Cargando…

Transcatheter interruption of large residual flow after device closure of “Type A” patent ductus arteriosus

We report a case of 3-year-old girl who had persistence of large residual flow following transcatheter closure of a 6 mm ‘Type A’ patent ductus arteriosus using a 12 × 10 mm duct occluder. Angiography revealed a large left-to-right shunt coursing through and exiting around the implanted device. Near...

Descripción completa

Detalles Bibliográficos
Autores principales: Sridhar, Anuradha, Subramanyan, Raghavan, Rajasekar, Premsekar
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3487218/
https://www.ncbi.nlm.nih.gov/pubmed/23129919
http://dx.doi.org/10.4103/0974-2069.99632
_version_ 1782248447748866048
author Sridhar, Anuradha
Subramanyan, Raghavan
Rajasekar, Premsekar
author_facet Sridhar, Anuradha
Subramanyan, Raghavan
Rajasekar, Premsekar
author_sort Sridhar, Anuradha
collection PubMed
description We report a case of 3-year-old girl who had persistence of large residual flow following transcatheter closure of a 6 mm ‘Type A’ patent ductus arteriosus using a 12 × 10 mm duct occluder. Angiography revealed a large left-to-right shunt coursing through and exiting around the implanted device. Near total abolition of the residual shunt was achieved by initial implantation of an embolization coil within the duct occluder and subsequently an Amplatzer duct occluder (ADO II) adjacent to the duct occluder. This challenging case describes an additional technique of abolishing a large residual flow in and around a Nitinol duct occluder device.
format Online
Article
Text
id pubmed-3487218
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-34872182012-11-05 Transcatheter interruption of large residual flow after device closure of “Type A” patent ductus arteriosus Sridhar, Anuradha Subramanyan, Raghavan Rajasekar, Premsekar Ann Pediatr Cardiol Case Report We report a case of 3-year-old girl who had persistence of large residual flow following transcatheter closure of a 6 mm ‘Type A’ patent ductus arteriosus using a 12 × 10 mm duct occluder. Angiography revealed a large left-to-right shunt coursing through and exiting around the implanted device. Near total abolition of the residual shunt was achieved by initial implantation of an embolization coil within the duct occluder and subsequently an Amplatzer duct occluder (ADO II) adjacent to the duct occluder. This challenging case describes an additional technique of abolishing a large residual flow in and around a Nitinol duct occluder device. Medknow Publications & Media Pvt Ltd 2012 /pmc/articles/PMC3487218/ /pubmed/23129919 http://dx.doi.org/10.4103/0974-2069.99632 Text en Copyright: © Annals of Pediatric Cardiology http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Case Report
Sridhar, Anuradha
Subramanyan, Raghavan
Rajasekar, Premsekar
Transcatheter interruption of large residual flow after device closure of “Type A” patent ductus arteriosus
title Transcatheter interruption of large residual flow after device closure of “Type A” patent ductus arteriosus
title_full Transcatheter interruption of large residual flow after device closure of “Type A” patent ductus arteriosus
title_fullStr Transcatheter interruption of large residual flow after device closure of “Type A” patent ductus arteriosus
title_full_unstemmed Transcatheter interruption of large residual flow after device closure of “Type A” patent ductus arteriosus
title_short Transcatheter interruption of large residual flow after device closure of “Type A” patent ductus arteriosus
title_sort transcatheter interruption of large residual flow after device closure of “type a” patent ductus arteriosus
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3487218/
https://www.ncbi.nlm.nih.gov/pubmed/23129919
http://dx.doi.org/10.4103/0974-2069.99632
work_keys_str_mv AT sridharanuradha transcatheterinterruptionoflargeresidualflowafterdeviceclosureoftypeapatentductusarteriosus
AT subramanyanraghavan transcatheterinterruptionoflargeresidualflowafterdeviceclosureoftypeapatentductusarteriosus
AT rajasekarpremsekar transcatheterinterruptionoflargeresidualflowafterdeviceclosureoftypeapatentductusarteriosus