Evaluation of beam hardening and photon scatter by brass compensator for IMRT

When a brass compensator is set in a treatment beam, beam hardening may take place. This variation of the energy spectrum may affect the accuracy of dose calculation by a treatment planning system and the results of dose measurement of brass compensator intensity modulated radiation therapy (IMRT)....

Descripción completa

Detalles Bibliográficos
Autores principales: Hashimoto, Shimpei, Karasawa, Katsuyuki, Fujita, Yukio, Miyashita, Hisayuki, Chang, Weishan, Kawachi, Toru, Katayose, Tetsurou, Kobayashi, Nao, Kunieda, Etsuo, Saitoh, Hidetoshi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3483849/
https://www.ncbi.nlm.nih.gov/pubmed/22915779
http://dx.doi.org/10.1093/jrr/rrs048
_version_ 1782248074640359424
author Hashimoto, Shimpei
Karasawa, Katsuyuki
Fujita, Yukio
Miyashita, Hisayuki
Chang, Weishan
Kawachi, Toru
Katayose, Tetsurou
Kobayashi, Nao
Kunieda, Etsuo
Saitoh, Hidetoshi
author_facet Hashimoto, Shimpei
Karasawa, Katsuyuki
Fujita, Yukio
Miyashita, Hisayuki
Chang, Weishan
Kawachi, Toru
Katayose, Tetsurou
Kobayashi, Nao
Kunieda, Etsuo
Saitoh, Hidetoshi
author_sort Hashimoto, Shimpei
collection PubMed
description When a brass compensator is set in a treatment beam, beam hardening may take place. This variation of the energy spectrum may affect the accuracy of dose calculation by a treatment planning system and the results of dose measurement of brass compensator intensity modulated radiation therapy (IMRT). In addition, when X-rays pass the compensator, scattered photons are generated within the compensator. Scattered photons may affect the monitor unit (MU) calculation. In this study, to evaluate the variation of dose distribution by the compensator, dose distribution was measured and energy spectrum was simulated using the Monte Carlo method. To investigate the influence of beam hardening for dose measurement using an ionization chamber, the beam quality correction factor was determined. Moreover, to clarify the effect of scattered photons generated within the compensator for the MU calculation, the head scatter factor was measured and energy spectrum analyses were performed. As a result, when X-rays passed the brass compensator, beam hardening occurred and dose distribution was varied. The variation of dose distribution and energy spectrum was larger with decreasing field size. This means that energy spectrum should be reproduced correctly to obtain high accuracy of dose calculation for the compensator IMRT. On the other hand, the influence of beam hardening on k(Q) was insignificant. Furthermore, scattered photons were generated within the compensator, and scattered photons affect the head scatter factor. These results show that scattered photons must be taken into account for MU calculation for brass compensator IMRT.
format Online
Article
Text
id pubmed-3483849
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-34838492012-10-30 Evaluation of beam hardening and photon scatter by brass compensator for IMRT Hashimoto, Shimpei Karasawa, Katsuyuki Fujita, Yukio Miyashita, Hisayuki Chang, Weishan Kawachi, Toru Katayose, Tetsurou Kobayashi, Nao Kunieda, Etsuo Saitoh, Hidetoshi J Radiat Res Technology When a brass compensator is set in a treatment beam, beam hardening may take place. This variation of the energy spectrum may affect the accuracy of dose calculation by a treatment planning system and the results of dose measurement of brass compensator intensity modulated radiation therapy (IMRT). In addition, when X-rays pass the compensator, scattered photons are generated within the compensator. Scattered photons may affect the monitor unit (MU) calculation. In this study, to evaluate the variation of dose distribution by the compensator, dose distribution was measured and energy spectrum was simulated using the Monte Carlo method. To investigate the influence of beam hardening for dose measurement using an ionization chamber, the beam quality correction factor was determined. Moreover, to clarify the effect of scattered photons generated within the compensator for the MU calculation, the head scatter factor was measured and energy spectrum analyses were performed. As a result, when X-rays passed the brass compensator, beam hardening occurred and dose distribution was varied. The variation of dose distribution and energy spectrum was larger with decreasing field size. This means that energy spectrum should be reproduced correctly to obtain high accuracy of dose calculation for the compensator IMRT. On the other hand, the influence of beam hardening on k(Q) was insignificant. Furthermore, scattered photons were generated within the compensator, and scattered photons affect the head scatter factor. These results show that scattered photons must be taken into account for MU calculation for brass compensator IMRT. Oxford University Press 2012-11 2012-08-21 /pmc/articles/PMC3483849/ /pubmed/22915779 http://dx.doi.org/10.1093/jrr/rrs048 Text en © The Author 2012. Published by Oxford University Press on behalf of The Japan Radiation Research Society and Japanese Society for Therapeutic Radiology and Oncology. http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/), which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technology
Hashimoto, Shimpei
Karasawa, Katsuyuki
Fujita, Yukio
Miyashita, Hisayuki
Chang, Weishan
Kawachi, Toru
Katayose, Tetsurou
Kobayashi, Nao
Kunieda, Etsuo
Saitoh, Hidetoshi
Evaluation of beam hardening and photon scatter by brass compensator for IMRT
title Evaluation of beam hardening and photon scatter by brass compensator for IMRT
title_full Evaluation of beam hardening and photon scatter by brass compensator for IMRT
title_fullStr Evaluation of beam hardening and photon scatter by brass compensator for IMRT
title_full_unstemmed Evaluation of beam hardening and photon scatter by brass compensator for IMRT
title_short Evaluation of beam hardening and photon scatter by brass compensator for IMRT
title_sort evaluation of beam hardening and photon scatter by brass compensator for imrt
topic Technology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3483849/
https://www.ncbi.nlm.nih.gov/pubmed/22915779
http://dx.doi.org/10.1093/jrr/rrs048
work_keys_str_mv AT hashimotoshimpei evaluationofbeamhardeningandphotonscatterbybrasscompensatorforimrt
AT karasawakatsuyuki evaluationofbeamhardeningandphotonscatterbybrasscompensatorforimrt
AT fujitayukio evaluationofbeamhardeningandphotonscatterbybrasscompensatorforimrt
AT miyashitahisayuki evaluationofbeamhardeningandphotonscatterbybrasscompensatorforimrt
AT changweishan evaluationofbeamhardeningandphotonscatterbybrasscompensatorforimrt
AT kawachitoru evaluationofbeamhardeningandphotonscatterbybrasscompensatorforimrt
AT katayosetetsurou evaluationofbeamhardeningandphotonscatterbybrasscompensatorforimrt
AT kobayashinao evaluationofbeamhardeningandphotonscatterbybrasscompensatorforimrt
AT kuniedaetsuo evaluationofbeamhardeningandphotonscatterbybrasscompensatorforimrt
AT saitohhidetoshi evaluationofbeamhardeningandphotonscatterbybrasscompensatorforimrt