Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement
PURPOSE: To evaluate reliability of Fourier-domain optical coherence tomography (OCT) (RTVue), time-domain OCT (Visante), and slit-lamp adapted ultrasonic pachymetry (SL-US) in the measurement of central corneal thickness (CCT). MATERIALS AND METHODS: Thirty healthy volunteers visited our clinic 3 t...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Yonsei University College of Medicine
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3343442/ https://www.ncbi.nlm.nih.gov/pubmed/22477010 http://dx.doi.org/10.3349/ymj.2012.53.3.634 |
_version_ | 1782231825291149312 |
---|---|
author | Hong, Jin Pyo Nam, Sang Min Kim, Tae-Im Seo, Kyoung Yul Lee, Sang Yeop Meduri, Alessandro Kim, Eung Kweon |
author_facet | Hong, Jin Pyo Nam, Sang Min Kim, Tae-Im Seo, Kyoung Yul Lee, Sang Yeop Meduri, Alessandro Kim, Eung Kweon |
author_sort | Hong, Jin Pyo |
collection | PubMed |
description | PURPOSE: To evaluate reliability of Fourier-domain optical coherence tomography (OCT) (RTVue), time-domain OCT (Visante), and slit-lamp adapted ultrasonic pachymetry (SL-US) in the measurement of central corneal thickness (CCT). MATERIALS AND METHODS: Thirty healthy volunteers visited our clinic 3 times and fifty eyes were measured by one physician. RTVue and Visante were randomly performed, and then SL-US, in which the ultrasound probe was inserted into the Goldmann tonometry mount, was done. During the second visit, each measurement was repeated 3 times. Measurements on the second visit were averaged, and agreement among the instruments was investigated with Bland-Altman plots. RESULTS: RTVue showed smaller repeatability coefficient than Visante and SL-US (4.7, 8.3, and 7.7 µm, respectively). Intersession reproducibility of RTVue and Visante was worse than their repeatability. CCT of RTVue was estimated to be maximally different by 11.8 µm from CCT of Visante and 8.8 µm from CCT of SL-US. The repeatability coefficient of SL-US was 7.7 µm and its reproducibility was similar to the repeatability. CONCLUSION: CCT measured by RTVue showed good reliability and generally agreed with Visante and SL-US. SL-US was as reliable as triple touching conventional US even with a single touch on the cornea. |
format | Online Article Text |
id | pubmed-3343442 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | Yonsei University College of Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-33434422012-05-15 Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement Hong, Jin Pyo Nam, Sang Min Kim, Tae-Im Seo, Kyoung Yul Lee, Sang Yeop Meduri, Alessandro Kim, Eung Kweon Yonsei Med J Original Article PURPOSE: To evaluate reliability of Fourier-domain optical coherence tomography (OCT) (RTVue), time-domain OCT (Visante), and slit-lamp adapted ultrasonic pachymetry (SL-US) in the measurement of central corneal thickness (CCT). MATERIALS AND METHODS: Thirty healthy volunteers visited our clinic 3 times and fifty eyes were measured by one physician. RTVue and Visante were randomly performed, and then SL-US, in which the ultrasound probe was inserted into the Goldmann tonometry mount, was done. During the second visit, each measurement was repeated 3 times. Measurements on the second visit were averaged, and agreement among the instruments was investigated with Bland-Altman plots. RESULTS: RTVue showed smaller repeatability coefficient than Visante and SL-US (4.7, 8.3, and 7.7 µm, respectively). Intersession reproducibility of RTVue and Visante was worse than their repeatability. CCT of RTVue was estimated to be maximally different by 11.8 µm from CCT of Visante and 8.8 µm from CCT of SL-US. The repeatability coefficient of SL-US was 7.7 µm and its reproducibility was similar to the repeatability. CONCLUSION: CCT measured by RTVue showed good reliability and generally agreed with Visante and SL-US. SL-US was as reliable as triple touching conventional US even with a single touch on the cornea. Yonsei University College of Medicine 2012-05-01 2012-03-28 /pmc/articles/PMC3343442/ /pubmed/22477010 http://dx.doi.org/10.3349/ymj.2012.53.3.634 Text en © Copyright: Yonsei University College of Medicine 2012 http://creativecommons.org/licenses/by-nc/3.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Hong, Jin Pyo Nam, Sang Min Kim, Tae-Im Seo, Kyoung Yul Lee, Sang Yeop Meduri, Alessandro Kim, Eung Kweon Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement |
title | Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement |
title_full | Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement |
title_fullStr | Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement |
title_full_unstemmed | Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement |
title_short | Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement |
title_sort | reliability of rtvue, visante, and slit-lamp adapted ultrasonic pachymetry for central corneal thickness measurement |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3343442/ https://www.ncbi.nlm.nih.gov/pubmed/22477010 http://dx.doi.org/10.3349/ymj.2012.53.3.634 |
work_keys_str_mv | AT hongjinpyo reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement AT namsangmin reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement AT kimtaeim reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement AT seokyoungyul reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement AT leesangyeop reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement AT medurialessandro reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement AT kimeungkweon reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement |