Comparison of Venous and Capillary Differential Leukocyte Counts Using a Standard Hematology Analyzer and a Novel Microfluidic Impedance Cytometer
Capillary blood sampling has been identified as a potentially suitable technique for use in diagnostic testing of the full blood count (FBC) at the point-of-care (POC), for which a recent need has been highlighted. In this study we assess the accuracy of capillary blood counts and evaluate the poten...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3447872/ https://www.ncbi.nlm.nih.gov/pubmed/23028467 http://dx.doi.org/10.1371/journal.pone.0043702 |
_version_ | 1782244183080173568 |
---|---|
author | Hollis, Veronica S. Holloway, Judith A. Harris, Scott Spencer, Daniel van Berkel, Cees Morgan, Hywel |
author_facet | Hollis, Veronica S. Holloway, Judith A. Harris, Scott Spencer, Daniel van Berkel, Cees Morgan, Hywel |
author_sort | Hollis, Veronica S. |
collection | PubMed |
description | Capillary blood sampling has been identified as a potentially suitable technique for use in diagnostic testing of the full blood count (FBC) at the point-of-care (POC), for which a recent need has been highlighted. In this study we assess the accuracy of capillary blood counts and evaluate the potential of a miniaturized cytometer developed for POC testing. Differential leukocyte counts in the normal clinical range from fingerprick (capillary) and venous blood samples were measured and compared using a standard hematology analyzer. The accuracy of our novel microfluidic impedance cytometer (MIC) was then tested by comparing same-site measurements to those obtained with the standard analyzer. The concordance between measurements of fingerprick and venous blood samples using the standard hematology analyzer was high, with no clinically relevant differences observed between the mean differential leukocyte counts. Concordance data between the MIC and the standard analyzer on same-site measurements presented significantly lower leukocyte counts determined by the MIC. This systematic undercount was consistent across the measured (normal) concentration range, suggesting that an internal correction factor could be applied. Differential leukocyte counts obtained from fingerprick samples accurately reflect those from venous blood, which confirms the potential of capillary blood sampling for POC testing of the FBC. Furthermore, the MIC device demonstrated here presents a realistic technology for the future development of FBC and related tests for use at the site of patient care. |
format | Online Article Text |
id | pubmed-3447872 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-34478722012-10-01 Comparison of Venous and Capillary Differential Leukocyte Counts Using a Standard Hematology Analyzer and a Novel Microfluidic Impedance Cytometer Hollis, Veronica S. Holloway, Judith A. Harris, Scott Spencer, Daniel van Berkel, Cees Morgan, Hywel PLoS One Research Article Capillary blood sampling has been identified as a potentially suitable technique for use in diagnostic testing of the full blood count (FBC) at the point-of-care (POC), for which a recent need has been highlighted. In this study we assess the accuracy of capillary blood counts and evaluate the potential of a miniaturized cytometer developed for POC testing. Differential leukocyte counts in the normal clinical range from fingerprick (capillary) and venous blood samples were measured and compared using a standard hematology analyzer. The accuracy of our novel microfluidic impedance cytometer (MIC) was then tested by comparing same-site measurements to those obtained with the standard analyzer. The concordance between measurements of fingerprick and venous blood samples using the standard hematology analyzer was high, with no clinically relevant differences observed between the mean differential leukocyte counts. Concordance data between the MIC and the standard analyzer on same-site measurements presented significantly lower leukocyte counts determined by the MIC. This systematic undercount was consistent across the measured (normal) concentration range, suggesting that an internal correction factor could be applied. Differential leukocyte counts obtained from fingerprick samples accurately reflect those from venous blood, which confirms the potential of capillary blood sampling for POC testing of the FBC. Furthermore, the MIC device demonstrated here presents a realistic technology for the future development of FBC and related tests for use at the site of patient care. Public Library of Science 2012-09-20 /pmc/articles/PMC3447872/ /pubmed/23028467 http://dx.doi.org/10.1371/journal.pone.0043702 Text en © 2012 Hollis et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Hollis, Veronica S. Holloway, Judith A. Harris, Scott Spencer, Daniel van Berkel, Cees Morgan, Hywel Comparison of Venous and Capillary Differential Leukocyte Counts Using a Standard Hematology Analyzer and a Novel Microfluidic Impedance Cytometer |
title | Comparison of Venous and Capillary Differential Leukocyte Counts Using a Standard Hematology Analyzer and a Novel Microfluidic Impedance Cytometer |
title_full | Comparison of Venous and Capillary Differential Leukocyte Counts Using a Standard Hematology Analyzer and a Novel Microfluidic Impedance Cytometer |
title_fullStr | Comparison of Venous and Capillary Differential Leukocyte Counts Using a Standard Hematology Analyzer and a Novel Microfluidic Impedance Cytometer |
title_full_unstemmed | Comparison of Venous and Capillary Differential Leukocyte Counts Using a Standard Hematology Analyzer and a Novel Microfluidic Impedance Cytometer |
title_short | Comparison of Venous and Capillary Differential Leukocyte Counts Using a Standard Hematology Analyzer and a Novel Microfluidic Impedance Cytometer |
title_sort | comparison of venous and capillary differential leukocyte counts using a standard hematology analyzer and a novel microfluidic impedance cytometer |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3447872/ https://www.ncbi.nlm.nih.gov/pubmed/23028467 http://dx.doi.org/10.1371/journal.pone.0043702 |
work_keys_str_mv | AT hollisveronicas comparisonofvenousandcapillarydifferentialleukocytecountsusingastandardhematologyanalyzerandanovelmicrofluidicimpedancecytometer AT hollowayjuditha comparisonofvenousandcapillarydifferentialleukocytecountsusingastandardhematologyanalyzerandanovelmicrofluidicimpedancecytometer AT harrisscott comparisonofvenousandcapillarydifferentialleukocytecountsusingastandardhematologyanalyzerandanovelmicrofluidicimpedancecytometer AT spencerdaniel comparisonofvenousandcapillarydifferentialleukocytecountsusingastandardhematologyanalyzerandanovelmicrofluidicimpedancecytometer AT vanberkelcees comparisonofvenousandcapillarydifferentialleukocytecountsusingastandardhematologyanalyzerandanovelmicrofluidicimpedancecytometer AT morganhywel comparisonofvenousandcapillarydifferentialleukocytecountsusingastandardhematologyanalyzerandanovelmicrofluidicimpedancecytometer |