Dosimetric evaluation of the impacts of different heterogeneity correction algorithms on target doses in stereotactic body radiation therapy for lung tumors
Heterogeneity correction algorithms can have a large impact on the dose distributions of stereotactic body radiation therapy (SBRT) for lung tumors. Treatment plans of 20 patients who underwent SBRT for lung tumors with the prescribed dose of 48 Gy in four fractions at the isocenter were reviewed re...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3430415/ https://www.ncbi.nlm.nih.gov/pubmed/22843364 http://dx.doi.org/10.1093/jrr/rrs026 |
_version_ | 1782241937970954240 |
---|---|
author | Narabayashi, Masaru Mizowaki, Takashi Matsuo, Yukinori Nakamura, Mitsuhiro Takayama, Kenji Norihisa, Yoshiki Sakanaka, Katsuyuki Hiraoka, Masahiro |
author_facet | Narabayashi, Masaru Mizowaki, Takashi Matsuo, Yukinori Nakamura, Mitsuhiro Takayama, Kenji Norihisa, Yoshiki Sakanaka, Katsuyuki Hiraoka, Masahiro |
author_sort | Narabayashi, Masaru |
collection | PubMed |
description | Heterogeneity correction algorithms can have a large impact on the dose distributions of stereotactic body radiation therapy (SBRT) for lung tumors. Treatment plans of 20 patients who underwent SBRT for lung tumors with the prescribed dose of 48 Gy in four fractions at the isocenter were reviewed retrospectively and recalculated with different heterogeneity correction algorithms: the pencil beam convolution algorithm with a Batho power-law correction (BPL) in Eclipse, the radiological path length algorithm (RPL), and the X-ray Voxel Monte Carlo algorithm (XVMC) in iPlan. The doses at the periphery (minimum dose and D95) of the planning target volume (PTV) were compared using the same monitor units among the three heterogeneity correction algorithms, and the monitor units were compared between two methods of dose prescription, that is, an isocenter dose prescription (IC prescription) and dose–volume based prescription (D95 prescription). Mean values of the dose at the periphery of the PTV were significantly lower with XVMC than with BPL using the same monitor units (P < 0.001). In addition, under IC prescription using BPL, RPL and XVMC, the ratios of mean values of monitor units were 1, 0.959 and 0.986, respectively. Under D95 prescription, they were 1, 0.937 and 1.088, respectively. These observations indicated that the application of XVMC under D95 prescription results in an increase in the actually delivered dose by 8.8% on average compared with the application of BPL. The appropriateness of switching heterogeneity correction algorithms and dose prescription methods should be carefully validated from a clinical viewpoint. |
format | Online Article Text |
id | pubmed-3430415 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | Oxford University Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-34304152012-08-29 Dosimetric evaluation of the impacts of different heterogeneity correction algorithms on target doses in stereotactic body radiation therapy for lung tumors Narabayashi, Masaru Mizowaki, Takashi Matsuo, Yukinori Nakamura, Mitsuhiro Takayama, Kenji Norihisa, Yoshiki Sakanaka, Katsuyuki Hiraoka, Masahiro J Radiat Res Technology Heterogeneity correction algorithms can have a large impact on the dose distributions of stereotactic body radiation therapy (SBRT) for lung tumors. Treatment plans of 20 patients who underwent SBRT for lung tumors with the prescribed dose of 48 Gy in four fractions at the isocenter were reviewed retrospectively and recalculated with different heterogeneity correction algorithms: the pencil beam convolution algorithm with a Batho power-law correction (BPL) in Eclipse, the radiological path length algorithm (RPL), and the X-ray Voxel Monte Carlo algorithm (XVMC) in iPlan. The doses at the periphery (minimum dose and D95) of the planning target volume (PTV) were compared using the same monitor units among the three heterogeneity correction algorithms, and the monitor units were compared between two methods of