Comparison of HemoCue(® )hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan

BACKGROUND: Assessment of hemoglobin is one of the most reliable indicators for anemia, and is widely used to screen for anemia among pregnant women. The HemoCue(® )has been widely used for as a point-of-care device for hemoglobin estimation in health facilities. Previous studies showed contradictor...

Descripción completa

Detalles Bibliográficos
Autores principales: Adam, Ishag, Ahmed, Samah, Mahmoud, Mahmoud H, Yassin, Mohammed I
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3342090/
https://www.ncbi.nlm.nih.gov/pubmed/22436620
http://dx.doi.org/10.1186/1746-1596-7-30
_version_ 1782231627741528064
author Adam, Ishag
Ahmed, Samah
Mahmoud, Mahmoud H
Yassin, Mohammed I
author_facet Adam, Ishag
Ahmed, Samah
Mahmoud, Mahmoud H
Yassin, Mohammed I
author_sort Adam, Ishag
collection PubMed
description BACKGROUND: Assessment of hemoglobin is one of the most reliable indicators for anemia, and is widely used to screen for anemia among pregnant women. The HemoCue(® )has been widely used for as a point-of-care device for hemoglobin estimation in health facilities. Previous studies showed contradictory results regarding the accuracy of HemoCue(®). METHODS: This was a hospital-based cross sectional study carried- out among pregnant women at Khartoum hospital in Sudan to find out whether the measurement of hemoglobin concentration by HemoCue(® )using venous or capillary samples was comparable to that of the automated hematology analyzer as standard. Bland and Altman method was used to compare the measurements with an acceptable difference of ± 1.0 g/dl. RESULTS: Among the 108 subjects in this study the mean (SD) level of hemoglobin level using HemoCue(® )venous sample, HemoCue(® )capillary sample and automated hematology analyzer were 12.70 (1.77), 12.87 (2.04) and 11.53 (1.63) g/dl, respectively. Although the correlations between the measurements were all significant there was no agreement between HemoCue(® )and automated hematology analyzer. The bias + SD (limits of agreement) for HemoCue(® )venous versus hematology analyzer was 1.17 ± 1.57 (-1.97, 4.31) g/dl, HemoCue(® )capillary versus hematology analyzer was 1.34 ± 1.85 (-2.36, 5.04) g/dl, and HemoCue(® )venous versus HemoCue(® )capillary samples was 017 ± 1.90 and (3.97-3.63) g/dl. CONCLUSION: Hemoglobin concentration assessment by HemoCue(® )using either venous or capillary blood samples has shown unacceptable agreement with automated hematology analyzer. VIRTUAL SLIDES: The virtual slide(s) for this article can be found here: http://www.diagnosticpathology.diagnomx.eu/vs/8797022296725036
format Online
Article
Text
id pubmed-3342090
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-33420902012-05-03 Comparison of HemoCue(® )hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan Adam, Ishag Ahmed, Samah Mahmoud, Mahmoud H Yassin, Mohammed I Diagn Pathol Research BACKGROUND: Assessment of hemoglobin is one of the most reliable indicators for anemia, and is widely used to screen for anemia among pregnant women. The HemoCue(® )has been widely used for as a point-of-care device for hemoglobin estimation in health facilities. Previous studies showed contradictory results regarding the accuracy of HemoCue(®). METHODS: This was a hospital-based cross sectional study carried- out among pregnant women at Khartoum hospital in Sudan to find out whether the measurement of hemoglobin concentration by HemoCue(® )using venous or capillary samples was comparable to that of the automated hematology analyzer as standard. Bland and Altman method was used to compare the measurements with an acceptable difference of ± 1.0 g/dl. RESULTS: Among the 108 subjects in this study the mean (SD) level of hemoglobin level using HemoCue(® )venous sample, HemoCue(® )capillary sample and automated hematology analyzer were 12.70 (1.77), 12.87 (2.04) and 11.53 (1.63) g/dl, respectively. Although the correlations between the measurements were all significant there was no agreement between HemoCue(® )and automated hematology analyzer. The bias + SD (limits of agreement) for HemoCue(® )venous versus hematology analyzer was 1.17 ± 1.57 (-1.97, 4.31) g/dl, HemoCue(® )capillary versus hematology analyzer was 1.34 ± 1.85 (-2.36, 5.04) g/dl, and HemoCue(® )venous versus HemoCue(® )capillary samples was 017 ± 1.90 and (3.97-3.63) g/dl. CONCLUSION: Hemoglobin concentration assessment by HemoCue(® )using either venous or capillary blood samples has shown unacceptable agreement with automated hematology analyzer. VIRTUAL SLIDES: The virtual slide(s) for this article can be found here: http://www.diagnosticpathology.diagnomx.eu/vs/8797022296725036 BioMed Central 2012-03-21 /pmc/articles/PMC3342090/ /pubmed/22436620 http://dx.doi.org/10.1186/1746-1596-7-30 Text en Copyright ©2012 Adam et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Adam, Ishag
Ahmed, Samah
Mahmoud, Mahmoud H
Yassin, Mohammed I
Comparison of HemoCue(® )hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan
title Comparison of HemoCue(® )hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan
title_full Comparison of HemoCue(® )hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan
title_fullStr Comparison of HemoCue(® )hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan
title_full_unstemmed Comparison of HemoCue(® )hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan
title_short Comparison of HemoCue(® )hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan
title_sort comparison of hemocue(® )hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at khartoum hospital, sudan
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3342090/
https://www.ncbi.nlm.nih.gov/pubmed/22436620
http://dx.doi.org/10.1186/1746-1596-7-30
work_keys_str_mv AT adamishag comparisonofhemocuehemoglobinmeterandautomatedhematologyanalyzerinmeasurementofhemoglobinlevelsinpregnantwomenatkhartoumhospitalsudan
AT ahmedsamah comparisonofhemocuehemoglobinmeterandautomatedhematologyanalyzerinmeasurementofhemoglobinlevelsinpregnantwomenatkhartoumhospitalsudan
AT mahmoudmahmoudh comparisonofhemocuehemoglobinmeterandautomatedhematologyanalyzerinmeasurementofhemoglobinlevelsinpregnantwomenatkhartoumhospitalsudan
AT yassinmohammedi comparisonofhemocuehemoglobinmeterandautomatedhematologyanalyzerinmeasurementofhemoglobinlevelsinpregnantwomenatkhartoumhospitalsudan