Modeling Focused Ultrasound Exposure for the Optimal Control of Thermal Dose Distribution

Preclinical studies indicate that focused ultrasound at exposure conditions close to the threshold for thermal damage can increase drug delivery at the focal region. Although these results are promising, the optimal control of temperature still remains a challenge. To address this issue, computer-si...

Descripción completa

Detalles Bibliográficos
Autores principales: Sassaroli, E., Li, K. C. P., O'Neill, B. E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Scientific World Journal 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3349131/
https://www.ncbi.nlm.nih.gov/pubmed/22593669
http://dx.doi.org/10.1100/2012/252741
_version_ 1782232471488692224
author Sassaroli, E.
Li, K. C. P.
O'Neill, B. E.
author_facet Sassaroli, E.
Li, K. C. P.
O'Neill, B. E.
author_sort Sassaroli, E.
collection PubMed
description Preclinical studies indicate that focused ultrasound at exposure conditions close to the threshold for thermal damage can increase drug delivery at the focal region. Although these results are promising, the optimal control of temperature still remains a challenge. To address this issue, computer-simulated ultrasound treatments have been performed. When the treatments are delivered without taking into account the cooling effect exerted by the blood flow, the resulting thermal dose is highly variable with regions of thermal damage, regions of underdosage close to the vessels, and areas in between these two extremes. When the power deposition is adjusted so that the peak thermal dose remains close to the threshold for thermal damage, the thermal dose is more uniformly distributed but under-dosage is still visible around the thermally significant vessels. The results of these simulations suggest that, for focused ultrasound, as for other delivery methods, the only way to control temperature is to adjust the average energy deposition to compensate for the presence of thermally significant vessels in the target area. By doing this, we have shown that it is possible to reduce the temperature heterogeneity observed in focused ultrasound thermal treatments.
format Online
Article
Text
id pubmed-3349131
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher The Scientific World Journal
record_format MEDLINE/PubMed
spelling pubmed-33491312012-05-16 Modeling Focused Ultrasound Exposure for the Optimal Control of Thermal Dose Distribution Sassaroli, E. Li, K. C. P. O'Neill, B. E. ScientificWorldJournal Research Article Preclinical studies indicate that focused ultrasound at exposure conditions close to the threshold for thermal damage can increase drug delivery at the focal region. Although these results are promising, the optimal control of temperature still remains a challenge. To address this issue, computer-simulated ultrasound treatments have been performed. When the treatments are delivered without taking into account the cooling effect exerted by the blood flow, the resulting thermal dose is highly variable with regions of thermal damage, regions of underdosage close to the vessels, and areas in between these two extremes. When the power deposition is adjusted so that the peak thermal dose remains close to the threshold for thermal damage, the thermal dose is more uniformly distributed but under-dosage is still visible around the thermally significant vessels. The results of these simulations suggest that, for focused ultrasound, as for other delivery methods, the only way to control temperature is to adjust the average energy deposition to compensate for the presence of thermally significant vessels in the target area. By doing this, we have shown that it is possible to reduce the temperature heterogeneity observed in focused ultrasound thermal treatments. The Scientific World Journal 2012-04-19 /pmc/articles/PMC3349131/ /pubmed/22593669 http://dx.doi.org/10.1100/2012/252741 Text en Copyright © 2012 E. Sassaroli et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Sassaroli, E.
Li, K. C. P.
O'Neill, B. E.
Modeling Focused Ultrasound Exposure for the Optimal Control of Thermal Dose Distribution
title Modeling Focused Ultrasound Exposure for the Optimal Control of Thermal Dose Distribution
title_full Modeling Focused Ultrasound Exposure for the Optimal Control of Thermal Dose Distribution
title_fullStr Modeling Focused Ultrasound Exposure for the Optimal Control of Thermal Dose Distribution
title_full_unstemmed Modeling Focused Ultrasound Exposure for the Optimal Control of Thermal Dose Distribution
title_short Modeling Focused Ultrasound Exposure for the Optimal Control of Thermal Dose Distribution
title_sort modeling focused ultrasound exposure for the optimal control of thermal dose distribution
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3349131/
https://www.ncbi.nlm.nih.gov/pubmed/22593669
http://dx.doi.org/10.1100/2012/252741
work_keys_str_mv AT sassarolie modelingfocusedultrasoundexposurefortheoptimalcontrolofthermaldosedistribution
AT likcp modelingfocusedultrasoundexposurefortheoptimalcontrolofthermaldosedistribution
AT oneillbe modelingfocusedultrasoundexposurefortheoptimalcontrolofthermaldosedistribution