Radiation Dose Reduction during Radial Cardiac Catheterization: Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape
Objectives. Radiation scatter protection shield drapes have been designed with the goal of decreasing radiation dose to the operators during transfemoral catheterization. We sought to investigate the impact on operator radiation exposure of various shielding drapes specifically designed for the radi...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
International Scholarly Research Network
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3439952/ https://www.ncbi.nlm.nih.gov/pubmed/22988525 http://dx.doi.org/10.5402/2012/769167 |
_version_ | 1782243099990294528 |
---|---|
author | Ertel, Andrew Nadelson, Jeffrey Shroff, Adhir R. Sweis, Ranya Ferrera, Dean Vidovich, Mladen I. |
author_facet | Ertel, Andrew Nadelson, Jeffrey Shroff, Adhir R. Sweis, Ranya Ferrera, Dean Vidovich, Mladen I. |
author_sort | Ertel, Andrew |
collection | PubMed |
description | Objectives. Radiation scatter protection shield drapes have been designed with the goal of decreasing radiation dose to the operators during transfemoral catheterization. We sought to investigate the impact on operator radiation exposure of various shielding drapes specifically designed for the radial approach. Background. Radial access for cardiac catheterization has increased due to improved patient comfort and decreased bleeding complications. There are concerns for increased radiation exposure to patients and operators. Methods. Radiation doses to a simulated operator were measured with a RadCal Dosimeter in the cardiac catheterization laboratory. The mock patient was a 97.5 kg fission product phantom. Three lead-free drape designs were studied. The drapes were placed just proximal to the right wrist and extended medially to phantom's trunk. Simulated diagnostic coronary angiography included 6 minutes of fluoroscopy time and 32 seconds of cineangiography time at 4 standard angulated views (8 s each), both 15 frames/s. ANOVA with Bonferroni correction was used for statistical analysis. Results. All drape designs led to substantial reductions in operator radiation exposure compared to control (P < 0.0001). The greatest decrease in radiation exposure (72%) was with the L-shaped design. Conclusions. Dedicated radial shielding drapes decrease radiation exposure to the operator by up to 72% during simulated cardiac catheterization. |
format | Online Article Text |
id | pubmed-3439952 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | International Scholarly Research Network |
record_format | MEDLINE/PubMed |
spelling | pubmed-34399522012-09-17 Radiation Dose Reduction during Radial Cardiac Catheterization: Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape Ertel, Andrew Nadelson, Jeffrey Shroff, Adhir R. Sweis, Ranya Ferrera, Dean Vidovich, Mladen I. ISRN Cardiol Research Article Objectives. Radiation scatter protection shield drapes have been designed with the goal of decreasing radiation dose to the operators during transfemoral catheterization. We sought to investigate the impact on operator radiation exposure of various shielding drapes specifically designed for the radial approach. Background. Radial access for cardiac catheterization has increased due to improved patient comfort and decreased bleeding complications. There are concerns for increased radiation exposure to patients and operators. Methods. Radiation doses to a simulated operator were measured with a RadCal Dosimeter in the cardiac catheterization laboratory. The mock patient was a 97.5 kg fission product phantom. Three lead-free drape designs were studied. The drapes were placed just proximal to the right wrist and extended medially to phantom's trunk. Simulated diagnostic coronary angiography included 6 minutes of fluoroscopy time and 32 seconds of cineangiography time at 4 standard angulated views (8 s each), both 15 frames/s. ANOVA with Bonferroni correction was used for statistical analysis. Results. All drape designs led to substantial reductions in operator radiation exposure compared to control (P < 0.0001). The greatest decrease in radiation exposure (72%) was with the L-shaped design. Conclusions. Dedicated radial shielding drapes decrease radiation exposure to the operator by up to 72% during simulated cardiac catheterization. International Scholarly Research Network 2012-09-04 /pmc/articles/PMC3439952/ /pubmed/22988525 http://dx.doi.org/10.5402/2012/769167 Text en Copyright © 2012 Andrew Ertel et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Ertel, Andrew Nadelson, Jeffrey Shroff, Adhir R. Sweis, Ranya Ferrera, Dean Vidovich, Mladen I. Radiation Dose Reduction during Radial Cardiac Catheterization: Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape |
title | Radiation Dose Reduction during Radial Cardiac Catheterization: Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape |
title_full | Radiation Dose Reduction during Radial Cardiac Catheterization: Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape |
title_fullStr | Radiation Dose Reduction during Radial Cardiac Catheterization: Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape |
title_full_unstemmed | Radiation Dose Reduction during Radial Cardiac Catheterization: Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape |
title_short | Radiation Dose Reduction during Radial Cardiac Catheterization: Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape |
title_sort | radiation dose reduction during radial cardiac catheterization: evaluation of a dedicated radial angiography absorption shielding drape |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3439952/ https://www.ncbi.nlm.nih.gov/pubmed/22988525 http://dx.doi.org/10.5402/2012/769167 |
work_keys_str_mv | AT ertelandrew radiationdosereductionduringradialcardiaccatheterizationevaluationofadedicatedradialangiographyabsorptionshieldingdrape AT nadelsonjeffrey radiationdosereductionduringradialcardiaccatheterizationevaluationofadedicatedradialangiographyabsorptionshieldingdrape AT shroffadhirr radiationdosereductionduringradialcardiaccatheterizationevaluationofadedicatedradialangiographyabsorptionshieldingdrape AT sweisranya radiationdosereductionduringradialcardiaccatheterizationevaluationofadedicatedradialangiographyabsorptionshieldingdrape AT ferreradean radiationdosereductionduringradialcardiaccatheterizationevaluationofadedicatedradialangiographyabsorptionshieldingdrape AT vidovichmladeni radiationdosereductionduringradialcardiaccatheterizationevaluationofadedicatedradialangiographyabsorptionshieldingdrape |