In-vivo validation of a new non-invasive continuous ventricular stroke volume monitoring system in an animal model

INTRODUCTION: Recently, a non-invasive, continuous ventricular stroke volume monitoring system using skin electrodes has been developed. In contrast to impedance-based methods, the new technique (ventricular field recognition) enables measurement of changes in ventricular volume. A prototype using t...

Descripción completa

Detalles Bibliográficos
Autores principales: Konings, Maurits K, Grundeman, Paul F, Goovaerts, Henk G, Roosendaal, Maarten R, Hoefer, Imo E, Doevendans, Pieter A, Rademakers, Frank E, Buhre, Wolfgang F
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3387602/
https://www.ncbi.nlm.nih.gov/pubmed/21745380
http://dx.doi.org/10.1186/cc10306
Descripción
Sumario:INTRODUCTION: Recently, a non-invasive, continuous ventricular stroke volume monitoring system using skin electrodes has been developed. In contrast to impedance-based methods, the new technique (ventricular field recognition) enables measurement of changes in ventricular volume. A prototype using this new method was built (the hemologic cardiac profiler, HCP) and validated against a reference method in a pig model during variations in cardiac output. METHODS: In six Dalland pigs, cardiac output was simultaneously measured with the HCP (CO-HCP), and an invasive ultrasonic flow-probe around the ascending aorta (CO-FP). Variations in CO were achieved by change in ventricular loading conditions, cardiac pacing, and dobutamine administration. Data were analysed according to Bland-Altman analysis and Pearson's correlation. RESULTS: Pearson's correlation between the CO-HCP and the CO-FP was r = 0.978. Bland-Altman analysis showed a bias of - 0.114 L/minute, and a variability of the bias (2 standard deviations, 2SD) of 0.55 L/minute. CONCLUSIONS: The results of the present study demonstrate that CO-HCP is comparable to CO-FP in an animal model of cardiac output measurements during a wide variation of CO. Therefore, the HCP has the potential to become a clinical applicable cardiac output monitor.