Experimental Evaluation of Accelerated T1rho Relaxation Quantification in Human Liver Using Limited Spin-Lock Times

OBJECTIVE: It was reported lately that to obtain consistent liver T1rho measurement, at 3T MRI using six spin-lock times (SLTs), is feasible. In this study, the feasibility of using three or two SLT points to measure liver T1rho relaxation time was explored. MATERIALS AND METHODS: Seventeen healthy...

Descripción completa

Detalles Bibliográficos
Autores principales: Zhao, Feng, Deng, Min, Yuan, Jing, Teng, Gao-Jun, Ahuja, Anil T, Wang, Yi-Xiang J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Society of Radiology 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3484294/
https://www.ncbi.nlm.nih.gov/pubmed/23118572
http://dx.doi.org/10.3348/kjr.2012.13.6.736
_version_ 1782248127543115776
author Zhao, Feng
Deng, Min
Yuan, Jing
Teng, Gao-Jun
Ahuja, Anil T
Wang, Yi-Xiang J.
author_facet Zhao, Feng
Deng, Min
Yuan, Jing
Teng, Gao-Jun
Ahuja, Anil T
Wang, Yi-Xiang J.
author_sort Zhao, Feng
collection PubMed
description OBJECTIVE: It was reported lately that to obtain consistent liver T1rho measurement, at 3T MRI using six spin-lock times (SLTs), is feasible. In this study, the feasibility of using three or two SLT points to measure liver T1rho relaxation time was explored. MATERIALS AND METHODS: Seventeen healthy volunteers underwent 36 examinations. Three representative axial slices were selected to cut through the upper, middle, and lower liver. A rotary echo spin-lock pulse was implemented in a 2D fast field echo sequence. Spin-lock frequency was 500 Hz and the spin-lock times of 1, 10, 20, 30, 40, and 50 milliseconds (ms) were used for T1rho mapping. T1rho maps were constructed by using all 6 SLT points, three SLT points of 1, 20, and 50 ms, or two SLTs of 1 and 50 ms, respectively. Intra-class correlation coefficient (ICC) and Bland and Altman plot were used to assess the measurement agreement. RESULTS: Two examinations were excluded, due to motion artifact at the SLT of 50 ms. With the remaining 34 examinations, the ICC for 6-SLT vs. 3-SLT T1rho measurements was 0.922, while the ICC for 6-SLT vs. 2-SLT T1rho measurement was 0.756. The Bland and Altman analysis showed a mean difference of 0.19 (95% limits of agreement: -1.34, 1.73) for 6-SLT vs. 3-SLT T1rho measurement, and the mean difference of 0.89 (95% limits of agreement: -1.67, 3.45) for 6-SLT vs. 2-SLT T1rho measurement. The scan re-scan reproducibility ICC (n = 11 subjects) was 0.755, 0.727, and 0.528 for 6-SLT measurement, 3-SLT measurement, and 2-SLT measurement, respectively. CONCLUSION: Adopting 3 SLTs of 1, 20, and 50 ms can be an acceptable alternative for the liver T1rho measurement, while 2 SLTs of 1 and 50 ms do not provide reliable measurement.
format Online
Article
Text
id pubmed-3484294
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher The Korean Society of Radiology
record_format MEDLINE/PubMed
spelling pubmed-34842942012-11-02 Experimental Evaluation of Accelerated T1rho Relaxation Quantification in Human Liver Using Limited Spin-Lock Times Zhao, Feng Deng, Min Yuan, Jing Teng, Gao-Jun Ahuja, Anil T Wang, Yi-Xiang J. Korean J Radiol Original Article OBJECTIVE: It was reported lately that to obtain consistent liver T1rho measurement, at 3T MRI using six spin-lock times (SLTs), is feasible. In this study, the feasibility of using three or two SLT points to measure liver T1rho relaxation time was explored. MATERIALS AND METHODS: Seventeen healthy volunteers underwent 