BACKGROUND: Malaria rapid diagnostic tests (RDTs) are protected from humidity-caused degradation by a desiccant added to the device packaging. The present study assessed malaria RDT products for the availability, type and design of desiccants and their information supplied in the instructions for us...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3493316/ https://www.ncbi.nlm.nih.gov/pubmed/22974115 http://dx.doi.org/10.1186/1475-2875-11-326 |
_version_ | 1782249241172770816 |
---|---|
author | Barbé, Barbara Gillet, Philippe Beelaert, Greet Fransen, Katrien Jacobs, Jan |
author_facet | Barbé, Barbara Gillet, Philippe Beelaert, Greet Fransen, Katrien Jacobs, Jan |
author_sort | Barbé, Barbara |
collection | PubMed |
description | BACKGROUND: Malaria rapid diagnostic tests (RDTs) are protected from humidity-caused degradation by a desiccant added to the device packaging. The present study assessed malaria RDT products for the availability, type and design of desiccants and their information supplied in the instructions for use (IFU). METHODS: Criteria were based on recommendations of the World Health Organization (WHO), the European Community (CE) and own observations. Silica gel sachets were defined as self-indicating (all beads coated with a humidity indicator that changes colour upon saturation), partial-indicating (part of beads coated) and non-indicating (none of the beads coated). Indicating silica gel sachets were individually assessed for humidity saturation and (in case of partial-indicating silica gels) for the presence of indicating beads. RESULTS: Fifty malaria RDT products from 25 manufacturers were assessed, 14 (28%) products were listed by the “Global Fund Quality Assurance Policy” and 31 (62%) were CE-marked. All but one product contained a desiccant, mostly (47/50, 94%) silica gel. Twenty (40%) RDT products (one with no desiccant and 19 with non-indicating desiccant) did not meet the WHO guidelines recommending indicating desiccant. All RDT products with self- or partial-indicating silica gel (n = 22 and 8 respectively) contained the toxic cobalt dichloride as humidity indicator. Colour change indicating humidity saturation was observed for 8/16 RDT products, at a median incidence of 0.8% (range 0.05%-4.6%) of sachets inspected. In all RDTs with partial-indicating silica gel, sachets with no colour indicating beads were found (median proportion 13.5% (0.6% - 17.8%) per product) and additional light was needed to assess the humidity colour. Less than half (14/30, 47%) IFUs of RDT products with indicating desiccants mentioned to check the humidity saturation before using the test. Information on properties, safety hazards and disposal of the desiccant was not included in any of the IFUs. There were no differences between Global Fund-listed and CE marked RDT products compared to those which were not. Similar findings were noted for a panel of 11 HIV RDTs that was assessed with the same checklist as the malaria RDTs. CONCLUSION: RDTs showed shortcomings in desiccant type and information supplied in the IFU. |
format | Online Article Text |
id | pubmed-3493316 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-34933162012-11-09 Assessment of desiccants and their instructions for use in rapid diagnostic tests Barbé, Barbara Gillet, Philippe Beelaert, Greet Fransen, Katrien Jacobs, Jan Malar J Research BACKGROUND: Malaria rapid diagnostic tests (RDTs) are protected from humidity-caused degradation by a desiccant added to the device packaging. The present study assessed malaria RDT products for the availability, type and design of desiccants and their information supplied in the instructions for use (IFU). METHODS: Criteria were based on recommendations of the World Health Organization (WHO), the European Community (CE) and own observations. Silica gel sachets were defined as self-indicating (all beads coated with a humidity indicator that changes colour upon saturation), partial-indicating (part of beads coated) and non-indicating (none of the beads coated). Indicating silica gel sachets were individually assessed for humidity saturation and (in case of partial-indicating silica gels) for the presence of indicating beads. RESULTS: Fifty malaria RDT products from 25 manufacturers were assessed, 14 (28%) products were listed by the “Global Fund Quality Assurance Policy” and 31 (62%) were CE-marked. All but one product contained a desiccant, mostly (47/50, 94%) silica gel. Twenty (40%) RDT products (one with no desiccant and 19 with non-indicating desiccant) did not meet the WHO guidelines recommending indicating desiccant. All RDT products with self- or partial-indicating silica gel (n = 22 and 8 respectively) contained the toxic cobalt dichloride as humidity indicator. Colour change indicating humidity saturation was observed for 8/16 RDT products, at a median incidence of 0.8% (range 0.05%-4.6%) of sachets inspected. In all RDTs with partial-indicating silica gel, sachets with no colour indicating beads were found (median proportion 13.5% (0.6% - 17.8%) per product) and additional light was needed to assess the humidity colour. Less than half (14/30, 47%) IFUs of RDT products with indicating desiccants mentioned to check the humidity saturation before using the test. Information on properties, safety hazards and disposal of the desiccant was not included in any of the IFUs. There were no differences between Global Fund-listed and CE marked RDT products compared to those which were not. Similar findings were noted for a panel of 11 HIV RDTs that was assessed with the same checklist as the malaria RDTs. CONCLUSION: RDTs showed shortcomings in desiccant type and information supplied in the IFU. BioMed Central 2012-09-13 /pmc/articles/PMC3493316/ /pubmed/22974115 http://dx.doi.org/10.1186/1475-2875-11-326 Text en Copyright ©2012 Barbé et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Barbé, Barbara Gillet, Philippe Beelaert, Greet Fransen, Katrien Jacobs, Jan Assessment of desiccants and their instructions for use in rapid diagnostic tests |
title | Assessment of desiccants and their instructions for use in rapid diagnostic tests |
title_full | Assessment of desiccants and their instructions for use in rapid diagnostic tests |
title_fullStr | Assessment of desiccants and their instructions for use in rapid diagnostic tests |
title_full_unstemmed | Assessment of desiccants and their instructions for use in rapid diagnostic tests |
title_short | Assessment of desiccants and their instructions for use in rapid diagnostic tests |
title_sort | assessment of desiccants and their instructions for use in rapid diagnostic tests |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3493316/ https://www.ncbi.nlm.nih.gov/pubmed/22974115 http://dx.doi.org/10.1186/1475-2875-11-326 |
work_keys_str_mv | AT barbebarbara assessmentofdesiccantsandtheirinstructionsforuseinrapiddiagnostictests AT gilletphilippe assessmentofdesiccantsandtheirinstructionsforuseinrapiddiagnostictests AT beelaertgreet assessmentofdesiccantsandtheirinstructionsforuseinrapiddiagnostictests AT fransenkatrien assessmentofdesiccantsandtheirinstructionsforuseinrapiddiagnostictests AT jacobsjan assessmentofdesiccantsandtheirinstructionsforuseinrapiddiagnostictests |