Dosimetric evaluation of a 320 detector row CT scanner unit

BACKGROUND: The technologic improvements in Multislice scanners include the increment in the X-ray beam width. Some new CT scanners are equipped with a 320 detector row which allows a longitudinal coverage of 160 mm and a total of 640 slices for a single rotation. When such parameters are used the l...

Descripción completa

Detalles Bibliográficos
Autores principales: de Denaro, Mario, Bregant, Paola
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Versita, Warsaw 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3423722/
https://www.ncbi.nlm.nih.gov/pubmed/22933937
http://dx.doi.org/10.2478/v10019-010-0049-1
_version_ 1782241136928096256
author de Denaro, Mario
Bregant, Paola
author_facet de Denaro, Mario
Bregant, Paola
author_sort de Denaro, Mario
collection PubMed
description BACKGROUND: The technologic improvements in Multislice scanners include the increment in the X-ray beam width. Some new CT scanners are equipped with a 320 detector row which allows a longitudinal coverage of 160 mm and a total of 640 slices for a single rotation. When such parameters are used the length of the traditional pencil chamber (10 cm) is no more appropriate to measure the standard weighted computed tomography dose index (CTDI(w)) value. MATERIALS AND METHODS. Dosimetric measurements were performed on a 640 slices Toshiba Aquilion One CT scanner using common instrumentation available in Medical Physics Departments. RESULTS: For the measurements in air, two different ionization chambers were completely exposed to the beam. Dosimeters showed an acceptable agreement in the measurements. To evaluate the actual shape of the dose profile strips of Gafchromic XRQA film were used. Films were previously calibrated on site. From the graphic response of the scanned film it is possible to evaluate the full width at half maximum (FWHM) of the dose profile which represent the actual beam width. CONCLUSIONS: Computed Tomography Dose Index (CTDI) and Dose Length Product (DLP) need to be changed when the beam width of the CT scanner is over 100 mm. To perform dose evaluation with the conventional instrumentation, two parameters should be considered: the average absorbed dose and the actual beam width. To measure the average absorbed dose, the conventional ionization chamber can be used. For the measurement of the width of the dose profile, Gafchromic XRQA film seemed to be suitable.
format Online
Article
Text
id pubmed-3423722
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher Versita, Warsaw
record_format MEDLINE/PubMed
spelling pubmed-34237222012-08-29 Dosimetric evaluation of a 320 detector row CT scanner unit de Denaro, Mario Bregant, Paola Radiol Oncol Research Article BACKGROUND: The technologic improvements in Multislice scanners include the increment in the X-ray beam width. Some new CT scanners are equipped with a 320 detector row which allows a longitudinal coverage of 160 mm and a total of 640 slices for a single rotation. When such parameters are used the length of the traditional pencil chamber (10 cm) is no more appropriate to measure the standard weighted computed tomography dose index (CTDI(w)) value. MATERIALS AND METHODS. Dosimetric measurements were performed on a 640 slices Toshiba Aquilion One CT scanner using common instrumentation available in Medical Physics Departments. RESULTS: For the measurements in air, two different ionization chambers were completely exposed to the beam. Dosimeters showed an acceptable agreement in the measurements. To evaluate the actual shape of the dose profile strips of Gafchromic XRQA film were used. Films were previously calibrated on site. From the graphic response of the scanned film it is possible to evaluate the full width at half maximum (FWHM) of the dose profile which represent the actual beam width. CONCLUSIONS: Computed Tomography Dose Index (CTDI) and Dose Length Product (DLP) need to be changed when the beam width of the CT scanner is over 100 mm. To perform dose evaluation with the conventional instrumentation, two parameters should be considered: the average absorbed dose and the actual beam width. To measure the average absorbed dose, the conventional ionization chamber can be used. For the measurement of the width of the dose profile, Gafchromic XRQA film seemed to be suitable. Versita, Warsaw 2010-11-25 /pmc/articles/PMC3423722/ /pubmed/22933937 http://dx.doi.org/10.2478/v10019-010-0049-1 Text en Copyright © by Association of Radiology & Oncology http://creativecommons.org/licenses/by/3.0 This article is an open-access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/).
spellingShingle Research Article
de Denaro, Mario
Bregant, Paola
Dosimetric evaluation of a 320 detector row CT scanner unit
title Dosimetric evaluation of a 320 detector row CT scanner unit
title_full Dosimetric evaluation of a 320 detector row CT scanner unit
title_fullStr Dosimetric evaluation of a 320 detector row CT scanner unit
title_full_unstemmed Dosimetric evaluation of a 320 detector row CT scanner unit
title_short Dosimetric evaluation of a 320 detector row CT scanner unit
title_sort dosimetric evaluation of a 320 detector row ct scanner unit
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3423722/
https://www.ncbi.nlm.nih.gov/pubmed/22933937
http://dx.doi.org/10.2478/v10019-010-0049-1
work_keys_str_mv AT dedenaromario dosimetricevaluationofa320detectorrowctscannerunit
AT bregantpaola dosimetricevaluationofa320detectorrowctscannerunit