Cargando…

Complete bacterial filter blockage by a plastic film

We describe a case where the expiratory bacterial filter was completely blocked by a plastic membrane. Significantly increased airway pressures and inability to deliver adequate tidal volume to the patient were experienced. The cause of machine failure could not be identified until a complete preane...

Descripción completa

Detalles Bibliográficos
Autores principales: Huang, Jiapeng, Dodwani, Kishin, Simon, Paul, Zhou, Jing
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3339731/
https://www.ncbi.nlm.nih.gov/pubmed/22557749
http://dx.doi.org/10.4103/0970-9185.94902
Descripción
Sumario:We describe a case where the expiratory bacterial filter was completely blocked by a plastic membrane. Significantly increased airway pressures and inability to deliver adequate tidal volume to the patient were experienced. The cause of machine failure could not be identified until a complete preanesthesia machine manual check was performed. All machines and circuits should be checked manually, even with the availability of new and advanced self-tests in machines. Backup ventilation equipments and familiarity with signs of expiratory and inspiratory limb obstruction should not be underestimated.