Cardiovascular magnetic resonance evaluation of aortic stenosis severity using single plane measurement of effective orifice area

BACKGROUND: Transthoracic echocardiography (TTE) is the standard method for the evaluation of the severity of aortic stenosis (AS). Valve effective orifice area (EOA) measured by the continuity equation is one of the most frequently used stenotic indices. However, TTE measurement of aortic valve EOA...

Descripción completa

Detalles Bibliográficos
Autores principales: Garcia, Julio, Marrufo, Oscar R, Rodriguez, Alfredo O, Larose, Eric, Pibarot, Philippe, Kadem, Lyes
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3366866/
https://www.ncbi.nlm.nih.gov/pubmed/22480269
http://dx.doi.org/10.1186/1532-429X-14-23
_version_ 1782234775273078784
author Garcia, Julio
Marrufo, Oscar R
Rodriguez, Alfredo O
Larose, Eric
Pibarot, Philippe
Kadem, Lyes
author_facet Garcia, Julio
Marrufo, Oscar R
Rodriguez, Alfredo O
Larose, Eric
Pibarot, Philippe
Kadem, Lyes
author_sort Garcia, Julio
collection PubMed
description BACKGROUND: Transthoracic echocardiography (TTE) is the standard method for the evaluation of the severity of aortic stenosis (AS). Valve effective orifice area (EOA) measured by the continuity equation is one of the most frequently used stenotic indices. However, TTE measurement of aortic valve EOA is not feasible or not reliable in a significant proportion of patients. Cardiovascular magnetic resonance (CMR) has emerged as a non-invasive alternative to evaluate EOA using velocity measurements. The objectives of this study were: 1) to validate a new CMR method using jet shear layer detection (JSLD) based on acoustical source term (AST) concept to estimate the valve EOA; 2) to introduce a simplified JSLD method not requiring vorticity field derivation. METHODS AND RESULTS: We performed an in vitro study where EOA was measured by CMR in 4 fixed stenoses (EOA = 0.48, 1.00, 1.38 and 2.11 cm(2)) under the same steady flow conditions (4-20 L/min). The in vivo study included eight (8) healthy subjects and 37 patients with mild to severe AS (0.72 cm(2 )≤ EOA ≤ 1.71 cm(2)). All subjects underwent TTE and CMR examinations. EOA was determinated by TTE with the use of continuity equation method (TTE(CONT)). For CMR estimation of EOA, we used 3 methods: 1) Continuity equation (CMR(CONT)); 2) Shear layer detection (CMR(JSLD)), which was computed from the velocity field of a single CMR velocity profile at the peak systolic phase; 3) Single plane velocity truncation (CMR(SPVT)), which is a simplified version of CMR(JSLD )method. There was a good agreement between the EOAs obtained in vitro by the different CMR methods and the EOA predicted from the potential flow theory. In the in vivo study, there was good correlation and concordance between the EOA measured by the TTE(CONT )method versus those measured by each of the CMR methods: CMR(CONT )(r = 0.88), CMR(JSLD )(r = 0.93) and CMR(SPVT )(r = 0.93). The intra- and inter- observer variability of EOA measurements was 5 ± 5% and 9 ± 5% for TTE(CONT), 2 ± 1% and 7 ± 5% for CMR(CONT), 7 ± 5% and 8 ± 7% for CMR(JSLD), 1 ± 2% and 3 ± 2% for CMR(SPVT). When repeating image acquisition, reproducibility of measurements was 10 ± 8% and 12 ± 5% for TTE(CONT), 9 ± 9% and 8 ± 8% for CMR(CONT), 6 ± 5% and 7 ± 4% for CMR(JSLD )and 3 ± 2% and 2 ± 2% for CMR(SPVT). CONCLUSION: There was an excellent agreement between the EOA estimated by the CMR(JSLD )or CMR(SPVT )methods and: 1) the theoretical EOA in vitro, and 2) the TTE(CONT )EOA in vivo. The CMR(SPVT )method was superior to the TTE and other CMR methods in terms of measurement variability. The novel CMR-based methods proposed in this study may be helpful to corroborate stenosis severity in patients for whom Doppler-echocardiography exam is inconclusive.
format Online
Article
Text
id pubmed-3366866
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-33668662012-06-05 Cardiovascular magnetic resonance evaluation of aortic stenosis severity using single plane measurement of effective orifice area Garcia, Julio Marrufo, Oscar R Rodriguez, Alfredo O Larose, Eric Pibarot, Philippe Kadem, Lyes J Cardiovasc Magn Reson Research BACKGROUND: Transthoracic echocardiography (TTE) is the standard method for the evaluation of the severity of aortic stenosis (AS). Valve effective orifice area (EOA) measured by the continuity equation is one of the most frequently used stenotic indices. However, TTE measurement of aortic valve EOA is not feasible or not reliable in a significant proportion of patients. Cardiovascular magnetic resonance (CMR) has emerged as a non-invasive alternative to evaluate EOA using velocity measurements. The objectives of this study were: 1) to validate a new CMR method using jet shear layer detection (JSLD) based on acoustical source term (AST) concept to estimate the valve EOA; 2) to introduce a simplified JSLD method not requiring vorticity field derivation. METHODS AND RESULTS: We performed an in vitro study where EOA was measured by CMR in 4 fixed stenoses (EOA = 0.48, 1.00, 1.38 and 2.11 cm(2)) under the same steady flow conditions (4-20 L/min). The in vivo study included eight (8) healthy subjects and 37 patients with mild to severe AS (0.72 cm(2 )≤ EOA ≤ 1.71 cm(2)). All subjects underwent TTE and CMR examinations. EOA was determinated by TTE with the use of continuity equation method (TTE(CONT)). For CMR estimation of EOA, we used 3 methods: 1) Continuity equation (CMR(CONT)); 2) Shear layer detection (CMR(JSLD)), which was computed from the velocity field of a single CMR velocity profile at the peak systolic phase; 3) Single plane velocity truncation (CMR(SPVT)), which is a simplified version of CMR(JSLD )method. There was a good agreement between the EOAs obtained in vitro by the different CMR methods and the EOA predicted from the potential flow theory. In the in vivo study, there was good correlation and concordance between the EOA measured by the TTE(CONT )method versus those measured by each of the CMR methods: CMR(CONT )(r = 0.88), CMR(JSLD )(r = 0.93) and CMR(SPVT )(r = 0.93). The intra- and inter- observer variability of EOA measurements was 5 ± 5% and 9 ± 5% for TTE(CONT), 2 ± 1% and 7 ± 5% for CMR(CONT), 7 ± 5% and 8 ± 7% for CMR(JSLD), 1 ± 2% and 3 ± 2% for CMR(SPVT). When repeating image acquisition, reproducibility of measurements was 10 ± 8% and 12 ± 5% for TTE(CONT), 9 ± 9% and 8 ± 8% for CMR(CONT), 6 ± 5% and 7 ± 4% for CMR(JSLD )and 3 ± 2% and 2 ± 2% for CMR(SPVT). CONCLUSION: There was an excellent agreement between the EOA estimated by the CMR(JSLD )or CMR(SPVT )methods and: 1) the theoretical EOA in vitro, and 2) the TTE(CONT )EOA in vivo. The CMR(SPVT )method was superior to the TTE and other CMR methods in terms of measurement variability. The novel CMR-based methods proposed in this study may be helpful to corroborate stenosis severity in patients for whom Doppler-echocardiography exam is inconclusive. BioMed Central 2012-04-06 /pmc/articles/PMC3366866/ /pubmed/22480269 http://dx.doi.org/10.1186/1532-429X-14-23 Text en Copyright ©2012 Garcia et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Garcia, Julio
Marrufo, Oscar R
Rodriguez, Alfredo O
Larose, Eric
Pibarot, Philippe
Kadem, Lyes
Cardiovascular magnetic resonance evaluation of aortic stenosis severity using single plane measurement of effective orifice area
title Cardiovascular magnetic resonance evaluation of aortic stenosis severity using single plane measurement of effective orifice area
title_full Cardiovascular magnetic resonance evaluation of aortic stenosis severity using single plane measurement of effective orifice area
title_fullStr Cardiovascular magnetic resonance evaluation of aortic stenosis severity using single plane measurement of effective orifice area
title_full_unstemmed Cardiovascular magnetic resonance evaluation of aortic stenosis severity using single plane measurement of effective orifice area
title_short Cardiovascular magnetic resonance evaluation of aortic stenosis severity using single plane measurement of effective orifice area
title_sort cardiovascular magnetic resonance evaluation of aortic stenosis severity using single plane measurement of effective orifice area
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3366866/
https://www.ncbi.nlm.nih.gov/pubmed/22480269
http://dx.doi.org/10.1186/1532-429X-14-23
work_keys_str_mv AT garciajulio cardiovascularmagneticresonanceevaluationofaorticstenosisseverityusingsingleplanemeasurementofeffectiveorificearea
AT marrufooscarr cardiovascularmagneticresonanceevaluationofaorticstenosisseverityusingsingleplanemeasurementofeffectiveorificearea
AT rodriguezalfredoo cardiovascularmagneticresonanceevaluationofaorticstenosisseverityusingsingleplanemeasurementofeffectiveorificearea
AT laroseeric cardiovascularmagneticresonanceevaluationofaorticstenosisseverityusingsingleplanemeasurementofeffectiveorificearea
AT pibarotphilippe cardiovascularmagneticresonanceevaluationofaorticstenosisseverityusingsingleplanemeasurementofeffectiveorificearea
AT kademlyes cardiovascularmagneticresonanceevaluationofaorticstenosisseverityusingsingleplanemeasurementofeffectiveorificearea