A Comprehensive Assessment of the Precision and Agreement of Anterior Corneal Power Measurements Obtained Using 8 Different Devices
PURPOSE: To comprehensively assess the precision and agreement of anterior corneal power measurements using 8 different devices. METHODS: Thirty-five eyes from 35 healthy subjects were included in the prospective study. In the first session, a single examiner performed on each subject randomly measu...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3458095/ https://www.ncbi.nlm.nih.gov/pubmed/23049823 http://dx.doi.org/10.1371/journal.pone.0045607 |
_version_ | 1782244626846973952 |
---|---|
author | Wang, Qinmei Savini, Giacomo Hoffer, Kenneth J. Xu, Zhen Feng, Yifan Wen, Daizong Hua, Yanjun Yang, Feng Pan, Chao Huang, Jinhai |
author_facet | Wang, Qinmei Savini, Giacomo Hoffer, Kenneth J. Xu, Zhen Feng, Yifan Wen, Daizong Hua, Yanjun Yang, Feng Pan, Chao Huang, Jinhai |
author_sort | Wang, Qinmei |
collection | PubMed |
description | PURPOSE: To comprehensively assess the precision and agreement of anterior corneal power measurements using 8 different devices. METHODS: Thirty-five eyes from 35 healthy subjects were included in the prospective study. In the first session, a single examiner performed on each subject randomly measurements with the RC-5000 (Tomey Corp., Japan), KR-8000 (Topcon, Japan), IOLMaster (Carl Zeiss Meditec, Germany), E300 (Medmont International, Australia), Allegro Topolyzer (Wavelight AG, Germany), Vista (EyeSys, TX), Pentacam (Oculus, Germany) and Sirius (CSO, Italy). Measurements were repeated in the second session (1 to 2 weeks later). Repeatability and reproducibility of corneal power measurements were assessed based on the intrasession and intersession within-subject standard deviation (Sw), repeatability (2.77Sw), coefficient of variation (COV), and intraclass correlation coefficient (ICC). Agreement was evaluated by 95% limits of agreement (LoA). RESULTS: All devices demonstrated high repeatability and reproducibility of the keratometric values (2.77Sw<0.36D, COV<0.3%, ICC>0.98). Repeated-measures analysis of variance with Bonferroni post test showed statistically significant differences (P<0.01) among mean keratometric values of most instruments; the largest differences were observed between the EyeSys Vista and Medmont E300. Good agreement (i.e., 95%LoA within ±0.5D) was found between most instruments for flat, steep and mean keratometry, except for EyeSys and Medmont. Repeatability and reproducibility of vectors J(0) and J(45) was good, as the ICCs were higher than 0.9, except J(45) of Medmont and Pentacam. For the 95% LoAs of J(0) and J(45), they were all ≤ ±0.31 among any two paired devices. CONCLUSIONS: The 8 devices showed excellent repeatability and reproducibility. The results obtained using the RC-5000, KR-8000, IOLMaster, Allegro Topolyzer, Pentacam and Sirius were comparable, suggesting that they could be used interchangeably in most clinical settings. Caution is warranted with the measurements of the EyeSys Vista and Medmont E300, which should not be used interchangeably with other devices due to lower agreement. TRIAL REGISTRATION: ClinicalTrials.gov NCT01587287 |
format | Online Article Text |
id | pubmed-3458095 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-34580952012-10-03 A Comprehensive Assessment of the Precision and Agreement of Anterior Corneal Power Measurements Obtained Using 8 Different Devices Wang, Qinmei Savini, Giacomo Hoffer, Kenneth J. Xu, Zhen Feng, Yifan Wen, Daizong Hua, Yanjun Yang, Feng Pan, Chao Huang, Jinhai PLoS One Research Article PURPOSE: To comprehensively assess the precision and agreement of anterior corneal power measurements using 8 different devices. METHODS: Thirty-five eyes from 35 healthy subjects were included in the prospective study. In the first session, a single examiner performed on each subject randomly measurements with the RC-5000 (Tomey Corp., Japan), KR-8000 (Topcon, Japan), IOLMaster (Carl Zeiss Meditec, Germany), E300 (Medmont International, Australia), Allegro Topolyzer (Wavelight AG, Germany), Vista (EyeSys, TX), Pentacam (Oculus, Germany) and Sirius (CSO, Italy). Measurements were repeated in the second session (1 to 2 weeks later). Repeatability and reproducibility of corneal power measurements were assessed based on the intrasession