Low tube voltage CT for improved detection of pancreatic cancer: detection threshold for small, simulated lesions
BACKGROUND: Pancreatic ductal adenocarcinoma is associated with dismal prognosis. The detection of small pancreatic tumors which are still resectable is still a challenging problem. The aim of this study was to investigate the effect of decreasing the tube voltage from 120 to 80 kV on the detection...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3414789/ https://www.ncbi.nlm.nih.gov/pubmed/22828284 http://dx.doi.org/10.1186/1471-2342-12-20 |
_version_ | 1782240257075314688 |
---|---|
author | Holm, Jon Loizou, Louiza Albiin, Nils Kartalis, Nikolaos Leidner, Bertil Sundin, Anders |
author_facet | Holm, Jon Loizou, Louiza Albiin, Nils Kartalis, Nikolaos Leidner, Bertil Sundin, Anders |
author_sort | Holm, Jon |
collection | PubMed |
description | BACKGROUND: Pancreatic ductal adenocarcinoma is associated with dismal prognosis. The detection of small pancreatic tumors which are still resectable is still a challenging problem. The aim of this study was to investigate the effect of decreasing the tube voltage from 120 to 80 kV on the detection of pancreatic tumors. METHODS: Three scanning protocols was used; one using the standard tube voltage (120 kV) and current (160 mA) and two using 80 kV but with different tube currents (500 and 675 mA) to achieve equivalent dose (15 mGy) and noise (15 HU) as that of the standard protocol. Tumors were simulated into collected CT phantom images. The attenuation in normal parenchyma at 120 kV was set at 130 HU, as measured previously in clinical examinations, and the tumor attenuation was assumed to differ 20 HU and was set at 110HU. By scanning and measuring of iodine solution with different concentrations the corresponding tumor and parenchyma attenuation at 80 kV was found to be 185 and 219 HU, respectively. To objectively evaluate the differences between the three protocols, a multi-reader multi-case receiver operating characteristic study was conducted, using three readers and 100 cases, each containing 0–3 lesions. RESULTS: The highest reader averaged figure-of-merit (FOM) was achieved for 80 kV and 675 mA (FOM = 0,850), and the lowest for 120 kV (FOM = 0,709). There was a significant difference between the three protocols (p < 0,0001), when making an analysis of variance (ANOVA). Post-hoc analysis (students t-test) shows that there was a significant difference between 120 and 80 kV, but not between the two levels of tube currents at 80 kV. CONCLUSION: We conclude that when decreasing the tube voltage there is a significant improvement in tumor conspicuity. |
format | Online Article Text |
id | pubmed-3414789 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-34147892012-08-13 Low tube voltage CT for improved detection of pancreatic cancer: detection threshold for small, simulated lesions Holm, Jon Loizou, Louiza Albiin, Nils Kartalis, Nikolaos Leidner, Bertil Sundin, Anders BMC Med Imaging Research Article BACKGROUND: Pancreatic ductal adenocarcinoma is associated with dismal prognosis. The detection of small pancreatic tumors which are still resectable is still a challenging problem. The aim of this study was to investigate the effect of decreasing the tube voltage from 120 to 80 kV on the detection of pancreatic tumors. METHODS: Three scanning protocols was used; one using the standard tube voltage (120 kV) and current (160 mA) and two using 80 kV but with different tube currents (500 and 675 mA) to achieve equivalent dose (15 mGy) and noise (15 HU) as that of the standard protocol. Tumors were simulated into collected CT phantom images. The attenuation in normal parenchyma at 120 kV was set at 130 HU, as measured previously in clinical examinations, and the tumor attenuation was assumed to differ 20 HU and was set at 110HU. By scanning and measuring of iodine solution with different concentrations the corresponding tumor and parenchyma attenuation at 80 kV was found to be 185 and 219 HU, respectively. To objectively evaluate the differences between the three protocols, a multi-reader multi-case receiver operating characteristic study was conducted, using three readers and 100 cases, each containing 0–3 lesions. RESULTS: The highest reader averaged figure-of-merit (FOM) was achieved for 80 kV and 675 mA (FOM = 0,850), and the lowest for 120 kV (FOM = 0,709). There was a significant difference between the three protocols (p < 0,0001), when making an analysis of variance (ANOVA). Post-hoc analysis (students t-test) shows that there was a significant difference between 120 and 80 kV, but not between the two levels of tube currents at 80 kV. CONCLUSION: We conclude that when decreasing the tube voltage there is a significant improvement in tumor conspicuity. BioMed Central 2012-07-24 /pmc/articles/PMC3414789/ /pubmed/22828284 http://dx.doi.org/10.1186/1471-2342-12-20 Text en Copyright ©2012 Holm et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Holm, Jon Loizou, Louiza Albiin, Nils Kartalis, Nikolaos Leidner, Bertil Sundin, Anders Low tube voltage CT for improved detection of pancreatic cancer: detection threshold for small, simulated lesions |
title | Low tube voltage CT for improved detection of pancreatic cancer: detection threshold for small, simulated lesions |
title_full | Low tube voltage CT for improved detection of pancreatic cancer: detection threshold for small, simulated lesions |
title_fullStr | Low tube voltage CT for improved detection of pancreatic cancer: detection threshold for small, simulated lesions |
title_full_unstemmed | Low tube voltage CT for improved detection of pancreatic cancer: detection threshold for small, simulated lesions |
title_short | Low tube voltage CT for improved detection of pancreatic cancer: detection threshold for small, simulated lesions |
title_sort | low tube voltage ct for improved detection of pancreatic cancer: detection threshold for small, simulated lesions |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3414789/ https://www.ncbi.nlm.nih.gov/pubmed/22828284 http://dx.doi.org/10.1186/1471-2342-12-20 |
work_keys_str_mv | AT holmjon lowtubevoltagectforimproveddetectionofpancreaticcancerdetectionthresholdforsmallsimulatedlesions AT loizoulouiza lowtubevoltagectforimproveddetectionofpancreaticcancerdetectionthresholdforsmallsimulatedlesions AT albiinnils lowtubevoltagectforimproveddetectionofpancreaticcancerdetectionthresholdforsmallsimulatedlesions AT kartalisnikolaos lowtubevoltagectforimproveddetectionofpancreaticcancerdetectionthresholdforsmallsimulatedlesions AT leidnerbertil lowtubevoltagectforimproveddetectionofpancreaticcancerdetectionthresholdforsmallsimulatedlesions AT sundinanders lowtubevoltagectforimproveddetectionofpancreaticcancerdetectionthresholdforsmallsimulatedlesions |