Performance of a malaria microscopy image analysis slide reading device

BACKGROUND: Viewing Plasmodium in Romanovsky-stained blood has long been considered the gold standard for diagnosis and a cornerstone in management of the disease. This method however, requires a subjective evaluation by trained, experienced diagnosticians and establishing proficiency of diagnosis i...

Descripción completa

Detalles Bibliográficos
Autores principales: Prescott, William R, Jordan, Robert G, Grobusch, Martin P, Chinchilli, Vernon M, Kleinschmidt, Immo, Borovsky, Joseph, Plaskow, Mark, Torrez, Miguel, Mico, Maximo, Schwabe, Christopher
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3416659/
https://www.ncbi.nlm.nih.gov/pubmed/22559294
http://dx.doi.org/10.1186/1475-2875-11-155
_version_ 1782240416731496448
author Prescott, William R
Jordan, Robert G
Grobusch, Martin P
Chinchilli, Vernon M
Kleinschmidt, Immo
Borovsky, Joseph
Plaskow, Mark
Torrez, Miguel
Mico, Maximo
Schwabe, Christopher
author_facet Prescott, William R
Jordan, Robert G
Grobusch, Martin P
Chinchilli, Vernon M
Kleinschmidt, Immo
Borovsky, Joseph
Plaskow, Mark
Torrez, Miguel
Mico, Maximo
Schwabe, Christopher
author_sort Prescott, William R
collection PubMed
description BACKGROUND: Viewing Plasmodium in Romanovsky-stained blood has long been considered the gold standard for diagnosis and a cornerstone in management of the disease. This method however, requires a subjective evaluation by trained, experienced diagnosticians and establishing proficiency of diagnosis is fraught with many challenges. Reported here is an evaluation of a diagnostic system (a “device” consisting of a microscope, a scanner, and a computer algorithm) that evaluates scanned images of standard Giemsa-stained slides and reports species and parasitaemia. METHODS: The device was challenged with two independent tests: a 55 slide, expert slide reading test the composition of which has been published by the World Health Organization (“WHO55” test), and a second test in which slides were made from a sample of consenting subjects participating in a malaria incidence survey conducted in Equatorial Guinea (EGMIS test). These subjects’ blood was tested by malaria RDT as well as having the blood smear diagnosis unequivocally determined by a worldwide panel of a minimum of six reference microscopists. Only slides with unequivocal microscopic diagnoses were used for the device challenge, n = 119. RESULTS: On the WHO55 test, the device scored a “Level 4” using the WHO published grading scheme. Broken down by more traditional analysis parameters this result was translated to 89% and 70% sensitivity and specificity, respectively. Species were correctly identified in 61% of the slides and the quantification of parasites fell within acceptable range of the validated parasitaemia in 10% of the cases. On the EGMIS test it scored 100% and 94% sensitivity/specificity, with 64% of the species correct and 45% of the parasitaemia within an acceptable range. A pooled analysis of the 174 slides used for both tests resulted in an overall 92% sensitivity and 90% specificity with 61% species and 19% quantifications correct. CONCLUSIONS: In its current manifestation, the device performs at a level comparable to that of many human slide readers. Because its use requires minimal additional equipment and it uses standard stained slides as starting material, its widespread adoption may eliminate the current uncertainty about the quality of microscopic diagnoses worldwide.
format Online
Article
Text
id pubmed-3416659
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-34166592012-08-11 Performance of a malaria microscopy image analysis slide reading device Prescott, William R Jordan, Robert G Grobusch, Martin P Chinchilli, Vernon M Kleinschmidt, Immo Borovsky, Joseph Plaskow, Mark Torrez, Miguel Mico, Maximo Schwabe, Christopher Malar J Research BACKGROUND: Viewing Plasmodium in Romanovsky-stained blood has long been considered the gold standard for diagnosis and a cornerstone in management of the disease. This method however, requires a subjective evaluation by trained, experienced diagnosticians and establishing proficiency of diagnosis is fraught with many challenges. Reported here is an evaluation of a diagnostic system (a “device” consisting of a microscope, a scanner, and a computer algorithm) that evaluates scanned images of standard Giemsa-stained slides and reports species and parasitaemia. METHODS: The device was challenged with two independent tests: a 55 slide, expert slide reading test the composition of which has been published by the World Health Organization (“WHO55” test), and a second test in which slides were made from a sample of consenting subjects participating in a malaria incidence survey conducted in Equatorial Guinea (EGMIS test). These subjects’ blood was tested by malaria RDT as well as having the blood smear diagnosis unequivocally determined by a worldwide panel of a minimum of six reference microscopists. Only slides with unequivocal microscopic diagnoses were used for the device challenge, n = 119. RESULTS: On the WHO55 test, the device scored a “Level 4” using the WHO published grading scheme. Broken down by more traditional analysis parameters this result was translated to 89% and 70% sensitivity and specificity, respectively. Species were correctly identified in 61% of the slides and the quantification of parasites fell within acceptable range of the validated parasitaemia in 10% of the cases. On the EGMIS test it scored 100% and 94% sensitivity/specificity, with 64% of the species correct and 45% of the parasitaemia within an acceptable range. A pooled analysis of the 174 slides used for both tests resulted in an overall 92% sensitivity and 90% specificity with 61% species and 19% quantifications correct. CONCLUSIONS: In its current manifestation, the device performs at a level comparable to that of many human slide readers. Because its use requires minimal additional equipment and it uses standard stained slides as starting material, its widespread adoption may eliminate the current uncertainty about the quality of microscopic diagnoses worldwide. BioMed Central 2012-05-06 /pmc/articles/PMC3416659/ /pubmed/22559294 http://dx.doi.org/10.1186/1475-2875-11-155 Text en Copyright ©2012 Prescott et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Prescott, William R
Jordan, Robert G
Grobusch, Martin P
Chinchilli, Vernon M
Kleinschmidt, Immo
Borovsky, Joseph
Plaskow, Mark
Torrez, Miguel
Mico, Maximo
Schwabe, Christopher
Performance of a malaria microscopy image analysis slide reading device
title Performance of a malaria microscopy image analysis slide reading device
title_full Performance of a malaria microscopy image analysis slide reading device
title_fullStr Performance of a malaria microscopy image analysis slide reading device
title_full_unstemmed Performance of a malaria microscopy image analysis slide reading device
title_short Performance of a malaria microscopy image analysis slide reading device
title_sort performance of a malaria microscopy image analysis slide reading device
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3416659/
https://www.ncbi.nlm.nih.gov/pubmed/22559294
http://dx.doi.org/10.1186/1475-2875-11-155
work_keys_str_mv AT prescottwilliamr performanceofamalariamicroscopyimageanalysisslidereadingdevice
AT jordanrobertg performanceofamalariamicroscopyimageanalysisslidereadingdevice
AT grobuschmartinp performanceofamalariamicroscopyimageanalysisslidereadingdevice
AT chinchillivernonm performanceofamalariamicroscopyimageanalysisslidereadingdevice
AT kleinschmidtimmo performanceofamalariamicroscopyimageanalysisslidereadingdevice
AT borovskyjoseph performanceofamalariamicroscopyimageanalysisslidereadingdevice
AT plaskowmark performanceofamalariamicroscopyimageanalysisslidereadingdevice
AT torrezmiguel performanceofamalariamicroscopyimageanalysisslidereadingdevice
AT micomaximo performanceofamalariamicroscopyimageanalysisslidereadingdevice
AT schwabechristopher performanceofamalariamicroscopyimageanalysisslidereadingdevice