Boston type I keratoprosthesis-donor cornea interface evaluated by high-definition spectral-domain anterior segment optical coherence tomography

BACKGROUND: The purpose of this study was to assess whether the resolution offered by two different, recently commercially available high-resolution, spectral-domain anterior segment optical coherence tomography (AS-OCT) instruments allows for detailed anatomic characterization of the critical devic...

Descripción completa

Detalles Bibliográficos
Autores principales: Fernandez, Ana G Alzaga, Radcliffe, Nathan M, Sippel, Kimberly C, Rosenblatt, Mark I, Sood, Priyanka, Starr, Christopher E, Ciralsky, Jessica B, D’Amico, Donald J, Kiss, Szilárd
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3437949/
https://www.ncbi.nlm.nih.gov/pubmed/22969280
http://dx.doi.org/10.2147/OPTH.S34787
_version_ 1782242840428937216
author Fernandez, Ana G Alzaga
Radcliffe, Nathan M
Sippel, Kimberly C
Rosenblatt, Mark I
Sood, Priyanka
Starr, Christopher E
Ciralsky, Jessica B
D’Amico, Donald J
Kiss, Szilárd
author_facet Fernandez, Ana G Alzaga
Radcliffe, Nathan M
Sippel, Kimberly C
Rosenblatt, Mark I
Sood, Priyanka
Starr, Christopher E
Ciralsky, Jessica B
D’Amico, Donald J
Kiss, Szilárd
author_sort Fernandez, Ana G Alzaga
collection PubMed
description BACKGROUND: The purpose of this study was to assess whether the resolution offered by two different, recently commercially available high-resolution, spectral-domain anterior segment optical coherence tomography (AS-OCT) instruments allows for detailed anatomic characterization of the critical device-donor cornea interface in eyes implanted with the Boston type I permanent keratoprosthesis. METHODS: Eighteen eyes of 17 patients implanted with the Boston type I keratoprosthesis were included in this retrospective case series. All eyes were quantitatively evaluated using the Cirrus HD-OCT while a subset (five eyes) was also qualitatively imaged using the Spectralis Anterior Segment Module. Images from these instruments were analyzed for evidence of epithelial migration onto the anterior surface of the keratoprosthesis front plate, and presence of a vertical gap between the posterior surface of the front plate and the underlying carrier donor corneal tissue. Quantitative data was obtained utilizing the caliper function on the Cirrus HD-OCT. RESULTS: The mean duration between AS-OCT imaging and keratoprosthesis placement was 29 months. As assessed by the Cirrus HD-OCT, 83% of eyes exhibited epithelial migration over the edge of the front plate. Fifty-six percent of the keratoprosthesis devices displayed good apposition of the device with the carrier corneal donor tissue. When a vertical gap was present (44% of eyes), the mean gap was 40 (range 8–104) microns. The Spectralis Anterior Segment Module also displayed sufficient resolution to allow for similar characterization of the device-donor cornea interface. CONCLUSION: Spectral-domain AS-OCT permits high resolution imaging of the keratoprosthesis device-donor cornea interface. Both the Cirrus HD-OCT and the Spectralis Anterior Segment module allowed for visualization of epithelial coverage of the device-donor cornea interface, as well as identification of physical gaps. These imaging modalities, by yielding information in regard to integration of the keratoprosthesis with surrounding corneal tissue, may help identify those at risk for keratoprosthesis-related complications, such as extrusion and endophthalmitis, and hence guide clinical management.
format Online
Article
Text
id pubmed-3437949
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-34379492012-09-11 Boston type I keratoprosthesis-donor cornea interface evaluated by high-definition spectral-domain anterior segment optical coherence tomography Fernandez, Ana G Alzaga Radcliffe, Nathan M Sippel, Kimberly C Rosenblatt, Mark I Sood, Priyanka Starr, Christopher E Ciralsky, Jessica B D’Amico, Donald J Kiss, Szilárd Clin Ophthalmol Case Series BACKGROUND: The purpose of this study was to assess whether the resolution offered by two different, recently commercially available high-resolution, spectral-domain anterior segment optical coherence tomography (AS-OCT) instruments allows for detailed anatomic characterization of the critical device-donor cornea interface in eyes implanted with the Boston type I permanent keratoprosthesis. METHODS: Eighteen eyes of 17 patients implanted with the Boston type I keratoprosthesis were included in this retrospective case series. All eyes were quantitatively evaluated using the Cirrus HD-OCT while a subset (five eyes) was also qualitatively imaged using the Spectralis Anterior