RSA prediction of high failure rate for the uncoated Interax TKA confirmed by meta-analysis
BACKGROUND AND PURPOSE: In a previous radiostereometric (RSA) trial the uncoated, uncemented, Interax tibial components showed excessive migration within 2 years compared to HA-coated and cemented tibial components. It was predicted that this type of fixation would have a high failure rate. The purp...
Autores principales: | Pijls, Bart G, Nieuwenhuijse, Marc J, Schoones, Jan W, Middeldorp, Saskia, Valstar, Edward R, Nelissen, Rob G H H |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Informa Healthcare
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3339527/ https://www.ncbi.nlm.nih.gov/pubmed/22530953 http://dx.doi.org/10.3109/17453674.2012.672092 |
Ejemplares similares
-
Early proximal migration of cups is associated with late revision in THA: A systematic review and meta-analysis of 26 RSA studies and 49 survival studies
por: Pijls, Bart G, et al.
Publicado: (2012) -
Early subsidence of shape-closed hip arthroplasty stems is associated with late revision: A systematic review and meta-analysis of 24 RSA studies and 56 survival studies
por: van der Voort, Paul, et al.
Publicado: (2015) -
5-year clinical and radiostereometric analysis (RSA) follow-up of 39 CUT femoral neck total hip prostheses in young osteoarthritis patients
por: Nieuwenhuijse, Marc J, et al.
Publicado: (2012) -
The Exeter femoral stem continues to migrate during its first decade after implantation: 10–12 years of follow-up with radiostereometric analysis (RSA)
por: Nieuwenhuijse, Marc J, et al.
Publicado: (2012) -
The beneficial effect of hydroxyapatite lasts: A randomized radiostereometric trial comparing hydroxyapatite-coated, uncoated, and cemented tibial components for up to 16 years
por: Pijls, Bart G, et al.
Publicado: (2012)