Cargando…
Measurement delay associated with the Guardian® RT continuous glucose monitoring system
AIMS: Using compartment modelling, we assessed the time delay between blood glucose and sensor glucose measured by the Guardian® RT continuous glucose monitoring system in young subjects with Type 1 diabetes (T1D). METHODS: Twelve children and adolescents with T1D treated by continuous subcutaneous...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Blackwell Publishing Ltd
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3427888/ https://www.ncbi.nlm.nih.gov/pubmed/20121899 http://dx.doi.org/10.1111/j.1464-5491.2009.02887.x |
Sumario: | AIMS: Using compartment modelling, we assessed the time delay between blood glucose and sensor glucose measured by the Guardian® RT continuous glucose monitoring system in young subjects with Type 1 diabetes (T1D). METHODS: Twelve children and adolescents with T1D treated by continuous subcutaneous insulin infusion (male/female 7/5; age 13.1 ± 4.2 years; body mass index 21.9 ± 4.3 kg/m(2); mean ± sd) were studied over 19 h in a Clinical Research Facility. Guardian® RT was calibrated every 6 h and sensor glucose measured every 5 min. Reference blood glucose was measured every 15 min using a YSI 2300 STAT Plus Analyser. A population compartment model of sensor glucose–blood glucose kinetics was adopted to estimate the time delay, the calibration scale and the calibration shift. RESULTS: The population median of the time delay was 15.8 (interquartile range 15.2, 16.5) min, which was corroborated by correlation analysis between blood glucose and 15-min delayed sensor glucose. The delay has a relatively low intersubject variability, with 95% of individuals predicted to have delays between 10.4 and 24.3 min. Population medians (interquartile range) for the scale and shift are 0.800 (0.777, 0.823) (unitless) and 1.66 (1.47, 1.84) mmol/l, respectively. CONCLUSIONS: In young subjects with T1D, the total time delay associated with the Guardian® RT system was approximately 15 min. This is twice that expected on physiological grounds, suggesting a 5- to 10-min delay because of data processing. Delays above 25 min are rarely to be observed. |
---|