Comparison of the Tendon Damage Caused by Four Different Anchor Systems Used in Transtendon Rotator Cuff Repair

Objectives. The objective of this study was to compare the damage to the rotator cuff tendons caused by four different anchor systems. Methods. 20 cadaveric human shoulder joints were used for transtendon insertion of four anchor systems. The Healix Peek, Fastin RC, Bio-Corkscrew Suture, and Healix...

Descripción completa

Detalles Bibliográficos
Autores principales: Zhang, Qing-Song, Liu, Sen, Zhang, Qiuyang, Xue, Yun, Ge, Dongxia, O'Brien, Michael J., Savoie, Felix H., You, Zongbing
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3395215/
https://www.ncbi.nlm.nih.gov/pubmed/22811923
http://dx.doi.org/10.1155/2012/798521
_version_ 1782237953884422144
author Zhang, Qing-Song
Liu, Sen
Zhang, Qiuyang
Xue, Yun
Ge, Dongxia
O'Brien, Michael J.
Savoie, Felix H.
You, Zongbing
author_facet Zhang, Qing-Song
Liu, Sen
Zhang, Qiuyang
Xue, Yun
Ge, Dongxia
O'Brien, Michael J.
Savoie, Felix H.
You, Zongbing
author_sort Zhang, Qing-Song
collection PubMed
description Objectives. The objective of this study was to compare the damage to the rotator cuff tendons caused by four different anchor systems. Methods. 20 cadaveric human shoulder joints were used for transtendon insertion of four anchor systems. The Healix Peek, Fastin RC, Bio-Corkscrew Suture, and Healix Transtend anchors were inserted through the tendons using standard transtendon procedures. The areas of tendon damage were measured. Results. The areas of tendon damage (mean ± standard deviation, n = 7) were 29.1 ± 4.3 mm(2) for the Healix Peek anchor, 20.4 ± 2.3 mm(2) for the Fastin RC anchor, 23.4 ± 1.2 mm(2) for the Bio-Corkscrew Suture anchor, 13.7 ± 3.2 mm(2) for the Healix Transtend anchor inserted directly, and 9.1 ± 2.1 mm(2) for the Healix Transtend anchor inserted through the Percannula system (P < 0.001 or P < 0.001, compared to other anchors). Conclusions. In a cadaver transtendon rotator cuff repair model, smaller anchors caused less damage to the tendon tissues. The Healix Transtend implant system caused the least damage to the tendon tissues. Our findings suggest that smaller anchors should be considered when performing transtendon procedures to repair partial rotator cuff tears.
format Online
Article
Text
id pubmed-3395215
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-33952152012-07-18 Comparison of the Tendon Damage Caused by Four Different Anchor Systems Used in Transtendon Rotator Cuff Repair Zhang, Qing-Song Liu, Sen Zhang, Qiuyang Xue, Yun Ge, Dongxia O'Brien, Michael J. Savoie, Felix H. You, Zongbing Adv Orthop Research Article Objectives. The objective of this study was to compare the damage to the rotator cuff tendons caused by four different anchor systems. Methods. 20 cadaveric human shoulder joints were used for transtendon insertion of four anchor systems. The Healix Peek, Fastin RC, Bio-Corkscrew Suture, and Healix Transtend anchors were inserted through the tendons using standard transtendon procedures. The areas of tendon damage were measured. Results. The areas of tendon damage (mean ± standard deviation, n = 7) were 29.1 ± 4.3 mm(2) for the Healix Peek anchor, 20.4 ± 2.3 mm(2) for the Fastin RC anchor, 23.4 ± 1.2 mm(2) for the Bio-Corkscrew Suture anchor, 13.7 ± 3.2 mm(2) for the Healix Transtend anchor inserted directly, and 9.1 ± 2.1 mm(2) for the Healix Transtend anchor inserted through the Percannula system (P < 0.001 or P < 0.001, compared to other anchors). Conclusions. In a cadaver transtendon rotator cuff repair model, smaller anchors caused less damage to the tendon tissues. The Healix Transtend implant system caused the least damage to the tendon tissues. Our findings suggest that smaller anchors should be considered when performing transtendon procedures to repair partial rotator cuff tears. Hindawi Publishing Corporation 2012 2012-07-01 /pmc/articles/PMC3395215/ /pubmed/22811923 http://dx.doi.org/10.1155/2012/798521 Text en Copyright © 2012 Qing-Song Zhang et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Zhang, Qing-Song
Liu, Sen
Zhang, Qiuyang
Xue, Yun
Ge, Dongxia
O'Brien, Michael J.
Savoie, Felix H.
You, Zongbing
Comparison of the Tendon Damage Caused by Four Different Anchor Systems Used in Transtendon Rotator Cuff Repair
title Comparison of the Tendon Damage Caused by Four Different Anchor Systems Used in Transtendon Rotator Cuff Repair
title_full Comparison of the Tendon Damage Caused by Four Different Anchor Systems Used in Transtendon Rotator Cuff Repair
title_fullStr Comparison of the Tendon Damage Caused by Four Different Anchor Systems Used in Transtendon Rotator Cuff Repair
title_full_unstemmed Comparison of the Tendon Damage Caused by Four Different Anchor Systems Used in Transtendon Rotator Cuff Repair
title_short Comparison of the Tendon Damage Caused by Four Different Anchor Systems Used in Transtendon Rotator Cuff Repair
title_sort comparison of the tendon damage caused by four different anchor systems used in transtendon rotator cuff repair
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3395215/
https://www.ncbi.nlm.nih.gov/pubmed/22811923
http://dx.doi.org/10.1155/2012/798521
work_keys_str_mv AT zhangqingsong comparisonofthetendondamagecausedbyfourdifferentanchorsystemsusedintranstendonrotatorcuffrepair
AT liusen comparisonofthetendondamagecausedbyfourdifferentanchorsystemsusedintranstendonrotatorcuffrepair
AT zhangqiuyang comparisonofthetendondamagecausedbyfourdifferentanchorsystemsusedintranstendonrotatorcuffrepair
AT xueyun comparisonofthetendondamagecausedbyfourdifferentanchorsystemsusedintranstendonrotatorcuffrepair
AT gedongxia comparisonofthetendondamagecausedbyfourdifferentanchorsystemsusedintranstendonrotatorcuffrepair
AT obrienmichaelj comparisonofthetendondamagecausedbyfourdifferentanchorsystemsusedintranstendonrotatorcuffrepair
AT savoiefelixh comparisonofthetendondamagecausedbyfourdifferentanchorsystemsusedintranstendonrotatorcuffrepair
AT youzongbing comparisonofthetendondamagecausedbyfourdifferentanchorsystemsusedintranstendonrotatorcuffrepair