The usefulness of an independent patient-specific treatment planning verification method using a benchmark plan in high-dose-rate intracavitary brachytherapy for carcinoma of the uterine cervix
To develop an easy independent patient-specific quality assurance (QA) method using a benchmark plan for high-dose-rate intracavitary brachytherapy for cervix cancer, we conducted benchmark treatment planning with various sizes and combinations of tandem-ovoid and tandem-cylinder applications with ‘...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3483840/ https://www.ncbi.nlm.nih.gov/pubmed/22843371 http://dx.doi.org/10.1093/jrr/rrs035 |
Sumario: | To develop an easy independent patient-specific quality assurance (QA) method using a benchmark plan for high-dose-rate intracavitary brachytherapy for cervix cancer, we conducted benchmark treatment planning with various sizes and combinations of tandem-ovoid and tandem-cylinder applications with ‘ideal’ geometry outside the patient. Two-dimensional-based treatment planning was conducted based on the Manchester method. We predicted the total dwell time of individual treatment plans from the air kerma strength, total dwell time and prescription dose of the benchmark plan. In addition, we recorded the height (dh), width (dw) and thickness (dt) covered with 100% isodose line. These parameters were compared with 169 and 29 clinical cases for tandem-ovoid or tandem-cylinder cases, respectively. With regard to tandem-ovoid cases, differences in total dwell time, dh, dt and dw between benchmark and individual plans were on average –0.2% ± 3.8%, –1.0 mm ± 2.6 mm, 0.8 mm ± 1.3 mm and –0.1 mm ± 1.5 mm, respectively. With regard to tandem-cylinder cases, differences in total dwell time, dh(front) (the distance from tandem tip to tandem ring), dt and dw between benchmark and individual plans were on average –1.5% ± 3.1%, –1.5 mm ± 4.9 mm, 0.1 mm ± 1.0 mm and 0.2 mm ± 0.8 mm, respectively. Of two cases, more than 13% differences in total dwell time were observed between benchmark plans and the clinical cases, which turned out to be due to the use of the wrong source position setting. These results suggest that our method is easy and useful for independent verification of patient-specific treatment planning QA. |
---|