36 examinations. Three representative axial slices were selected to cut through the upper, middle, and lower liver. A rotary echo spin-lock pulse was implemented in a 2D fast field echo sequence. Spin-lock frequency was 500 Hz and the spin-lock times of 1, 10, 20, 30, 40, and 50 milliseconds (ms) were used for T1rho mapping. T1rho maps were constructed by using all 6 SLT points, three SLT points of 1, 20, and 50 ms, or two SLTs of 1 and 50 ms, respectively. Intra-class correlation coefficient (ICC) and Bland and Altman plot were used to assess the measurement agreement. RESULTS: Two examinations were excluded, due to motion artifact at the SLT of 50 ms. With the remaining 34 examinations, the ICC for 6-SLT vs. 3-SLT T1rho measurements was 0.922, while the ICC for 6-SLT vs. 2-SLT T1rho measurement was 0.756. The Bland and Altman analysis showed a mean difference of 0.19 (95% limits of agreement: -1.34, 1.73) for 6-SLT vs. 3-SLT T1rho measurement, and the mean difference of 0.89 (95% limits of agreement: -1.67, 3.45) for 6-SLT vs. 2-SLT T1rho measurement. The scan re-scan reproducibility ICC (n = 11 subjects) was 0.755, 0.727, and 0.528 for 6-SLT measurement, 3-SLT measurement, and 2-SLT measurement, respectively. CONCLUSION: Adopting 3 SLTs of 1, 20, and 50 ms can be an acceptable alternative for the liver T1rho measurement, while 2 SLTs of 1 and 50 ms do not provide reliable measurement. The Korean Society of Radiology 2012 2012-10-12 /pmc/articles/PMC3484294/ /pubmed/23118572 http://dx.doi.org/10.3348/kjr.2012.13.6.736 Text en Copyright © 2012 The Korean Society of Radiology http://creativecommons.org/licenses/by-nc/3.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Zhao, Feng
Deng, Min
Yuan, Jing
Teng, Gao-Jun
Ahuja, Anil T
Wang, Yi-Xiang J.
Experimental Evaluation of Accelerated T1rho Relaxation Quantification in Human Liver Using Limited Spin-Lock Times
title Experimental Evaluation of Accelerated T1rho Relaxation Quantification in Human Liver Using Limited Spin-Lock Times
title_full Experimental Evaluation of Accelerated T1rho Relaxation Quantification in Human Liver Using Limited Spin-Lock Times
title_fullStr Experimental Evaluation of Accelerated T1rho Relaxation Quantification in Human Liver Using Limited Spin-Lock Times
title_full_unstemmed Experimental Evaluation of Accelerated T1rho Relaxation Quantification in Human Liver Using Limited Spin-Lock Times
title_short Experimental Evaluation of Accelerated T1rho Relaxation Quantification in Human Liver Using Limited Spin-Lock Times
title_sort experimental evaluation of accelerated t1rho relaxation quantification in human liver using limited spin-lock times
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3484294/
https://www.ncbi.nlm.nih.gov/pubmed/23118572
http://dx.doi.org/10.3348/kjr.2012.13.6.736
work_keys_str_mv AT zhaofeng experimentalevaluationofacceleratedt1rhorelaxationquantificationinhumanliverusinglimitedspinlocktimes
AT dengmin experimentalevaluationofacceleratedt1rhorelaxationquantificationinhumanliverusinglimitedspinlocktimes
AT yuanjing experimentalevaluationofacceleratedt1rhorelaxationquantificationinhumanliverusinglimitedspinlocktimes
AT tenggaojun experimentalevaluationofacceleratedt1rhorelaxationquantificationinhumanliverusinglimitedspinlocktimes
AT ahujaanilt experimentalevaluationofacceleratedt1rhorelaxationquantificationinhumanliverusinglimitedspinlocktimes
AT wangyixiangj experimentalevaluationofacceleratedt1rhorelaxationquantificationinhumanliverusinglimitedspinlocktimes