Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes

PURPOSE: To compare the precision of central corneal thickness (CCT) measurements taken with the handheld ultrasound pachymeter (USP), ultrasound biomicroscopy (UBM), and the Artemis-2 very high frequency ultrasound scanner (VHFUS) on normal subjects. DESIGN: Prospective study. METHODS: One eye from...

Descripción completa

Detalles Bibliográficos
Autores principales: Al-Farhan, Haya M, Al-Otaibi, Wafa’a Majed
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3402126/
https://www.ncbi.nlm.nih.gov/pubmed/22848145
http://dx.doi.org/10.2147/OPTH.S32955
_version_ 1782238704655400960
author Al-Farhan, Haya M
Al-Otaibi, Wafa’a Majed
author_facet Al-Farhan, Haya M
Al-Otaibi, Wafa’a Majed
author_sort Al-Farhan, Haya M
collection PubMed
description PURPOSE: To compare the precision of central corneal thickness (CCT) measurements taken with the handheld ultrasound pachymeter (USP), ultrasound biomicroscopy (UBM), and the Artemis-2 very high frequency ultrasound scanner (VHFUS) on normal subjects. DESIGN: Prospective study. METHODS: One eye from each of 61 normal subjects was randomly selected for this study. The measurements of the CCT were taken with the USP, VHFUS, and UBM. Results were compared statistically using repeated-measures analysis of variance (ANOVA), Pearson’s correlation coefficient, and limits of agreement. RESULTS: The average CCT (± standard deviation) was 530.1 ± 30.5 μm, 554.9 ± 31.7 μm, and 559.5 ± 30.7 μm for UBM, VHFUS, and USP respectively. The intraobserver repeatability analyses of variance are not significant for USP, UBM, and VHFUS. P-values were 0.17, 0.19, and 0.37 respectively. Repeated-measures ANOVA showed a significant difference between the three different methods of measuring CCT (P = 0.0001). The ANOVA test revealed no statistically significant difference between USP and VHFUS (P > 0.05), yet statistical significant differences with UBM versus USP and UBM versus VHFUS (P < 0.001). There were high correlations between the three instruments (P < 0.0001). The mean differences (and upper/lower limits of agreement) for CCT measurements were 29.4 ± 14.3 (2.7/56), 4.6 ± 8.6 (−14.7/23.8), and −24.8 ± 13.1 (−50.4/0.8) for USP versus UBM, USP versus VHFUS, and UBM versus VHFUS, respectively. CONCLUSION: The UBM produces CCT measurements that vary significantly from those returned by the USP and the VHFUS, suggesting that the UBM may not be used interchangeably with either equipment for monitoring the CCT in the clinical setting.
format Online
Article
Text
id pubmed-3402126
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-34021262012-07-30 Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes Al-Farhan, Haya M Al-Otaibi, Wafa’a Majed Clin Ophthalmol Original Research PURPOSE: To compare the precision of central corneal thickness (CCT) measurements taken with the handheld ultrasound pachymeter (USP), ultrasound biomicroscopy (UBM), and the Artemis-2 very high frequency ultrasound scanner (VHFUS) on normal subjects. DESIGN: Prospective study. METHODS: One eye from each of 61 normal subjects was randomly selected for this study. The measurements of the CCT were taken with the USP, VHFUS, and UBM. Results were compared statistically using repeated-measures analysis of variance (ANOVA), Pearson’s correlation coefficient, and limits of agreement. RESULTS: The average CCT (± standard deviation) was 530.1 ± 30.5 μm, 554.9 ± 31.7 μm, and 559.5 ± 30.7 μm for UBM, VHFUS, and USP respectively. The intraobserver repeatability analyses of variance are not significant for USP, UBM, and VHFUS. P-values were 0.17, 0.19, and 0.37 respectively. Repeated-measures ANOVA showed a significant difference between the three different methods of measuring CCT (P = 0.0001). The ANOVA test revealed no statistically significant difference between USP and VHFUS (P > 0.05), yet statistical significant differences with UBM versus USP and UBM versus VHFUS (P < 0.001). There were high correlations between the three instruments (P < 0.0001). The mean differences (and upper/lower limits of agreement) for CCT measurements were 29.4 ± 14.3 (2.7/56), 4.6 ± 8.6 (−14.7/23.8), and −24.8 ± 13.1 (−50.4/0.8) for USP versus UBM, USP versus VHFUS, and UBM versus VHFUS, respectively. CONCLUSION: The UBM produces CCT measurements that vary significantly from those returned by the USP and the VHFUS, suggesting that the UBM may not be used interchangeably with either equipment for monitoring the CCT in the clinical setting. Dove Medical Press 2012 2012-07-06 /pmc/articles/PMC3402126/ /pubmed/22848145 http://dx.doi.org/10.2147/OPTH.S32955 Text en © 2012 Al-Farhan and Al-Otaibi, publisher and licensee Dove Medical Press Ltd. This is an Open Access article which permits unrestricted noncommercial use, provided the original work is properly cited.
spellingShingle Original Research
Al-Farhan, Haya M
Al-Otaibi, Wafa’a Majed
Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes
title Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes
title_full Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes
title_fullStr Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes
title_full_unstemmed Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes
title_short Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes
title_sort comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the artemis-2 vhf scanner in normal eyes
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3402126/
https://www.ncbi.nlm.nih.gov/pubmed/22848145
http://dx.doi.org/10.2147/OPTH.S32955
work_keys_str_mv AT alfarhanhayam comparisonofcentralcornealthicknessmeasurementsusingultrasoundpachymetryultrasoundbiomicroscopyandtheartemis2vhfscannerinnormaleyes
AT alotaibiwafaamajed comparisonofcentralcornealthicknessmeasurementsusingultrasoundpachymetryultrasoundbiomicroscopyandtheartemis2vhfscannerinnormaleyes