dose prescription, that is, an isocenter dose prescription (IC prescription) and dose–volume based prescription (D95 prescription). Mean values of the dose at the periphery of the PTV were significantly lower with XVMC than with BPL using the same monitor units (P < 0.001). In addition, under IC prescription using BPL, RPL and XVMC, the ratios of mean values of monitor units were 1, 0.959 and 0.986, respectively. Under D95 prescription, they were 1, 0.937 and 1.088, respectively. These observations indicated that the application of XVMC under D95 prescription results in an increase in the actually delivered dose by 8.8% on average compared with the application of BPL. The appropriateness of switching heterogeneity correction algorithms and dose prescription methods should be carefully validated from a clinical viewpoint. Oxford University Press 2012-09 2012-07-13 /pmc/articles/PMC3430415/ /pubmed/22843364 http://dx.doi.org/10.1093/jrr/rrs026 Text en © The Author 2012. Published by Oxford University Press on behalf of The Japan Radiation Research Society and Japanese Society for Therapeutic Radiology and Oncology. http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/), which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Technology Narabayashi, Masaru Mizowaki, Takashi Matsuo, Yukinori Nakamura, Mitsuhiro Takayama, Kenji Norihisa, Yoshiki Sakanaka, Katsuyuki Hiraoka, Masahiro Dosimetric evaluation of the impacts of different heterogeneity correction algorithms on target doses in stereotactic body radiation therapy for lung tumors |
title | Dosimetric evaluation of the impacts of different heterogeneity correction algorithms on target doses in stereotactic body radiation therapy for lung tumors |
title_full | Dosimetric evaluation of the impacts of different heterogeneity correction algorithms on target doses in stereotactic body radiation therapy for lung tumors |
title_fullStr | Dosimetric evaluation of the impacts of different heterogeneity correction algorithms on target doses in stereotactic body radiation therapy for lung tumors |
title_full_unstemmed | Dosimetric evaluation of the impacts of different heterogeneity correction algorithms on target doses in stereotactic body radiation therapy for lung tumors |
title_short | Dosimetric evaluation of the impacts of different heterogeneity correction algorithms on target doses in stereotactic body radiation therapy for lung tumors |
title_sort | dosimetric evaluation of the impacts of different heterogeneity correction algorithms on target doses in stereotactic body radiation therapy for lung tumors |
topic | Technology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3430415/ https://www.ncbi.nlm.nih.gov/pubmed/22843364 http://dx.doi.org/10.1093/jrr/rrs026 |
work_keys_str_mv | AT narabayashimasaru dosimetricevaluationoftheimpactsofdifferentheterogeneitycorrectionalgorithmsontargetdosesinstereotacticbodyradiationtherapyforlungtumors AT mizowakitakashi dosimetricevaluationoftheimpactsofdifferentheterogeneitycorrectionalgorithmsontargetdosesinstereotacticbodyradiationtherapyforlungtumors AT matsuoyukinori dosimetricevaluationoftheimpactsofdifferentheterogeneitycorrectionalgorithmsontargetdosesinstereotacticbodyradiationtherapyforlungtumors AT nakamuramitsuhiro dosimetricevaluationoftheimpactsofdifferentheterogeneitycorrectionalgorithmsontargetdosesinstereotacticbodyradiationtherapyforlungtumors AT takayamakenji dosimetricevaluationoftheimpactsofdifferentheterogeneitycorrectionalgorithmsontargetdosesinstereotacticbodyradiationtherapyforlungtumors AT norihisayoshiki dosimetricevaluationoftheimpactsofdifferentheterogeneitycorrectionalgorithmsontargetdosesinstereotacticbodyradiationtherapyforlungtumors AT sakanakakatsuyuki dosimetricevaluationoftheimpactsofdifferentheterogeneitycorrectionalgorithmsontargetdosesinstereotacticbodyradiationtherapyforlungtumors AT hiraokamasahiro dosimetricevaluationoftheimpactsofdifferentheterogeneitycorrectionalgorithmsontargetdosesinstereotacticbodyradiationtherapyforlungtumors |