Prospective ECG triggering reduces prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT
OBJECTIVES: Multidetector computed tomography (MDCT) has diagnostic value for the evaluation of prosthetic heart valve (PHV) dysfunction but it is hampered by artefacts. We hypothesised that image acquisition using prospective triggering instead of retrospective gating would reduce artefacts related...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer-Verlag
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3348472/ https://www.ncbi.nlm.nih.gov/pubmed/22205446 http://dx.doi.org/10.1007/s00330-011-2358-1 |
_version_ | 1782232389235245056 |
---|---|
author | Symersky, Petr Habets, Jesse Westers, Paul de Mol, Bas A. J. M. Prokop, Mathias Budde, Ricardo P. J. |
author_facet | Symersky, Petr Habets, Jesse Westers, Paul de Mol, Bas A. J. M. Prokop, Mathias Budde, Ricardo P. J. |
author_sort | Symersky, Petr |
collection | PubMed |
description | OBJECTIVES: Multidetector computed tomography (MDCT) has diagnostic value for the evaluation of prosthetic heart valve (PHV) dysfunction but it is hampered by artefacts. We hypothesised that image acquisition using prospective triggering instead of retrospective gating would reduce artefacts related to pulsating PHV. METHODS: In a pulsatile in vitro model, a mono- and bileaflet PHV were imaged using 256 MDCT at 60, 75 and 90 beats per minute (BPM) with either retrospective gating (120 kV, 600 mAs, pitch 0.2, CTDI(vol) 39.8 mGy) or prospective triggering (120 kV, 200 mAs, CTDI(vol) 13.3 mGy). Two thresholds (>175 and <−45HU), derived from the density of surrounding structures, were used for quantification of hyper- and hypodense artefacts. Image noise and artefacts were compared between protocols. RESULTS: Prospective triggering reduced hyperdense artefacts for both valves at every BPM (P = 0.001 all comparisons). Hypodense artefacts were reduced for the monoleaflet valve at 60 (P = 0.009), 75 (P = 0.016) and 90 BPM (P = 0.001), and for the bileaflet valves at 60 (P = 0.001), 90 (P = 0.001) but not at 75 BPM (P = 0.6). Prospective triggering reduced image noise at 60 (P = 0.001) and 75 (P < 0.03) but not at 90 BPM. CONCLUSIONS: Compared with retrospective gating, prospective triggering reduced most artefacts related to pulsating PHV in vitro. KEY POINTS: • Computed tomographic images are often degraded by prosthetic heart valve-induced artefacts • Prospective triggering reduces prosthetic heart valve-induced artefacts in vitro • Artefact reduction at 90 beats per minute occurs without image noise reduction • Prospective triggering may improve CT image quality of moving hyperdense structures |
format | Online Article Text |
id | pubmed-3348472 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2011 |
publisher | Springer-Verlag |
record_format | MEDLINE/PubMed |
spelling | pubmed-33484722012-05-30 Prospective ECG triggering reduces prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT Symersky, Petr Habets, Jesse Westers, Paul de Mol, Bas A. J. M. Prokop, Mathias Budde, Ricardo P. J. Eur Radiol Cardiac OBJECTIVES: Multidetector computed tomography (MDCT) has diagnostic value for the evaluation of prosthetic heart valve (PHV) dysfunction but it is hampered by artefacts. We hypothesised that image acquisition using prospective triggering instead of retrospective gating would reduce artefacts related to pulsating PHV. METHODS: In a pulsatile in vitro model, a mono- and bileaflet PHV were imaged using 256 MDCT at 60, 75 and 90 beats per minute (BPM) with either retrospective gating (120 kV, 600 mAs, pitch 0.2, CTDI(vol) 39.8 mGy) or prospective triggering (120 kV, 200 mAs, CTDI(vol) 13.3 mGy). Two thresholds (>175 and <−45HU), derived from the density of surrounding structures, were used for quantification of hyper- and hypodense artefacts. Image noise and artefacts were compared between protocols. RESULTS: Prospective triggering reduced hyperdense artefacts for both valves at every BPM (P = 0.001 all comparisons). Hypodense artefacts were reduced for the monoleaflet valve at 60 (P = 0.009), 75 (P = 0.016) and 90 BPM (P = 0.001), and for the bileaflet valves at 60 (P = 0.001), 90 (P = 0.001) but not at 75 BPM (P = 0.6). Prospective triggering reduced image noise at 60 (P = 0.001) and 75 (P < 0.03) but not at 90 BPM. CONCLUSIONS: Compared with retrospective gating, prospective triggering reduced most artefacts related to pulsating PHV in vitro. KEY POINTS: • Computed tomographic images are often degraded by prosthetic heart valve-induced artefacts • Prospective triggering reduces prosthetic heart valve-induced artefacts in vitro • Artefact reduction at 90 beats per minute occurs without image noise reduction • Prospective triggering may improve CT image quality of moving hyperdense structures Springer-Verlag 2011-12-30 2012 /pmc/articles/PMC3348472/ /pubmed/22205446 http://dx.doi.org/10.1007/s00330-011-2358-1 Text en © The Author(s) 2011 https://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited. |
spellingShingle | Cardiac Symersky, Petr Habets, Jesse Westers, Paul de Mol, Bas A. J. M. Prokop, Mathias Budde, Ricardo P. J. Prospective ECG triggering reduces prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT |
title | Prospective ECG triggering reduces prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT |
title_full | Prospective ECG triggering reduces prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT |
title_fullStr | Prospective ECG triggering reduces prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT |
title_full_unstemmed | Prospective ECG triggering reduces prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT |
title_short | Prospective ECG triggering reduces prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT |
title_sort | prospective ecg triggering reduces prosthetic heart valve-induced artefacts compared with retrospective ecg gating on 256-slice ct |
topic | Cardiac |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3348472/ https://www.ncbi.nlm.nih.gov/pubmed/22205446 http://dx.doi.org/10.1007/s00330-011-2358-1 |
work_keys_str_mv | AT symerskypetr prospectiveecgtriggeringreducesprostheticheartvalveinducedartefactscomparedwithretrospectiveecggatingon256slicect AT habetsjesse prospectiveecgtriggeringreducesprostheticheartvalveinducedartefactscomparedwithretrospectiveecggatingon256slicect AT westerspaul prospectiveecgtriggeringreducesprostheticheartvalveinducedartefactscomparedwithretrospectiveecggatingon256slicect AT demolbasajm prospectiveecgtriggeringreducesprostheticheartvalveinducedartefactscomparedwithretrospectiveecggatingon256slicect AT prokopmathias prospectiveecgtriggeringreducesprostheticheartvalveinducedartefactscomparedwithretrospectiveecggatingon256slicect AT buddericardopj prospectiveecgtriggeringreducesprostheticheartvalveinducedartefactscomparedwithretrospectiveecggatingon256slicect |