and intersession within-subject standard deviation (Sw), repeatability (2.77Sw), coefficient of variation (COV), and intraclass correlation coefficient (ICC). Agreement was evaluated by 95% limits of agreement (LoA). RESULTS: All devices demonstrated high repeatability and reproducibility of the keratometric values (2.77Sw<0.36D, COV<0.3%, ICC>0.98). Repeated-measures analysis of variance with Bonferroni post test showed statistically significant differences (P<0.01) among mean keratometric values of most instruments; the largest differences were observed between the EyeSys Vista and Medmont E300. Good agreement (i.e., 95%LoA within ±0.5D) was found between most instruments for flat, steep and mean keratometry, except for EyeSys and Medmont. Repeatability and reproducibility of vectors J(0) and J(45) was good, as the ICCs were higher than 0.9, except J(45) of Medmont and Pentacam. For the 95% LoAs of J(0) and J(45), they were all ≤ ±0.31 among any two paired devices. CONCLUSIONS: The 8 devices showed excellent repeatability and reproducibility. The results obtained using the RC-5000, KR-8000, IOLMaster, Allegro Topolyzer, Pentacam and Sirius were comparable, suggesting that they could be used interchangeably in most clinical settings. Caution is warranted with the measurements of the EyeSys Vista and Medmont E300, which should not be used interchangeably with other devices due to lower agreement. TRIAL REGISTRATION: ClinicalTrials.gov NCT01587287 Public Library of Science 2012-09-25 /pmc/articles/PMC3458095/ /pubmed/23049823 http://dx.doi.org/10.1371/journal.pone.0045607 Text en © 2012 Wang et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Wang, Qinmei Savini, Giacomo Hoffer, Kenneth J. Xu, Zhen Feng, Yifan Wen, Daizong Hua, Yanjun Yang, Feng Pan, Chao Huang, Jinhai A Comprehensive Assessment of the Precision and Agreement of Anterior Corneal Power Measurements Obtained Using 8 Different Devices |
title | A Comprehensive Assessment of the Precision and Agreement of Anterior Corneal Power Measurements Obtained Using 8 Different Devices |
title_full | A Comprehensive Assessment of the Precision and Agreement of Anterior Corneal Power Measurements Obtained Using 8 Different Devices |
title_fullStr | A Comprehensive Assessment of the Precision and Agreement of Anterior Corneal Power Measurements Obtained Using 8 Different Devices |
title_full_unstemmed | A Comprehensive Assessment of the Precision and Agreement of Anterior Corneal Power Measurements Obtained Using 8 Different Devices |
title_short | A Comprehensive Assessment of the Precision and Agreement of Anterior Corneal Power Measurements Obtained Using 8 Different Devices |
title_sort | comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3458095/ https://www.ncbi.nlm.nih.gov/pubmed/23049823 http://dx.doi.org/10.1371/journal.pone.0045607 |
work_keys_str_mv | AT wangqinmei acomprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT savinigiacomo acomprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT hofferkennethj acomprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT xuzhen acomprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT fengyifan acomprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT wendaizong acomprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT huayanjun acomprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT yangfeng acomprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT panchao acomprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT huangjinhai acomprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT wangqinmei comprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT savinigiacomo comprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT hofferkennethj comprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT xuzhen comprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT fengyifan comprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT wendaizong comprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT huayanjun comprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT yangfeng comprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT panchao comprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices AT huangjinhai comprehensiveassessmentoftheprecisionandagreementofanteriorcornealpowermeasurementsobtainedusing8differentdevices |