Complete bacterial filter blockage by a plastic film

We describe a case where the expiratory bacterial filter was completely blocked by a plastic membrane. Significantly increased airway pressures and inability to deliver adequate tidal volume to the patient were experienced. The cause of machine failure could not be identified until a complete preane...

Descripción completa

Detalles Bibliográficos
Autores principales: Huang, Jiapeng, Dodwani, Kishin, Simon, Paul, Zhou, Jing
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3339731/
https://www.ncbi.nlm.nih.gov/pubmed/22557749
http://dx.doi.org/10.4103/0970-9185.94902
_version_ 1782231406928199680
author Huang, Jiapeng
Dodwani, Kishin
Simon, Paul
Zhou, Jing
author_facet Huang, Jiapeng
Dodwani, Kishin
Simon, Paul
Zhou, Jing
author_sort Huang, Jiapeng
collection PubMed
description We describe a case where the expiratory bacterial filter was completely blocked by a plastic membrane. Significantly increased airway pressures and inability to deliver adequate tidal volume to the patient were experienced. The cause of machine failure could not be identified until a complete preanesthesia machine manual check was performed. All machines and circuits should be checked manually, even with the availability of new and advanced self-tests in machines. Backup ventilation equipments and familiarity with signs of expiratory and inspiratory limb obstruction should not be underestimated.
format Online
Article
Text
id pubmed-3339731
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-33397312012-05-03 Complete bacterial filter blockage by a plastic film Huang, Jiapeng Dodwani, Kishin Simon, Paul Zhou, Jing J Anaesthesiol Clin Pharmacol Case Report We describe a case where the expiratory bacterial filter was completely blocked by a plastic membrane. Significantly increased airway pressures and inability to deliver adequate tidal volume to the patient were experienced. The cause of machine failure could not be identified until a complete preanesthesia machine manual check was performed. All machines and circuits should be checked manually, even with the availability of new and advanced self-tests in machines. Backup ventilation equipments and familiarity with signs of expiratory and inspiratory limb obstruction should not be underestimated. Medknow Publications & Media Pvt Ltd 2012 /pmc/articles/PMC3339731/ /pubmed/22557749 http://dx.doi.org/10.4103/0970-9185.94902 Text en Copyright: © Journal of Anaesthesiology Clinical Pharmacology http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Case Report
Huang, Jiapeng
Dodwani, Kishin
Simon, Paul
Zhou, Jing
Complete bacterial filter blockage by a plastic film
title Complete bacterial filter blockage by a plastic film
title_full Complete bacterial filter blockage by a plastic film
title_fullStr Complete bacterial filter blockage by a plastic film
title_full_unstemmed Complete bacterial filter blockage by a plastic film
title_short Complete bacterial filter blockage by a plastic film
title_sort complete bacterial filter blockage by a plastic film
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3339731/
https://www.ncbi.nlm.nih.gov/pubmed/22557749
http://dx.doi.org/10.4103/0970-9185.94902
work_keys_str_mv AT huangjiapeng completebacterialfilterblockagebyaplasticfilm
AT dodwanikishin completebacterialfilterblockagebyaplasticfilm
AT simonpaul completebacterialfilterblockagebyaplasticfilm
AT zhoujing completebacterialfilterblockagebyaplasticfilm