A Simple Sign for Recognizing Off–Axis OCT Measurement Beam Placement in the Context of Multicentre Studies

PURPOSE: Optical coherence tomography (OCT) allows quantification of the thickness of the retinal nerve fibre layer (RNFL) thickness, a potential biomarker for neurodegeneration. The estimated annual RNFL loss in multiple sclerosis amounts to 2 μm using time domain OCT. The recognition of measuremen...

Descripción completa

Detalles Bibliográficos
Autores principales: Balk, Lisanne J., de Vries–Knoppert, Willemien A. E. J., Petzold, Axel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3493550/
https://www.ncbi.nlm.nih.gov/pubmed/23144857
http://dx.doi.org/10.1371/journal.pone.0048222
_version_ 1782249285294751744
author Balk, Lisanne J.
de Vries–Knoppert, Willemien A. E. J.
Petzold, Axel
author_facet Balk, Lisanne J.
de Vries–Knoppert, Willemien A. E. J.
Petzold, Axel
author_sort Balk, Lisanne J.
collection PubMed
description PURPOSE: Optical coherence tomography (OCT) allows quantification of the thickness of the retinal nerve fibre layer (RNFL) thickness, a potential biomarker for neurodegeneration. The estimated annual RNFL loss in multiple sclerosis amounts to 2 μm using time domain OCT. The recognition of measurement artifacts exceeding this limit is relevant for the successful use of OCT as a secondary outcome measure in clinical trials. METHODS: Prospective study design. An exploratory pilot study (ring and volume scans) followed by a cohort study (1,980 OCT ring scans). The OCT measurement beam was placed off–axis to the left, right, top and bottom of the subjects pupil and RNFL thickness of these scans were compared to the centrally placed reference scans. RESULTS: Off–axis placement of the OCT measurement beam resulted in significant artifacts in RNFL thickness measurements (95%CI 9μm, maximal size of error 42μm). Off–axis placement gave characteristic patterns of the OCT live images which are not necessarily saved for review. Off–axis placement also causes regional inhomogeneity of reflectivity in the outer nuclear (ONL) and outer plexiform layers (OPL) which remains visible on scans saved for review. CONCLUSION: Off–axis beam placement introduces measurement artifacts at a magnitude which may mask recognition of RNFL loss due to neurodegeneration in multiple sclerosis. The resulting pattern in the OCT live image can only be recognised by the technician capturing the scans. Once the averaged scans have been aligned this pattern is lost. Retrospective identification of this artifact is however possible by presence of regional inhomogeneity of ONL/OPL reflectivity. This simple and robust sign may be considered for quality control criteria in the setting of multicentre OCT studies. The practical advice of this study is to keep the OCT image in the acquisition window horizontally aligned whenever possible.
format Online
Article
Text
id pubmed-3493550
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-34935502012-11-09 A Simple Sign for Recognizing Off–Axis OCT Measurement Beam Placement in the Context of Multicentre Studies Balk, Lisanne J. de Vries–Knoppert, Willemien A. E. J. Petzold, Axel PLoS One Research Article PURPOSE: Optical coherence tomography (OCT) allows quantification of the thickness of the retinal nerve fibre layer (RNFL) thickness, a potential biomarker for neurodegeneration. The estimated annual RNFL loss in multiple sclerosis amounts to 2 μm using time domain OCT. The recognition of measurement artifacts exceeding this limit is relevant for the successful use of OCT as a secondary outcome measure in clinical trials. METHODS: Prospective study design. An exploratory pilot study (ring and volume scans) followed by a cohort study (1,980 OCT ring scans). The OCT measurement beam was placed off–axis to the left, right, top and bottom of the subjects pupil and RNFL thickness of these scans were compared to the centrally placed reference scans. RESULTS: Off–axis placement of the OCT measurement beam resulted in significant artifacts in RNFL thickness measurements (95%CI 9μm, maximal size of error 42μm). Off–axis placement gave characteristic patterns of the OCT live images which are not necessarily saved for review. Off–axis placement also causes regional inhomogeneity of reflectivity in the outer nuclear (ONL) and outer plexiform layers (OPL) which remains visible on scans saved for review. CONCLUSION: Off–axis beam placement introduces measurement artifacts at a magnitude which may mask recognition of RNFL loss due to neurodegeneration in multiple sclerosis. The resulting pattern in the OCT live image can only be recognised by the technician capturing the scans. Once the averaged scans have been aligned this pattern is lost. Retrospective identification of this artifact is however possible by presence of regional inhomogeneity of ONL/OPL reflectivity. This simple and robust sign may be considered for quality control criteria in the setting of multicentre OCT studies. The practical advice of this study is to keep the OCT image in the acquisition window horizontally aligned whenever possible. Public Library of Science 2012-11-08 /pmc/articles/PMC3493550/ /pubmed/23144857 http://dx.doi.org/10.1371/journal.pone.0048222 Text en © 2012 Balk et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Balk, Lisanne J.
de Vries–Knoppert, Willemien A. E. J.
Petzold, Axel
A Simple Sign for Recognizing Off–Axis OCT Measurement Beam Placement in the Context of Multicentre Studies
title A Simple Sign for Recognizing Off–Axis OCT Measurement Beam Placement in the Context of Multicentre Studies
title_full A Simple Sign for Recognizing Off–Axis OCT Measurement Beam Placement in the Context of Multicentre Studies
title_fullStr A Simple Sign for Recognizing Off–Axis OCT Measurement Beam Placement in the Context of Multicentre Studies
title_full_unstemmed A Simple Sign for Recognizing Off–Axis OCT Measurement Beam Placement in the Context of Multicentre Studies
title_short A Simple Sign for Recognizing Off–Axis OCT Measurement Beam Placement in the Context of Multicentre Studies
title_sort simple sign for recognizing off–axis oct measurement beam placement in the context of multicentre studies
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3493550/
https://www.ncbi.nlm.nih.gov/pubmed/23144857
http://dx.doi.org/10.1371/journal.pone.0048222
work_keys_str_mv AT balklisannej asimplesignforrecognizingoffaxisoctmeasurementbeamplacementinthecontextofmulticentrestudies
AT devriesknoppertwillemienaej asimplesignforrecognizingoffaxisoctmeasurementbeamplacementinthecontextofmulticentrestudies
AT petzoldaxel asimplesignforrecognizingoffaxisoctmeasurementbeamplacementinthecontextofmulticentrestudies
AT balklisannej simplesignforrecognizingoffaxisoctmeasurementbeamplacementinthecontextofmulticentrestudies
AT devriesknoppertwillemienaej simplesignforrecognizingoffaxisoctmeasurementbeamplacementinthecontextofmulticentrestudies
AT petzoldaxel simplesignforrecognizingoffaxisoctmeasurementbeamplacementinthecontextofmulticentrestudies