Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement

PURPOSE: To evaluate reliability of Fourier-domain optical coherence tomography (OCT) (RTVue), time-domain OCT (Visante), and slit-lamp adapted ultrasonic pachymetry (SL-US) in the measurement of central corneal thickness (CCT). MATERIALS AND METHODS: Thirty healthy volunteers visited our clinic 3 t...

Descripción completa

Detalles Bibliográficos
Autores principales: Hong, Jin Pyo, Nam, Sang Min, Kim, Tae-Im, Seo, Kyoung Yul, Lee, Sang Yeop, Meduri, Alessandro, Kim, Eung Kweon
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Yonsei University College of Medicine 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3343442/
https://www.ncbi.nlm.nih.gov/pubmed/22477010
http://dx.doi.org/10.3349/ymj.2012.53.3.634
_version_ 1782231825291149312
author Hong, Jin Pyo
Nam, Sang Min
Kim, Tae-Im
Seo, Kyoung Yul
Lee, Sang Yeop
Meduri, Alessandro
Kim, Eung Kweon
author_facet Hong, Jin Pyo
Nam, Sang Min
Kim, Tae-Im
Seo, Kyoung Yul
Lee, Sang Yeop
Meduri, Alessandro
Kim, Eung Kweon
author_sort Hong, Jin Pyo
collection PubMed
description PURPOSE: To evaluate reliability of Fourier-domain optical coherence tomography (OCT) (RTVue), time-domain OCT (Visante), and slit-lamp adapted ultrasonic pachymetry (SL-US) in the measurement of central corneal thickness (CCT). MATERIALS AND METHODS: Thirty healthy volunteers visited our clinic 3 times and fifty eyes were measured by one physician. RTVue and Visante were randomly performed, and then SL-US, in which the ultrasound probe was inserted into the Goldmann tonometry mount, was done. During the second visit, each measurement was repeated 3 times. Measurements on the second visit were averaged, and agreement among the instruments was investigated with Bland-Altman plots. RESULTS: RTVue showed smaller repeatability coefficient than Visante and SL-US (4.7, 8.3, and 7.7 µm, respectively). Intersession reproducibility of RTVue and Visante was worse than their repeatability. CCT of RTVue was estimated to be maximally different by 11.8 µm from CCT of Visante and 8.8 µm from CCT of SL-US. The repeatability coefficient of SL-US was 7.7 µm and its reproducibility was similar to the repeatability. CONCLUSION: CCT measured by RTVue showed good reliability and generally agreed with Visante and SL-US. SL-US was as reliable as triple touching conventional US even with a single touch on the cornea.
format Online
Article
Text
id pubmed-3343442
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher Yonsei University College of Medicine
record_format MEDLINE/PubMed
spelling pubmed-33434422012-05-15 Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement Hong, Jin Pyo Nam, Sang Min Kim, Tae-Im Seo, Kyoung Yul Lee, Sang Yeop Meduri, Alessandro Kim, Eung Kweon Yonsei Med J Original Article PURPOSE: To evaluate reliability of Fourier-domain optical coherence tomography (OCT) (RTVue), time-domain OCT (Visante), and slit-lamp adapted ultrasonic pachymetry (SL-US) in the measurement of central corneal thickness (CCT). MATERIALS AND METHODS: Thirty healthy volunteers visited our clinic 3 times and fifty eyes were measured by one physician. RTVue and Visante were randomly performed, and then SL-US, in which the ultrasound probe was inserted into the Goldmann tonometry mount, was done. During the second visit, each measurement was repeated 3 times. Measurements on the second visit were averaged, and agreement among the instruments was investigated with Bland-Altman plots. RESULTS: RTVue showed smaller repeatability coefficient than Visante and SL-US (4.7, 8.3, and 7.7 µm, respectively). Intersession reproducibility of RTVue and Visante was worse than their repeatability. CCT of RTVue was estimated to be maximally different by 11.8 µm from CCT of Visante and 8.8 µm from CCT of SL-US. The repeatability coefficient of SL-US was 7.7 µm and its reproducibility was similar to the repeatability. CONCLUSION: CCT measured by RTVue showed good reliability and generally agreed with Visante and SL-US. SL-US was as reliable as triple touching conventional US even with a single touch on the cornea. Yonsei University College of Medicine 2012-05-01 2012-03-28 /pmc/articles/PMC3343442/ /pubmed/22477010 http://dx.doi.org/10.3349/ymj.2012.53.3.634 Text en © Copyright: Yonsei University College of Medicine 2012 http://creativecommons.org/licenses/by-nc/3.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Hong, Jin Pyo
Nam, Sang Min
Kim, Tae-Im
Seo, Kyoung Yul
Lee, Sang Yeop
Meduri, Alessandro
Kim, Eung Kweon
Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement
title Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement
title_full Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement
title_fullStr Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement
title_full_unstemmed Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement
title_short Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement
title_sort reliability of rtvue, visante, and slit-lamp adapted ultrasonic pachymetry for central corneal thickness measurement
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3343442/
https://www.ncbi.nlm.nih.gov/pubmed/22477010
http://dx.doi.org/10.3349/ymj.2012.53.3.634
work_keys_str_mv AT hongjinpyo reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement
AT namsangmin reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement
AT kimtaeim reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement
AT seokyoungyul reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement
AT leesangyeop reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement
AT medurialessandro reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement
AT kimeungkweon reliabilityofrtvuevisanteandslitlampadaptedultrasonicpachymetryforcentralcornealthicknessmeasurement