Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments
BACKGROUND: Hemostasis is a central issue in laparoscopic surgery. Ultrasonic scissors and bipolar clamps are commonly used, with known advantages with each technique. METHODS: The prototype of new surgical scissors, delivering ultrasonically generated frictional heat energy and bipolar heat energy...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer-Verlag
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3427482/ https://www.ncbi.nlm.nih.gov/pubmed/22447285 http://dx.doi.org/10.1007/s00464-012-2229-0 |
_version_ | 1782241613941047296 |
---|---|
author | Seehofer, Daniel Mogl, Martina Boas-Knoop, Sabine Unger, Juliane Schirmeier, Anja Chopra, Sascha Eurich, Dennis |
author_facet | Seehofer, Daniel Mogl, Martina Boas-Knoop, Sabine Unger, Juliane Schirmeier, Anja Chopra, Sascha Eurich, Dennis |
author_sort | Seehofer, Daniel |
collection | PubMed |
description | BACKGROUND: Hemostasis is a central issue in laparoscopic surgery. Ultrasonic scissors and bipolar clamps are commonly used, with known advantages with each technique. METHODS: The prototype of new surgical scissors, delivering ultrasonically generated frictional heat energy and bipolar heat energy simultaneously (THUNDERBEAT(®) [TB]), was compared to ultrasonic scissors (Harmonic ACE(®) [HA]) and an advanced bipolar device (LigaSure(®) [LS]) using a pig model. As safety parameters, temperature profiles after single activation and after a defined cut were determined. As efficacy parameters, seal failures and the maximum burst pressure (BP) were measured after in vivo sealing of vessels of various types and diameters (categories 2–4 and 5–7 mm). Moreover, the vertical width of the tissue seal was measured on serial histological slices of selected arteries. The cutting speed was measured during division of isolated arteries and during dissection of a defined length of compound tissue (10 cm of mesentery). Burst-pressure measurement and histological analysis were performed by investigators blinded to the used sealing device. RESULTS: Using the TB, the burst pressure in larger arteries was significantly higher (734 ± 64 mmHg) than that of the HA (453 ± 50 mmHg). No differences in the rate of seal failures were observed. The cutting speed of the TB was significantly higher than that of all other devices. Safety evaluation revealed temperatures below 100 °C in the bipolar device. The maximum temperature of the HA and the TB was significantly higher. No relevant differences were observed between the HA and the TB. CONCLUSIONS: The ultrasonic and bipolar technique of the TB has the potential to surpass the dissection speed of ultrasonic devices with the sealing efficacy of bipolar clamps. However, heat production that is comparable to conventional ultrasonic scissors should be minded for clinical use. |
format | Online Article Text |
id | pubmed-3427482 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | Springer-Verlag |
record_format | MEDLINE/PubMed |
spelling | pubmed-34274822012-08-30 Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments Seehofer, Daniel Mogl, Martina Boas-Knoop, Sabine Unger, Juliane Schirmeier, Anja Chopra, Sascha Eurich, Dennis Surg Endosc Article BACKGROUND: Hemostasis is a central issue in laparoscopic surgery. Ultrasonic scissors and bipolar clamps are commonly used, with known advantages with each technique. METHODS: The prototype of new surgical scissors, delivering ultrasonically generated frictional heat energy and bipolar heat energy simultaneously (THUNDERBEAT(®) [TB]), was compared to ultrasonic scissors (Harmonic ACE(®) [HA]) and an advanced bipolar device (LigaSure(®) [LS]) using a pig model. As safety parameters, temperature profiles after single activation and after a defined cut were determined. As efficacy parameters, seal failures and the maximum burst pressure (BP) were measured after in vivo sealing of vessels of various types and diameters (categories 2–4 and 5–7 mm). Moreover, the vertical width of the tissue seal was measured on serial histological slices of selected arteries. The cutting speed was measured during division of isolated arteries and during dissection of a defined length of compound tissue (10 cm of mesentery). Burst-pressure measurement and histological analysis were performed by investigators blinded to the used sealing device. RESULTS: Using the TB, the burst pressure in larger arteries was significantly higher (734 ± 64 mmHg) than that of the HA (453 ± 50 mmHg). No differences in the rate of seal failures were observed. The cutting speed of the TB was significantly higher than that of all other devices. Safety evaluation revealed temperatures below 100 °C in the bipolar device. The maximum temperature of the HA and the TB was significantly higher. No relevant differences were observed between the HA and the TB. CONCLUSIONS: The ultrasonic and bipolar technique of the TB has the potential to surpass the dissection speed of ultrasonic devices with the sealing efficacy of bipolar clamps. However, heat production that is comparable to conventional ultrasonic scissors should be minded for clinical use. Springer-Verlag 2012-03-24 2012 /pmc/articles/PMC3427482/ /pubmed/22447285 http://dx.doi.org/10.1007/s00464-012-2229-0 Text en © The Author(s) 2012 https://creativecommons.org/licenses/by/4.0/ This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited. |
spellingShingle | Article Seehofer, Daniel Mogl, Martina Boas-Knoop, Sabine Unger, Juliane Schirmeier, Anja Chopra, Sascha Eurich, Dennis Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments |
title | Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments |
title_full | Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments |
title_fullStr | Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments |
title_full_unstemmed | Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments |
title_short | Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments |
title_sort | safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3427482/ https://www.ncbi.nlm.nih.gov/pubmed/22447285 http://dx.doi.org/10.1007/s00464-012-2229-0 |
work_keys_str_mv | AT seehoferdaniel safetyandefficacyofnewintegratedbipolarandultrasonicscissorscomparedtoconventionallaparoscopic5mmsealingandcuttinginstruments AT moglmartina safetyandefficacyofnewintegratedbipolarandultrasonicscissorscomparedtoconventionallaparoscopic5mmsealingandcuttinginstruments AT boasknoopsabine safetyandefficacyofnewintegratedbipolarandultrasonicscissorscomparedtoconventionallaparoscopic5mmsealingandcuttinginstruments AT ungerjuliane safetyandefficacyofnewintegratedbipolarandultrasonicscissorscomparedtoconventionallaparoscopic5mmsealingandcuttinginstruments AT schirmeieranja safetyandefficacyofnewintegratedbipolarandultrasonicscissorscomparedtoconventionallaparoscopic5mmsealingandcuttinginstruments AT choprasascha safetyandefficacyofnewintegratedbipolarandultrasonicscissorscomparedtoconventionallaparoscopic5mmsealingandcuttinginstruments AT eurichdennis safetyandefficacyofnewintegratedbipolarandultrasonicscissorscomparedtoconventionallaparoscopic5mmsealingandcuttinginstruments |