Segment Module. Images from these instruments were analyzed for evidence of epithelial migration onto the anterior surface of the keratoprosthesis front plate, and presence of a vertical gap between the posterior surface of the front plate and the underlying carrier donor corneal tissue. Quantitative data was obtained utilizing the caliper function on the Cirrus HD-OCT. RESULTS: The mean duration between AS-OCT imaging and keratoprosthesis placement was 29 months. As assessed by the Cirrus HD-OCT, 83% of eyes exhibited epithelial migration over the edge of the front plate. Fifty-six percent of the keratoprosthesis devices displayed good apposition of the device with the carrier corneal donor tissue. When a vertical gap was present (44% of eyes), the mean gap was 40 (range 8–104) microns. The Spectralis Anterior Segment Module also displayed sufficient resolution to allow for similar characterization of the device-donor cornea interface. CONCLUSION: Spectral-domain AS-OCT permits high resolution imaging of the keratoprosthesis device-donor cornea interface. Both the Cirrus HD-OCT and the Spectralis Anterior Segment module allowed for visualization of epithelial coverage of the device-donor cornea interface, as well as identification of physical gaps. These imaging modalities, by yielding information in regard to integration of the keratoprosthesis with surrounding corneal tissue, may help identify those at risk for keratoprosthesis-related complications, such as extrusion and endophthalmitis, and hence guide clinical management. Dove Medical Press 2012 2012-08-27 /pmc/articles/PMC3437949/ /pubmed/22969280 http://dx.doi.org/10.2147/OPTH.S34787 Text en © 2012 Alzaga Fernandez et al, publisher and licensee Dove Medical Press Ltd. This is an Open Access article which permits unrestricted noncommercial use, provided the original work is properly cited.
spellingShingle Case Series
Fernandez, Ana G Alzaga
Radcliffe, Nathan M
Sippel, Kimberly C
Rosenblatt, Mark I
Sood, Priyanka
Starr, Christopher E
Ciralsky, Jessica B
D’Amico, Donald J
Kiss, Szilárd
Boston type I keratoprosthesis-donor cornea interface evaluated by high-definition spectral-domain anterior segment optical coherence tomography
title Boston type I keratoprosthesis-donor cornea interface evaluated by high-definition spectral-domain anterior segment optical coherence tomography
title_full Boston type I keratoprosthesis-donor cornea interface evaluated by high-definition spectral-domain anterior segment optical coherence tomography
title_fullStr Boston type I keratoprosthesis-donor cornea interface evaluated by high-definition spectral-domain anterior segment optical coherence tomography
title_full_unstemmed Boston type I keratoprosthesis-donor cornea interface evaluated by high-definition spectral-domain anterior segment optical coherence tomography
title_short Boston type I keratoprosthesis-donor cornea interface evaluated by high-definition spectral-domain anterior segment optical coherence tomography
title_sort boston type i keratoprosthesis-donor cornea interface evaluated by high-definition spectral-domain anterior segment optical coherence tomography
topic Case Series
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3437949/
https://www.ncbi.nlm.nih.gov/pubmed/22969280
http://dx.doi.org/10.2147/OPTH.S34787
work_keys_str_mv AT fernandezanagalzaga bostontypeikeratoprosthesisdonorcorneainterfaceevaluatedbyhighdefinitionspectraldomainanteriorsegmentopticalcoherencetomography
AT radcliffenathanm bostontypeikeratoprosthesisdonorcorneainterfaceevaluatedbyhighdefinitionspectraldomainanteriorsegmentopticalcoherencetomography
AT sippelkimberlyc bostontypeikeratoprosthesisdonorcorneainterfaceevaluatedbyhighdefinitionspectraldomainanteriorsegmentopticalcoherencetomography
AT rosenblattmarki bostontypeikeratoprosthesisdonorcorneainterfaceevaluatedbyhighdefinitionspectraldomainanteriorsegmentopticalcoherencetomography
AT soodpriyanka bostontypeikeratoprosthesisdonorcorneainterfaceevaluatedbyhighdefinitionspectraldomainanteriorsegmentopticalcoherencetomography
AT starrchristophere bostontypeikeratoprosthesisdonorcorneainterfaceevaluatedbyhighdefinitionspectraldomainanteriorsegmentopticalcoherencetomography
AT ciralskyjessicab bostontypeikeratoprosthesisdonorcorneainterfaceevaluatedbyhighdefinitionspectraldomainanteriorsegmentopticalcoherencetomography
AT damicodonaldj bostontypeikeratoprosthesisdonorcorneainterfaceevaluatedbyhighdefinitionspectraldomainanteriorsegmentopticalcoherencetomography
AT kissszilard bostontypeikeratoprosthesisdonorcorneainterfaceevaluatedbyhighdefinitionspectraldomainanteriorsegmentopticalcoherencetomography