Hyperglycemia-Induced T-Wave Oversensing as a Cause of Cardiac Resynchronization Therapy (CRT) Failure
T-wave oversensing occurs when the counter starts giving dual beeps for every cardiac cycle instead of one. This usually happens when the monitoring lead displays a tall T wave, which is also sharp. R wave sensing algorithms of the devices do not sense T wave because the slow rate of the T wave is m...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Tehran University of Medical Sciences
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3466884/ https://www.ncbi.nlm.nih.gov/pubmed/23074632 |
_version_ | 1782245721794150400 |
---|---|
author | Assadian Rad, Mohammad Emkanjoo, Zahra Moladoust, Hassan |
author_facet | Assadian Rad, Mohammad Emkanjoo, Zahra Moladoust, Hassan |
author_sort | Assadian Rad, Mohammad |
collection | PubMed |
description | T-wave oversensing occurs when the counter starts giving dual beeps for every cardiac cycle instead of one. This usually happens when the monitoring lead displays a tall T wave, which is also sharp. R wave sensing algorithms of the devices do not sense T wave because the slow rate of the T wave is much less than that of the R wave. But the slow rate of T waves may change with time and also because of parameters like potassium levels and hyperglycemia. We present a 67-year-old female who underwent the implantation of cardiac resynchronization therapy (cardiac resynchronization and implantable cardioverter defibrilator [CRT-D]) because of severe left ventricular systolic dysfunction and ventricular dyssynchrony experienced recurrent inappropriate implantable cardioverter-defibrillator (ICD) shocks and CRT failure. Device analysis showed that the CRT failure was in consequence of T-wave oversensing due to hyperglycemia. Elimination of the T-wave oversensing after hyperglycemia control conferred good biventricular pacing and good response to CRT during a 6-month follow-up period. |
format | Online Article Text |
id | pubmed-3466884 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | Tehran University of Medical Sciences |
record_format | MEDLINE/PubMed |
spelling | pubmed-34668842012-10-16 Hyperglycemia-Induced T-Wave Oversensing as a Cause of Cardiac Resynchronization Therapy (CRT) Failure Assadian Rad, Mohammad Emkanjoo, Zahra Moladoust, Hassan J Tehran Heart Cent Case Report T-wave oversensing occurs when the counter starts giving dual beeps for every cardiac cycle instead of one. This usually happens when the monitoring lead displays a tall T wave, which is also sharp. R wave sensing algorithms of the devices do not sense T wave because the slow rate of the T wave is much less than that of the R wave. But the slow rate of T waves may change with time and also because of parameters like potassium levels and hyperglycemia. We present a 67-year-old female who underwent the implantation of cardiac resynchronization therapy (cardiac resynchronization and implantable cardioverter defibrilator [CRT-D]) because of severe left ventricular systolic dysfunction and ventricular dyssynchrony experienced recurrent inappropriate implantable cardioverter-defibrillator (ICD) shocks and CRT failure. Device analysis showed that the CRT failure was in consequence of T-wave oversensing due to hyperglycemia. Elimination of the T-wave oversensing after hyperglycemia control conferred good biventricular pacing and good response to CRT during a 6-month follow-up period. Tehran University of Medical Sciences 2012-02-28 2012 /pmc/articles/PMC3466884/ /pubmed/23074632 Text en Copyright © Tehran Heart Center, Tehran University of Medical Sciences http://creativecommons.org/licenses/by-nc/3.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution NonCommercial 3.0 License (CC BY-NC 3.0), which allows users to read, copy, distribute and make derivative works for non-commercial purposes from the material, as long as the author of the original work is cited properly. |
spellingShingle | Case Report Assadian Rad, Mohammad Emkanjoo, Zahra Moladoust, Hassan Hyperglycemia-Induced T-Wave Oversensing as a Cause of Cardiac Resynchronization Therapy (CRT) Failure |
title | Hyperglycemia-Induced T-Wave Oversensing as a Cause of Cardiac Resynchronization Therapy (CRT) Failure |
title_full | Hyperglycemia-Induced T-Wave Oversensing as a Cause of Cardiac Resynchronization Therapy (CRT) Failure |
title_fullStr | Hyperglycemia-Induced T-Wave Oversensing as a Cause of Cardiac Resynchronization Therapy (CRT) Failure |
title_full_unstemmed | Hyperglycemia-Induced T-Wave Oversensing as a Cause of Cardiac Resynchronization Therapy (CRT) Failure |
title_short | Hyperglycemia-Induced T-Wave Oversensing as a Cause of Cardiac Resynchronization Therapy (CRT) Failure |
title_sort | hyperglycemia-induced t-wave oversensing as a cause of cardiac resynchronization therapy (crt) failure |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3466884/ https://www.ncbi.nlm.nih.gov/pubmed/23074632 |
work_keys_str_mv | AT assadianradmohammad hyperglycemiainducedtwaveoversensingasacauseofcardiacresynchronizationtherapycrtfailure AT emkanjoozahra hyperglycemiainducedtwaveoversensingasacauseofcardiacresynchronizationtherapycrtfailure AT moladousthassan hyperglycemiainducedtwaveoversensingasacauseofcardiacresynchronizationtherapycrtfailure |