Comparison Between the Accuracy of NovApex Apex Locator and Radiographs in Determining Radiographic Apex

INTRODUCTION: Determination of the canal working length with radiographs has many drawbacks. Electronic apex locators have been developed to overcome some of these problems. Recently, a newly designed apex locator called NovApex has been introduced. All the studies conducted to determine the accurac...

Descripción completa

Detalles Bibliográficos
Autores principales: Shahi, Shahriar, Rahimi, Saeed, Milani, Amin Salem, Asghari Jafarabadi, Mohammad, Djoorabbaf Naghsh, Gholam Reza
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Iranian Center for Endodontic Research 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3487509/
https://www.ncbi.nlm.nih.gov/pubmed/23130055
Descripción
Sumario:INTRODUCTION: Determination of the canal working length with radiographs has many drawbacks. Electronic apex locators have been developed to overcome some of these problems. Recently, a newly designed apex locator called NovApex has been introduced. All the studies conducted to determine the accuracy of NovApex have been carried out in-vitro on extracted teeth. The aim of this in vivo study was to evaluate the accuracy of NovApex compared with traditional radiographic method. MATERIALS AND METHODS: Twenty-five patients whose lower molars that were candidates for extraction were selected. The teeth were accessed, and the radiographic working length was determined by measuring the length of the initial file 0.5mm short of the radiographic apex. Then, NovApex apex locator was used to measure the electronic working length. Subsequently, the tooth was extracted, and the actual working length was measured by introducing a size #15 K-file into each canal until the file tip was visible at the apex, then 1.5mm was subtracted to attain the working length. Cohen's Kappa was computed for each of the methods versus actual working length as a measure of reliability. The accuracies were compared using Chi-square test. RESULTS: The accuracy of NovApex apex locator and radiographic method in detecting the apical end point within ±0.5mm was 74.7% and 68%, respectively; this was not significantly different (P<0.001). CONCLUSION: The NovApex apex locator is useful in detecting the apical end point with the accuracy similar to radiographic method. However, neither technique is fully reliable in detecting the apical end point of the canal.