A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves

Multidetector-row CT is promising for prosthetic heart valve (PHV) assessment but retrospectively ECG-gated scanning has a considerable radiation dose. Recently introduced iterative reconstruction (IR) algorithms may enable radiation dose reduction with retained image quality. Furthermore, PHV image...

Descripción completa

Detalles Bibliográficos
Autores principales: Habets, Jesse, Symersky, Petr, de Mol, Bas A. J. M., Mali, Willem P. Th. M., Leiner, Tim, Budde, Ricardo P. J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Netherlands 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3463798/
https://www.ncbi.nlm.nih.gov/pubmed/22002686
http://dx.doi.org/10.1007/s10554-011-9954-7
_version_ 1782245327052472320
author Habets, Jesse
Symersky, Petr
de Mol, Bas A. J. M.
Mali, Willem P. Th. M.
Leiner, Tim
Budde, Ricardo P. J.
author_facet Habets, Jesse
Symersky, Petr
de Mol, Bas A. J. M.
Mali, Willem P. Th. M.
Leiner, Tim
Budde, Ricardo P. J.
author_sort Habets, Jesse
collection PubMed
description Multidetector-row CT is promising for prosthetic heart valve (PHV) assessment but retrospectively ECG-gated scanning has a considerable radiation dose. Recently introduced iterative reconstruction (IR) algorithms may enable radiation dose reduction with retained image quality. Furthermore, PHV image quality on the CT scan mainly depends on extent of PHV artifacts. IR may decrease streak artifacts. We compared image noise and artifact volumes in scans of mechanical PHVs reconstructed with conventional filtered back projection (FBP) to lower dose scans reconstructed with IR. Four different PHVs (St. Jude, Carbomedics, ON-X and Medtronic Hall) were scanned in a pulsatile in vitro model. Ten retrospectively ECG-gated CT scans were performed of each PHV at 120 kV, 600 mAs (high-dose CTDI(vol) 35.3 mGy) and 120 kV, 300 mAs (low-dose CTDI(vol) 17.7 mGy) on a 64 detector-row scanner. Diastolic and systolic images were reconstructed with FBP (high and low-dose) and the IR algorithm (low-dose only). Hypo- and hyperdense artifact volumes were determined using two threshold filters. Image noise was measured. Mean hypo- and hyperdense artifact volumes (mm(3)) were 1,235/5,346 (high-dose FBP); 2,405/6,877 (low-dose FBP) and 1,218/5,333 (low-dose IR). Low-dose IR reconstructions had similar image noise compared to high-dose FBP (16.5 ± 1.7 vs. 16.3 ± 1.6, mean ± SD, respectively, P = 1.0). IR allows ECG-gated PHV imaging with similar image noise and PHV artifacts at 50% less dose compared to conventional FBP in an pulsatile in vitro model.
format Online
Article
Text
id pubmed-3463798
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher Springer Netherlands
record_format MEDLINE/PubMed
spelling pubmed-34637982012-10-04 A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves Habets, Jesse Symersky, Petr de Mol, Bas A. J. M. Mali, Willem P. Th. M. Leiner, Tim Budde, Ricardo P. J. Int J Cardiovasc Imaging Original Paper Multidetector-row CT is promising for prosthetic heart valve (PHV) assessment but retrospectively ECG-gated scanning has a considerable radiation dose. Recently introduced iterative reconstruction (IR) algorithms may enable radiation dose reduction with retained image quality. Furthermore, PHV image quality on the CT scan mainly depends on extent of PHV artifacts. IR may decrease streak artifacts. We compared image noise and artifact volumes in scans of mechanical PHVs reconstructed with conventional filtered back projection (FBP) to lower dose scans reconstructed with IR. Four different PHVs (St. Jude, Carbomedics, ON-X and Medtronic Hall) were scanned in a pulsatile in vitro model. Ten retrospectively ECG-gated CT scans were performed of each PHV at 120 kV, 600 mAs (high-dose CTDI(vol) 35.3 mGy) and 120 kV, 300 mAs (low-dose CTDI(vol) 17.7 mGy) on a 64 detector-row scanner. Diastolic and systolic images were reconstructed with FBP (high and low-dose) and the IR algorithm (low-dose only). Hypo- and hyperdense artifact volumes were determined using two threshold filters. Image noise was measured. Mean hypo- and hyperdense artifact volumes (mm(3)) were 1,235/5,346 (high-dose FBP); 2,405/6,877 (low-dose FBP) and 1,218/5,333 (low-dose IR). Low-dose IR reconstructions had similar image noise compared to high-dose FBP (16.5 ± 1.7 vs. 16.3 ± 1.6, mean ± SD, respectively, P = 1.0). IR allows ECG-gated PHV imaging with similar image noise and PHV artifacts at 50% less dose compared to conventional FBP in an pulsatile in vitro model. Springer Netherlands 2011-10-15 2012 /pmc/articles/PMC3463798/ /pubmed/22002686 http://dx.doi.org/10.1007/s10554-011-9954-7 Text en © The Author(s) 2011 https://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
spellingShingle Original Paper
Habets, Jesse
Symersky, Petr
de Mol, Bas A. J. M.
Mali, Willem P. Th. M.
Leiner, Tim
Budde, Ricardo P. J.
A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves
title A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves
title_full A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves
title_fullStr A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves
title_full_unstemmed A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves
title_short A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves
title_sort novel iterative reconstruction algorithm allows reduced dose multidetector-row ct imaging of mechanical prosthetic heart valves
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3463798/
https://www.ncbi.nlm.nih.gov/pubmed/22002686
http://dx.doi.org/10.1007/s10554-011-9954-7
work_keys_str_mv AT habetsjesse anoveliterativereconstructionalgorithmallowsreduceddosemultidetectorrowctimagingofmechanicalprostheticheartvalves
AT symerskypetr anoveliterativereconstructionalgorithmallowsreduceddosemultidetectorrowctimagingofmechanicalprostheticheartvalves
AT demolbasajm anoveliterativereconstructionalgorithmallowsreduceddosemultidetectorrowctimagingofmechanicalprostheticheartvalves
AT maliwillempthm anoveliterativereconstructionalgorithmallowsreduceddosemultidetectorrowctimagingofmechanicalprostheticheartvalves
AT leinertim anoveliterativereconstructionalgorithmallowsreduceddosemultidetectorrowctimagingofmechanicalprostheticheartvalves
AT buddericardopj anoveliterativereconstructionalgorithmallowsreduceddosemultidetectorrowctimagingofmechanicalprostheticheartvalves
AT habetsjesse noveliterativereconstructionalgorithmallowsreduceddosemultidetectorrowctimagingofmechanicalprostheticheartvalves
AT symerskypetr noveliterativereconstructionalgorithmallowsreduceddosemultidetectorrowctimagingofmechanicalprostheticheartvalves
AT demolbasajm noveliterativereconstructionalgorithmallowsreduceddosemultidetectorrowctimagingofmechanicalprostheticheartvalves
AT maliwillempthm noveliterativereconstructionalgorithmallowsreduceddosemultidetectorrowctimagingofmechanicalprostheticheartvalves
AT leinertim noveliterativereconstructionalgorithmallowsreduceddosemultidetectorrowctimagingofmechanicalprostheticheartvalves
AT buddericardopj noveliterativereconstructionalgorithmallowsreduceddosemultidetectorrowctimagingofmechanicalprostheticheartvalves