short-term results have been reported. In this study, the mean length of follow-up was 9.4 years (SD 3.2 years). No patients were lost to follow-up. 15 patients died during follow-up. RESULTS: No stems were revised. In 4 stems, fractures of the cement mantle were noted within the first 3 postoperative years. In 3 stems, this resulted in a complete circumferential cement mantle discontinuity. For the 37 well-performing stems, continuous but small migration was measured between 2 and 12 years of follow-up. Continued subsidence of 0.08 mm/year (95% CI: 0.05–0.12, p < 0.001) was seen in combination with continued rotation in retroversion of 0.07°/year (95% CI: 0.02–0.12, p = 0.01). At 10 years of follow-up, mean subsidence was 2.1 (SD 1.2) mm and mean retroversion was 1.8° (SD 2.0). Two-thirds of this occurred during the first 2 postoperative years. In the 3 stems with a complete circumferential cement fracture, a sudden and disproportionately high increase in subsidence was measured in the time period of occurrence. INTERPRETATION: The Exeter femoral stem continues to migrate during the first decade after implantation. Absolute stability is not required for good long-term survival if this is compatible with the design of the implant. |
format | Online Article Text |
id | pubmed-3339525 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | Informa Healthcare |
record_format | MEDLINE/PubMed |
spelling | pubmed-33395252012-05-03 The Exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (RSA) Nieuwenhuijse, Marc J Valstar, Edward R Kaptein, Bart L Nelissen, Rob G H H Acta Orthop Original Papers BACKGROUND: Due to its collarless, double-tapered polished design, the Exeter femoral stem is known to migrate distally in the first 5 years after implantation. However, its long-term migration pattern has not been investigated. PATIENTS AND METHODS: 39 consecutive patients (41 total hip arthroplasties) received a cemented Exeter stem and had prospective clinical and RSA follow-up. Patients were evaluated postoperatively at 6, 12, 26, and 52 weeks, and annually thereafter. Short-term results have been reported. In this study, the mean length of follow-up was 9.4 years (SD 3.2 years). No patients were lost to follow-up. 15 patients died during follow-up. RESULTS: No stems were revised. In 4 stems, fractures of the cement mantle were noted within the first 3 postoperative years. In 3 stems, this resulted in a complete circumferential cement mantle discontinuity. For the 37 well-performing stems, continuous but small migration was measured between 2 and 12 years of follow-up. Continued subsidence of 0.08 mm/year (95% CI: 0.05–0.12, p < 0.001) was seen in combination with continued rotation in retroversion of 0.07°/year (95% CI: 0.02–0.12, p = 0.01). At 10 years of follow-up, mean subsidence was 2.1 (SD 1.2) mm and mean retroversion was 1.8° (SD 2.0). Two-thirds of this occurred during the first 2 postoperative years. In the 3 stems with a complete circumferential cement fracture, a sudden and disproportionately high increase in subsidence was measured in the time period of occurrence. INTERPRETATION: The Exeter femoral stem continues to migrate during the first decade after implantation. Absolute stability is not required for good long-term survival if this is compatible with the design of the implant. Informa Healthcare 2012-04 2012-04-24 /pmc/articles/PMC3339525/ /pubmed/22401676 http://dx.doi.org/10.3109/17453674.2012.672093 Text en Copyright: © Nordic Orthopaedic Federation http://creativecommons.org/licenses/by/2.5/ This is an open-access article distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the source is credited. |
spellingShingle | Original Papers Nieuwenhuijse, Marc J Valstar, Edward R Kaptein, Bart L Nelissen, Rob G H H The Exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (RSA) |
title | The Exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (RSA) |
title_full | The Exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (RSA) |
title_fullStr | The Exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (RSA) |
title_full_unstemmed | The Exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (RSA) |
title_short | The Exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (RSA) |
title_sort | exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (rsa) |
topic | Original Papers |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3339525/ https://www.ncbi.nlm.nih.gov/pubmed/22401676 http://dx.doi.org/10.3109/17453674.2012.672093 |
work_keys_str_mv | AT nieuwenhuijsemarcj theexeterfemoralstemcontinuestomigrateduringitsfirstdecadeafterimplantation1012yearsoffollowupwithradiostereometricanalysisrsa AT valstaredwardr theexeterfemoralstemcontinuestomigrateduringitsfirstdecadeafterimplantation1012yearsoffollowupwithradiostereometricanalysisrsa AT kapteinbartl theexeterfemoralstemcontinuestomigrateduringitsfirstdecadeafterimplantation1012yearsoffollowupwithradiostereometricanalysisrsa AT nelissenrobghh theexeterfemoralstemcontinuestomigrateduringitsfirstdecadeafterimplantation1012yearsoffollowupwithradiostereometricanalysisrsa AT nieuwenhuijsemarcj exeterfemoralstemcontinuestomigrateduringitsfirstdecadeafterimplantation1012yearsoffollowupwithradiostereometricanalysisrsa AT valstaredwardr exeterfemoralstemcontinuestomigrateduringitsfirstdecadeafterimplantation1012yearsoffollowupwithradiostereometricanalysisrsa AT kapteinbartl exeterfemoralstemcontinuestomigrateduringitsfirstdecadeafterimplantation1012yearsoffollowupwithradiostereometricanalysisrsa AT nelissenrobghh exeterfemoralstemcontinuestomigrateduringitsfirstdecadeafterimplantation1012yearsoffollowupwithradiostereometricanalysisrsa |